Clinical Evaluation of Morcher Artificial Iris Diaphragms

Primary Purpose

Aniridia, Iris; Deformity

Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
Morcher iris diaphragm implantation
Sponsored by
University of California, Los Angeles
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Aniridia focused on measuring aniridia, Morcher iris diaphragms, modified capsule tension ring, artificial iris, iris reconstruction lens, glare sensitivity, photophobia

Eligibility Criteria

18 Years - undefined (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • Patient must be 18 years old and have partial or complete aniridia.
  • Be willing and able to comply with all follow-up requirements.
  • Must have increased light and/or glare sensitivity or complete aniridia.
  • Patients may be phakic, aphakic, or pseudophakic.
  • Phakic patients will require simultaneous cataract surgery.
  • Aphakic patients will require secondary intraocular lens implantation.

Exclusion Criteria:

  • The presence of any ocular condition that may cause complications from the surgical procedure
  • Active ocular infection or inflammation
  • Patients with allergies to operative and/or postoperative medications
  • Pregnant or lactating women
  • Persons who, in the determination of the investigator, are not competent to understand the procedure or the actions asked of them as research subjects

Sites / Locations

  • Stein Eye Institute, UCLA

Arms of the Study

Arm 1

Arm Type

Experimental

Arm Label

Morcher iris diaphragm implantation

Arm Description

This is a non-randomized, non-comparative interventional surgical series. Patients will undergo Morcher iris diaphragm implantation in their affected eye(s). After surgery, patients will complete 5 postoperative examinations. At each examination, they will be evaluated for changes in light and glare sensitivity and visual acuity. They will also be monitored for adverse reactions.

Outcomes

Primary Outcome Measures

Change in Light and Glare Sensitivity as Determined by a Clinical Glare Test (Primary Efficacy Measure)
The primary efficacy measure of the study was the change in corrected distance visual acuity (CDVA) under glare conditions. A transilluminator light was held just in front of the distance corrected study eye in one of four quadrants (above, below, nasal, or temporal) to the line of sight to evoke a glare response. The direction that produced the worst CDVA was called the CDVA with glare. A ≥ 2 line improvement in Snellen CDVA with glare following Morcher device implantation was considered a positive clinical change (glare sensitivity better). A ≥ 2 line worsening of Snellen CDVA with glare following Morcher device implantation was considered a negative clinical change (glare sensitivity worse). A change of < 2 Snellen lines following Morcher device implantation was considered to be a neutral change (glare sensitivity the same).
Change in Best Corrected Visual Acuity (Primary Safety Measure)
The primary safety measure of the study was the change in best corrected distance visual acuity (CDVA) as measured using a Snellen eye chart. A ≥ 2 line improvement in Snellen CDVA following Morcher device implantation was considered a positive clinical change (visual acuity better). A ≥ 2 line worsening of Snellen CDVA following Morcher device implantation was considered a negative clinical change (visual acuity worse). A change of < 2 lines was considered to be a neutral change (visual acuity the same).

Secondary Outcome Measures

Change in Glare Sensitivity Under Day Time Lighting Conditions (Secondary Efficacy Measure)
A secondary efficacy measure of the study was day time glare disability as assessed by subjective questionnaire. Glare disability was rated using a 0 to 10 scale, where 0 was considered very slight and 10 was considered very significant.
Change in Glare Sensitivity Under Night Time Lighting Conditions (Secondary Efficacy Measure)
Another secondary efficacy measure of the study was night time glare disability as assessed by subjective questionnaire. Glare disability was rated using a 0 to 10 scale, where 0 was considered very slight and 10 was considered very significant.
Change in Endothelial Cell Count (Secondary Safety Measure)
A secondary safety measure of the study was the change in endothelial cell count. A loss of >10% of central corneal endothelial cells was considered clinically significant at the onset of the study. (Note, the 10% loss criterion was established before the 67B implant was added to the list of study devices. The 67B implant has an expected greater cell loss than that associated with the 96F, 96S, 50D, and 50F modified capsule tension rings because it requires a larger incision for implantation. Thirty one (48.4%) of the 64 patients were implanted with the 67B device.)
Need to Explant or Exchange a Morcher Iris Diaphragm (Secondary Safety Measure)
Another secondary safety measure of the study was the need to explant or exchange a Morcher iris diaphragm within 1 year of implantation. Explantation in < 25% of patients was considered to be clinically acceptable.

Full Information

First Posted
December 19, 2008
Last Updated
December 7, 2021
Sponsor
University of California, Los Angeles
search

1. Study Identification

Unique Protocol Identification Number
NCT00812708
Brief Title
Clinical Evaluation of Morcher Artificial Iris Diaphragms
Official Title
Clinical Evaluation of Morcher Artificial Iris Diaphragms to Treat Light and Glare Sensitivity in Partial or Complete Aniridia
Study Type
Interventional

2. Study Status

Record Verification Date
December 2021
Overall Recruitment Status
Completed
Study Start Date
April 9, 2003 (Actual)
Primary Completion Date
June 20, 2014 (Actual)
Study Completion Date
December 7, 2021 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
University of California, Los Angeles

4. Oversight

Data Monitoring Committee
Yes

5. Study Description

Brief Summary
The purpose of this study is to determine the safety and efficacy of Morcher iris diaphragms in the treatment of congenital and acquired aniridia. Morcher iris diaphragms are intraocular devices that are designed to provide an artificial pupil for patients suffering from partial or complete aniridia. These devices are constructed from clinical quality, ultraviolet light-absorbing, opaque black polymethylmethacrylate (PMMA). After surgical implantation, patients are monitored over the course of 1 year to measure any changes to visual acuity and improvements in light and glare sensitivity.
Detailed Description
Aniridia is a condition in which the iris, the colored portion of the eye, is either partially or completely absent. The iris is responsible for regulating the amount of light entering the eye by adjusting the size of the pupil. Aniridia can either be a congenital condition, often a genetically-based abnormality affecting the formation of the iris, or it can be acquired by blunt trauma with or without rupture, penetrating trauma, or severe intraocular inflammation. Aniridia can also vary in the level of severity. A mild case might involve a thinner iris or iris remnant with a normal pupil. A severe case might involve complete absence of the iris or impairment of the muscles responsible for adjusting pupil size. Iris defects can occur without the loss of stromal tissue as well. In some cases, the iris pigment epithelium is missing. In other cases, holes are present in the iris or the pupil is chronically enlarged or mydriatic. People who suffer from iris defects also commonly suffer from other eye conditions. Common comorbidities include glaucoma, cataract, and nystagmus. Other structures in the eye are often affected, including the cornea, crystalline lens, zonules, and retina. Iris defects can cause severe visual disability if untreated. Common treatments for iris defects include the use of colored or tinted glasses or contact lenses to reduce the amount of light entering the eye. In this trial, several Morcher iris diaphragms are being evaluated as a potential treatment to improve the symptoms associated with aniridia and other iris defects. Morcher iris diaphragms are manufactured in Germany by Morcher GmbH [website: http:// www.morcher.com]. These devices have been utilized in Europe to treat aniridic patients for over 40 years. Additionally, these devices hold the European CE mark of conformity. When a patient is implanted with a Morcher iris device, surgery always involves additional standard of care surgical procedures. Typically, a device is implanted during cataract surgery along with an intraocular lens (IOL). Morcher devices can also be implanted during an IOL exchange with or without an anterior vitrectomy. All implantation procedures take place in an operating room under local or general anesthesia. In some cases, a Morcher iris reconstruction lens containing both an artificial iris and a lens is implanted. Also, depending on the condition of the eye, the lens and iris device may need to be sutured to the sclera, the white portion of the eye. All patients undergoing surgery are prescribed 2 different eye drops, an antibiotic and a corticosteroid. These medications are used for up to 6 weeks following implantation of the device. There are 5 postoperative follow-up visits that each patient must complete. Visits occur at specific intervals over the course of a 1-year period. At each visit, standard of care procedures are performed and, at certain time points during the follow-up interval, digital photos and endothelial cell counts are obtained. In instances where both eyes are implanted, surgery for the second eye is scheduled 6 months or more later.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Aniridia, Iris; Deformity
Keywords
aniridia, Morcher iris diaphragms, modified capsule tension ring, artificial iris, iris reconstruction lens, glare sensitivity, photophobia

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
72 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Morcher iris diaphragm implantation
Arm Type
Experimental
Arm Description
This is a non-randomized, non-comparative interventional surgical series. Patients will undergo Morcher iris diaphragm implantation in their affected eye(s). After surgery, patients will complete 5 postoperative examinations. At each examination, they will be evaluated for changes in light and glare sensitivity and visual acuity. They will also be monitored for adverse reactions.
Intervention Type
Device
Intervention Name(s)
Morcher iris diaphragm implantation
Other Intervention Name(s)
Morcher device models: 96F, 96S, 50D, 50F, and 67B
Intervention Description
Surgical implantation of Morcher iris device(s)
Primary Outcome Measure Information:
Title
Change in Light and Glare Sensitivity as Determined by a Clinical Glare Test (Primary Efficacy Measure)
Description
The primary efficacy measure of the study was the change in corrected distance visual acuity (CDVA) under glare conditions. A transilluminator light was held just in front of the distance corrected study eye in one of four quadrants (above, below, nasal, or temporal) to the line of sight to evoke a glare response. The direction that produced the worst CDVA was called the CDVA with glare. A ≥ 2 line improvement in Snellen CDVA with glare following Morcher device implantation was considered a positive clinical change (glare sensitivity better). A ≥ 2 line worsening of Snellen CDVA with glare following Morcher device implantation was considered a negative clinical change (glare sensitivity worse). A change of < 2 Snellen lines following Morcher device implantation was considered to be a neutral change (glare sensitivity the same).
Time Frame
Preoperatively and 1 year postoperatively
Title
Change in Best Corrected Visual Acuity (Primary Safety Measure)
Description
The primary safety measure of the study was the change in best corrected distance visual acuity (CDVA) as measured using a Snellen eye chart. A ≥ 2 line improvement in Snellen CDVA following Morcher device implantation was considered a positive clinical change (visual acuity better). A ≥ 2 line worsening of Snellen CDVA following Morcher device implantation was considered a negative clinical change (visual acuity worse). A change of < 2 lines was considered to be a neutral change (visual acuity the same).
Time Frame
Preoperatively and 1 year postoperatively
Secondary Outcome Measure Information:
Title
Change in Glare Sensitivity Under Day Time Lighting Conditions (Secondary Efficacy Measure)
Description
A secondary efficacy measure of the study was day time glare disability as assessed by subjective questionnaire. Glare disability was rated using a 0 to 10 scale, where 0 was considered very slight and 10 was considered very significant.
Time Frame
Preoperatively and 3 months postoperatively
Title
Change in Glare Sensitivity Under Night Time Lighting Conditions (Secondary Efficacy Measure)
Description
Another secondary efficacy measure of the study was night time glare disability as assessed by subjective questionnaire. Glare disability was rated using a 0 to 10 scale, where 0 was considered very slight and 10 was considered very significant.
Time Frame
Preoperatively and 3 months postoperatively
Title
Change in Endothelial Cell Count (Secondary Safety Measure)
Description
A secondary safety measure of the study was the change in endothelial cell count. A loss of >10% of central corneal endothelial cells was considered clinically significant at the onset of the study. (Note, the 10% loss criterion was established before the 67B implant was added to the list of study devices. The 67B implant has an expected greater cell loss than that associated with the 96F, 96S, 50D, and 50F modified capsule tension rings because it requires a larger incision for implantation. Thirty one (48.4%) of the 64 patients were implanted with the 67B device.)
Time Frame
Preoperatively and 3 months postoperatively
Title
Need to Explant or Exchange a Morcher Iris Diaphragm (Secondary Safety Measure)
Description
Another secondary safety measure of the study was the need to explant or exchange a Morcher iris diaphragm within 1 year of implantation. Explantation in < 25% of patients was considered to be clinically acceptable.
Time Frame
Preoperatively and 1 year postoperatively

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Patient must be 18 years old and have partial or complete aniridia. Be willing and able to comply with all follow-up requirements. Must have increased light and/or glare sensitivity or complete aniridia. Patients may be phakic, aphakic, or pseudophakic. Phakic patients will require simultaneous cataract surgery. Aphakic patients will require secondary intraocular lens implantation. Exclusion Criteria: The presence of any ocular condition that may cause complications from the surgical procedure Active ocular infection or inflammation Patients with allergies to operative and/or postoperative medications Pregnant or lactating women Persons who, in the determination of the investigator, are not competent to understand the procedure or the actions asked of them as research subjects
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Kevin M Miller, MD
Organizational Affiliation
University of California, Los Angeles
Official's Role
Principal Investigator
Facility Information:
Facility Name
Stein Eye Institute, UCLA
City
Los Angeles
State/Province
California
ZIP/Postal Code
90095-7000
Country
United States

12. IPD Sharing Statement

Plan to Share IPD
No
Citations:
PubMed Identifier
19265334
Citation
Chung MY, Miller KM, Weissman BA. Morcher iris reconstruction lens and rigid contact lens for traumatic aniridia. Eye Contact Lens. 2009 Mar;35(2):108-10. doi: 10.1097/ICL.0b013e318199b00b.
Results Reference
background
PubMed Identifier
18812117
Citation
Olson MD, Masket S, Miller KM. Interim results of a compassionate-use clinical trial of Morcher iris diaphragm implantation: report 1. J Cataract Refract Surg. 2008 Oct;34(10):1674-80. doi: 10.1016/j.jcrs.2008.05.048.
Results Reference
result
PubMed Identifier
26603402
Citation
Date RC, Olson MD, Shah M, Masket S, Miller KM. Outcomes of a modified capsular tension ring with a single black occluder paddle for eyes with congenital and acquired iris defects: Report 2. J Cataract Refract Surg. 2015 Sep;41(9):1934-44. doi: 10.1016/j.jcrs.2015.10.001.
Results Reference
result
PubMed Identifier
27373394
Citation
Miller KM, Nicoli CM, Olson MD, Shah M, Masket S. Outcomes of implantation of modified capsule tension rings with multiple black occluder paddles for eyes with congenital and acquired iris defects: Report 3. J Cataract Refract Surg. 2016 Jun;42(6):870-8. doi: 10.1016/j.jcrs.2016.03.035.
Results Reference
result
PubMed Identifier
29515428
Citation
Lin SR, Miller KM. Lessons Learned from Implantation of Morcher 50D and 96S Artificial Iris Diaphragms. Case Rep Ophthalmol. 2017 Nov 23;8(3):527-534. doi: 10.1159/000484128. eCollection 2017 Sep-Dec.
Results Reference
result
PubMed Identifier
30041739
Citation
Miller KM, Kuo A, Olson MD, Masket S. Safety and efficacy of black iris diaphragm intraocular lens implantation in eyes with large iris defects: Report 4. J Cataract Refract Surg. 2018 Jun;44(6):686-700. doi: 10.1016/j.jcrs.2018.03.033. Epub 2018 Jul 4.
Results Reference
result
Links:
URL
http://bmctoday.net/advancedocularcare/2010/12/article.asp?f=iris-prosthetics
Description
Related article: Iris Prosthetics. Advanced Ocular Care. December 2010