Safety of the Colonoscope Magnetic Imaging Device (ScopeGuide®) in Patients With Implantable Cardiac Devices

Primary Purpose

Cardiac Arrhythmia

Status
Completed
Phase
Not Applicable
Locations
United Kingdom
Study Type
Interventional
Intervention
Exposure to ScopeGuide
Sponsored by
Cambridge University Hospitals NHS Foundation Trust
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for Cardiac Arrhythmia

Eligibility Criteria

18 Years - 80 Years (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • Participant is willing and able to give informed consent for participation in the study.
  • Male or Female, aged 18 years or above.
  • Must have a permanent pacemaker or implantable cardioverter defibrillator
  • Is attending for a routine pacemaker check during the study period

Exclusion Criteria:

The participant may not enter the study if ANY of the following apply:

  • Cardio-respiratory compromise on the day of the experiment (we propose NYHA Grade 3 or higher as criteria for this)
  • Vulnerable adults

Sites / Locations

  • Addenbrooke's Hospital

Arms of the Study

Arm 1

Arm Type

Experimental

Arm Label

Cardiac device patients.

Arm Description

Exposure to ScopeGuide - patients with cardiac devices exposed to ScopeGuide as per study protocol

Outcomes

Primary Outcome Measures

Recruitment target met.
The study has a target recruitment of 100 patients. When this is met the study will stop.

Secondary Outcome Measures

Interference seen with implanted cardiac device
If interference is seen in any of the recruited patients the study will be stopped.

Full Information

First Posted
April 3, 2013
Last Updated
April 5, 2013
Sponsor
Cambridge University Hospitals NHS Foundation Trust
search

1. Study Identification

Unique Protocol Identification Number
NCT01825525
Brief Title
Safety of the Colonoscope Magnetic Imaging Device (ScopeGuide®) in Patients With Implantable Cardiac Devices
Official Title
Safety of the Colonoscope Magnetic Imaging Device (ScopeGuide®) in Patients With Implantable Cardiac Devices
Study Type
Interventional

2. Study Status

Record Verification Date
April 2013
Overall Recruitment Status
Completed
Study Start Date
August 2012 (undefined)
Primary Completion Date
April 2013 (Actual)
Study Completion Date
April 2013 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
Cambridge University Hospitals NHS Foundation Trust

4. Oversight

Data Monitoring Committee
Yes

5. Study Description

Brief Summary
The Olympus ScopeGuide® is a useful tool in speeding up colonoscopy and making it less painful for patients. It is currently contraindicated by the manufacturer for use with patients with implantable electronic devices (such as cardiac pacemakers). There is no evidence to back up this contraindication and with the number of pacemakers being inserted increasing by around 30 000 annually in the UK, more patients will be denied the use of this device should they require a colonoscopy. Participants with Pacemakers or Implantable Cardioverter Defibrillators will attend their check appointment. Following this they will be asked to lie with a colonoscope connected to the ScopeGuide® placed onto their abdomen (with clothes on) for 2 minutes. During this time they will be monitored for any irregular cardiac activity which may indicate electromagnetic interference. After this a second device check will occur and the participant will be sent home. The hypothesis is that the ScopeGuide will not interfere with the cardiac devices.
Detailed Description
BACKGROUND AND RATIONALE Colonoscopy is the 'gold-standard' diagnostic procedure for visualising the inside of the large bowel. It involves the insertion of a flexible instrument called a "colonoscope" around the large bowel. During insertion, looping of the instrument inside the colon can lead to incomplete procedures and significant patient discomfort. The development of the colonoscope magnetic imaging device (Olympus ScopeGuide®) has allowed the doctor performing the examination to detect looping of the instrument more easily. It has been shown in several studies that the duration, patient experience and completion rate of colonoscopy can be significantly improved by the use of ScopeGuide®. The manufacturer advises against the use of ScopeGuide® for patients with implantable devices due to the theoretical hazard of electromagnetic interference leading to device malfunction. Therefore, patients with permanent pacemakers and implantable cardioverter-defibrillators (ICDs) who require colonoscopy are denied the potential benefits offered by ScopeGuide®. There is no documented evidence for this risk from case reports or studies and the investigators have used a simulator to demonstrate that ScopeGuide® caused no interference or malfunction. The investigators propose an experimental study to evaluate the impact of the electromagnetic field created by ScopeGuide® on the implanted cardiac devices in patients. The study will be conducted in the Pacemaker Outpatient Clinic at Addenbrooke's hospital by a team of investigators including gastroenterologists and cardiologists. After giving their consent to participate, enrolled patients will have the activity of their cardiac device closely monitored while in close proximity to ScopeGuide® and any effect of this on the cardiac devices will be recorded. The investigators hypothesise that ScopeGuide® does not create a magnetic field significant enough to interfere with the normal function of these devices. If true, patients with implanted cardiac devices can also be offered the benefit ScopeGuide® provides during colonoscopy. The justification for this research is that the ScopeGuide® is currently not used in patients with implantable electronic cardiac devices (pacemakers and implantable cardioverter defibrillators(ICDs)) due to the manufacturers advice. There are no case reports and no studies in the medical literature to confirm or refute the advice and the investigators have demonstrated no evidence of interference using a simulator. The number of patients with either a pacemaker or ICD is increasingly annually and with services such as the NHS bowel cancer screening programme in which colonoscopy is offered to patients with blood in their stool more of these patients are having this test. To deny these patients the use of the ScopeGuide®, which has been shown to reduce procedure duration, reduce patient discomfort and increase completion rate without evidence is unjustified. The investigators therefore believe that this project is justified to potentially allow patients with pacemakers or ICDs the benefit of ScopeGuide®. If the investigators demonstrate evidence of interference, it would demonstrate a lack of safety in this patient group. Due to the lack of data for the safety of ScopeGuide® for patients with implantable cardiac devices we seek to question the current absolute contraindication imposed by the manufacturer. The main question to be answered by this study is whether implantable cardiac devices exhibit interference when exposed to the ScopeGuide®. The population to be studied are individuals with implantable cardiac devices, attending their regular device check appointment at Addenbrooke's hospital. The age range will be 18 to 80 years. Those being studied will be able to give informed consent. OBJECTIVES Primary Objective The primary objective of this study is to demonstrate in 100 serial participants whether the ScopeGuide® causes interference with implantable cardiac devices. Secondary Objectives None STUDY DESIGN Summary of Study Design This study is an uncontrolled trial. The study method will be undertaken on 100 consecutive consenting participants. The participants will be required for one visit only. This will be during their routine implantable device check appointment. Participation will extend their routine check appointment time by around 30 minutes. Once they have completed the study procedure they will not be required for follow up visits. The duration of the study for an individual participant is as follows: Attend regular cardiac device check appointment (around 20 minutes) Exposure to ScopeGuide® device while being monitored (around 10 minutes) Repeat cardiac device check (around 20 minutes)

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Cardiac Arrhythmia

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
100 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Cardiac device patients.
Arm Type
Experimental
Arm Description
Exposure to ScopeGuide - patients with cardiac devices exposed to ScopeGuide as per study protocol
Intervention Type
Device
Intervention Name(s)
Exposure to ScopeGuide
Intervention Description
The 2 minute protocol exposure to the electromagnetic field of the ScopeGuide device.
Primary Outcome Measure Information:
Title
Recruitment target met.
Description
The study has a target recruitment of 100 patients. When this is met the study will stop.
Time Frame
6 to 8 months
Secondary Outcome Measure Information:
Title
Interference seen with implanted cardiac device
Description
If interference is seen in any of the recruited patients the study will be stopped.
Time Frame
6 to 8 months

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Maximum Age & Unit of Time
80 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Participant is willing and able to give informed consent for participation in the study. Male or Female, aged 18 years or above. Must have a permanent pacemaker or implantable cardioverter defibrillator Is attending for a routine pacemaker check during the study period Exclusion Criteria: The participant may not enter the study if ANY of the following apply: Cardio-respiratory compromise on the day of the experiment (we propose NYHA Grade 3 or higher as criteria for this) Vulnerable adults
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Ewen AB Cameron, MD MB ChB
Organizational Affiliation
Cambridge University Hospitals NHS Foundation Trust
Official's Role
Study Director
Facility Information:
Facility Name
Addenbrooke's Hospital
City
Cambridge
State/Province
Cambridgeshire
ZIP/Postal Code
CB2 0QQ
Country
United Kingdom

12. IPD Sharing Statement

Citations:
PubMed Identifier
24477369
Citation
Corbett GD, Lim YC, Lee JC, Chernolesskiy A, Pugh PJ, Cameron EA. Safety of the colonoscope magnetic imaging device (ScopeGuide) in patients with implantable cardiac devices. Endoscopy. 2014 Feb;46(2):135-8. doi: 10.1055/s-0033-1359044. Epub 2014 Jan 29.
Results Reference
derived

Learn more about this trial

Safety of the Colonoscope Magnetic Imaging Device (ScopeGuide®) in Patients With Implantable Cardiac Devices