Clinical Evaluation of Cardiac Output Estimation Based on Pulse Wave Transit Time (esCCO)

Primary Purpose

Myocardial Infarction, Myocardial Ischemia

Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
Pulmonary Artery Catheter (PAC)
Endotracheal Cardiac Output Monitor (ECOM)
Estimated Continuous Cardiac Output (esCCO)
Sponsored by
University of California, San Francisco
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for Myocardial Infarction focused on measuring cardiac output, pulse oximetry, hemodynamics

Eligibility Criteria

19 Years - undefined (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • Patients undergoing surgery where cardiac output will be measured with PAC and ECOM.

Exclusion Criteria:

  • Failure to obtain consent.

Sites / Locations

  • San Francisco VA Medical Center
  • Drexel University

Arms of the Study

Arm 1

Arm Type

Active Comparator

Arm Label

Cardiac output monitoring

Arm Description

Patients undergoing surgery who will have their cardiac output monitored by pulmonary artery catheter (PAC) and endotracheal cardiac output monitoring (ECOM) during surgery and during post-surgical recovery, will also have sensors placed on the arm, finger and leg to calculate pulse wave transit time (PWTT) using the estimated Continuous Cardiac Output system (ecCCO).

Outcomes

Primary Outcome Measures

esCCO (Estimated Continuous Cardiac Output) Monitor
6 Measurements of cardiac output derived from pulse oximeter measurements using the esCCO system were taken at each time point. The measurements deemed valid under the criteria in the protocol were averaged to represent the reference value at that point. The mean and standard deviation reported consist of the reference values from all time points measured. Data for this test were taken at the following timepoints: 1. Start of Sterenotomy; 2. Before Bypass; 3. 30 Min after bypass; 4. Closure; 5. ICU arrival; 6. 6 hours in ICU; 7. 12 Hours in ICU; 8. 18 Hours in ICU (if PAC still in); 9. 24 Hours in ICU (If PAC still in)
PAC (Pulmonary Artery Catheter).
Measurements of cardiac output derived from a PAC (pulmonary artery catheter) using the standard thermodilution technique. At each time point, at least 6 measurements were taken. If, in the opinion of the clinician taking the readings, some of these were in error, more readings were taken to ensure accuracy. In no case were more than 18 readings taken. The points deemed valid by the clinician were averaged to obtain the reference value for that time point. The mean and standard deviation reported consisted of the reference values from all time points measured. Data for this test were taken at the following timepoints: 1. Start of Sterenotomy; 2. Before Bypass; 3. 30 Min after bypass; 4. Closure; 5. ICU arrival; 6. 6 hours in ICU; 7. 12 Hours in ICU; 8. 18 Hours in ICU (if PAC still in); 9. 24 Hours in ICU (If PAC still in)
Cardiac Output as Measured by ECOM
Measurements of cardiac output derived from the Endotracheal Cardiac Output Monitor (ECOM): An FDA-approved medical device is inserted into the patient's throat; cardiac output is calculated by measuring how electricity moves through the chest.

Secondary Outcome Measures

Full Information

First Posted
December 5, 2012
Last Updated
December 30, 2014
Sponsor
University of California, San Francisco
Collaborators
Northern California Institute of Research and Education, San Francisco Veterans Affairs Medical Center, Drexel University, Nihon Kohden
search

1. Study Identification

Unique Protocol Identification Number
NCT01877941
Brief Title
Clinical Evaluation of Cardiac Output Estimation Based on Pulse Wave Transit Time
Acronym
esCCO
Official Title
Clinical Evaluation of Cardiac Output Estimation Based on Pulse Wave Transit Time
Study Type
Interventional

2. Study Status

Record Verification Date
December 2014
Overall Recruitment Status
Completed
Study Start Date
December 2012 (undefined)
Primary Completion Date
November 2013 (Actual)
Study Completion Date
December 2013 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
University of California, San Francisco
Collaborators
Northern California Institute of Research and Education, San Francisco Veterans Affairs Medical Center, Drexel University, Nihon Kohden

4. Oversight

Data Monitoring Committee
No

5. Study Description

Brief Summary
The specific aim of this study is to assess whether a device approved by the FDA to measure pulse oximetry can accurately estimate cardiac output.
Detailed Description
We will compare three methods for measuring cardiac output: esCCO, PAC, and ECOM. Cardiac output is an important physiological parameter that must be monitored closely during surgery and during an Intensive Care Unit (ICU) stay. The use of a pulmonary artery catheter (PAC) has been the gold standard for accurate cardiac output (CO) measurement. This method requires a catheter to be inserted into the pulmonary artery and cardiac output is indicated by the speed that a temperature gradient dissipates. While the use of the pulmonary artery catheter is widespread, inserting the pulmonary artery catheter is risky, time consuming and requires a high level of skill. For these reasons, new devices have been developed to measure cardiac output. A second method typically used is called ECOM (Endotracheal Cardiac Output Monitor). A device is inserted into the patient's throat that calculates cardiac output by measuring how electricity moves through the patient's chest. The third method, estimated Continuous Cardiac Output (esCCO), uses sensors placed on the patient's finger, arm and chest to calculate Pulse Wave Transit Time (PWTT); the time it takes for the heart beat pulse to travel through the patient's body. The esCCO system is FDA approved for safety and efficacy to measure noninvasive blood pressure and pulse oximetry. The purpose of this study is to assess whether the PSTT calculation provides a cardiac output measure that is comparable to the other methods. This study is sponsored by Nihon Kohden Corporation who owns and manufactures the esCCO system.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Myocardial Infarction, Myocardial Ischemia
Keywords
cardiac output, pulse oximetry, hemodynamics

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
Outcomes Assessor
Allocation
N/A
Enrollment
39 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Cardiac output monitoring
Arm Type
Active Comparator
Arm Description
Patients undergoing surgery who will have their cardiac output monitored by pulmonary artery catheter (PAC) and endotracheal cardiac output monitoring (ECOM) during surgery and during post-surgical recovery, will also have sensors placed on the arm, finger and leg to calculate pulse wave transit time (PWTT) using the estimated Continuous Cardiac Output system (ecCCO).
Intervention Type
Device
Intervention Name(s)
Pulmonary Artery Catheter (PAC)
Intervention Description
A catheter is inserted into the pulmonary artery, through the internal jugular vein; cardiac output is indicated by the speed that a temperature gradient dissipates.
Intervention Type
Device
Intervention Name(s)
Endotracheal Cardiac Output Monitor (ECOM)
Intervention Description
An FDA-approved medical device is inserted into the patient's throat; cardiac output is calculated by measuring how electricity moves through the chest.
Intervention Type
Device
Intervention Name(s)
Estimated Continuous Cardiac Output (esCCO)
Intervention Description
Sensors are placed on the arm, finger and leg to calculate Pulse Wave Transit Time (PWTT); the time it takes for the pulse of the heartbeat to travel through the body.
Primary Outcome Measure Information:
Title
esCCO (Estimated Continuous Cardiac Output) Monitor
Description
6 Measurements of cardiac output derived from pulse oximeter measurements using the esCCO system were taken at each time point. The measurements deemed valid under the criteria in the protocol were averaged to represent the reference value at that point. The mean and standard deviation reported consist of the reference values from all time points measured. Data for this test were taken at the following timepoints: 1. Start of Sterenotomy; 2. Before Bypass; 3. 30 Min after bypass; 4. Closure; 5. ICU arrival; 6. 6 hours in ICU; 7. 12 Hours in ICU; 8. 18 Hours in ICU (if PAC still in); 9. 24 Hours in ICU (If PAC still in)
Time Frame
During and after surgery, up to 24 hours
Title
PAC (Pulmonary Artery Catheter).
Description
Measurements of cardiac output derived from a PAC (pulmonary artery catheter) using the standard thermodilution technique. At each time point, at least 6 measurements were taken. If, in the opinion of the clinician taking the readings, some of these were in error, more readings were taken to ensure accuracy. In no case were more than 18 readings taken. The points deemed valid by the clinician were averaged to obtain the reference value for that time point. The mean and standard deviation reported consisted of the reference values from all time points measured. Data for this test were taken at the following timepoints: 1. Start of Sterenotomy; 2. Before Bypass; 3. 30 Min after bypass; 4. Closure; 5. ICU arrival; 6. 6 hours in ICU; 7. 12 Hours in ICU; 8. 18 Hours in ICU (if PAC still in); 9. 24 Hours in ICU (If PAC still in)
Time Frame
During and post-surgery, up to 24 hours
Title
Cardiac Output as Measured by ECOM
Description
Measurements of cardiac output derived from the Endotracheal Cardiac Output Monitor (ECOM): An FDA-approved medical device is inserted into the patient's throat; cardiac output is calculated by measuring how electricity moves through the chest.
Time Frame
During and post-surgery, up to 8 hours

10. Eligibility

Sex
All
Minimum Age & Unit of Time
19 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Patients undergoing surgery where cardiac output will be measured with PAC and ECOM. Exclusion Criteria: Failure to obtain consent.
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Arthur Wallace, MD, PhD
Organizational Affiliation
San Francisco Veteran's Admnistration Medical Center
Official's Role
Principal Investigator
Facility Information:
Facility Name
San Francisco VA Medical Center
City
San Francisco
State/Province
California
ZIP/Postal Code
94121
Country
United States
Facility Name
Drexel University
City
Philadelphia
State/Province
Pennsylvania
ZIP/Postal Code
19102
Country
United States

12. IPD Sharing Statement

Learn more about this trial

Clinical Evaluation of Cardiac Output Estimation Based on Pulse Wave Transit Time

We'll reach out to this number within 24 hrs