In Vitro Assessment of a Breath-synchronized Vibrating Mesh Nebulizer During Non Invasive Ventilation (Synchro-Neb)
Primary Purpose
Respiratory Diseases, Lung Diseases, Cystic Fibrosis
Status
Completed
Phase
Not Applicable
Locations
Belgium
Study Type
Interventional
Intervention
Nebulization of Amikacin during NIV (RR: 15 cycles/minute)
Nebulization of Amikacin during NIV (RR: 25 cycles/minute)
Sponsored by
About this trial
This is an interventional treatment trial for Respiratory Diseases focused on measuring aerosol delivery, vibrating mesh nebulizer, breath-actuated nebulizer, non invasive ventilation, bilevel ventilator, single limb circuit ventilator, Amikacin
Eligibility Criteria
Inclusion Criteria:
- Not applicable (in vitro study)
Exclusion Criteria:
- hypersensitivity (allergic) reactions to aminoglycosides
Sites / Locations
- University Hospital St Luc
Arms of the Study
Arm 1
Arm 2
Arm Type
Experimental
Experimental
Arm Label
Breath-actuated vibrating mesh nebulizer
Conventional vibrating mesh nebulizer
Arm Description
500 mg/4 mL of Amikacin solution delivered with an experimental breath-actuated vibrating mesh nebulizer associated with a single limb circuit ventilator.
500 mg/4 mL of Amikacin solution delivered with a conventional vibrating mesh nebulizer (in continuous mode) associated with a single limb circuit ventilator.
Outcomes
Primary Outcome Measures
Inhaled dose
The inhaled dose assessed by residual gravimetric method
Secondary Outcome Measures
Expiratory wasted dose
The dose expelled in the ambient air through the exhalation port assessed by residual gravimetric method
Full Information
NCT ID
NCT02084043
First Posted
March 8, 2014
Last Updated
June 5, 2015
Sponsor
University Hospital St Luc, Brussels
Collaborators
Université Catholique de Louvain, University of Applied Sciences of Western Switzerland, School of Gestion and Engineering Vaud, Switzerland
1. Study Identification
Unique Protocol Identification Number
NCT02084043
Brief Title
In Vitro Assessment of a Breath-synchronized Vibrating Mesh Nebulizer During Non Invasive Ventilation
Acronym
Synchro-Neb
Official Title
In Vitro Comparison of Continuous and Breath-synchronized Vibrating Mesh Nebulizer During Non Invasive Ventilation: Analysis of Inhaled and Lost Doses.
Study Type
Interventional
2. Study Status
Record Verification Date
June 2015
Overall Recruitment Status
Completed
Study Start Date
March 2014 (undefined)
Primary Completion Date
June 2015 (Actual)
Study Completion Date
June 2015 (Actual)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
University Hospital St Luc, Brussels
Collaborators
Université Catholique de Louvain, University of Applied Sciences of Western Switzerland, School of Gestion and Engineering Vaud, Switzerland
4. Oversight
Data Monitoring Committee
No
5. Study Description
Brief Summary
Using an adult lung bench model of non invasive ventilation, the aim of the study is to compare an experimental system of breath-synchronized vibrating mesh nebulizer to a conventional vibrating mesh nebulizer during non invasive ventilation in terms of inhaled and lost doses.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Respiratory Diseases, Lung Diseases, Cystic Fibrosis, COPD, Asthma
Keywords
aerosol delivery, vibrating mesh nebulizer, breath-actuated nebulizer, non invasive ventilation, bilevel ventilator, single limb circuit ventilator, Amikacin
7. Study Design
Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Crossover Assignment
Masking
None (Open Label)
Allocation
Randomized
Enrollment
3 (Actual)
8. Arms, Groups, and Interventions
Arm Title
Breath-actuated vibrating mesh nebulizer
Arm Type
Experimental
Arm Description
500 mg/4 mL of Amikacin solution delivered with an experimental breath-actuated vibrating mesh nebulizer associated with a single limb circuit ventilator.
Arm Title
Conventional vibrating mesh nebulizer
Arm Type
Experimental
Arm Description
500 mg/4 mL of Amikacin solution delivered with a conventional vibrating mesh nebulizer (in continuous mode) associated with a single limb circuit ventilator.
Intervention Type
Drug
Intervention Name(s)
Nebulization of Amikacin during NIV (RR: 15 cycles/minute)
Other Intervention Name(s)
Amikacine sulfate
Intervention Description
500 mg/4 mL Amikacin nebulized using vibrating mesh nebulizer associated with a single limb bilevel ventilator. The nebulizations are considered as finished when there is no visible evidence of nebulization for a period of 30 seconds. The NIV is set with an IPAP of 15 cmH2O and EPAP of 5 cmH2O. The lung model is simulated with a respiratory rate of 15 cycles/minute
Intervention Type
Drug
Intervention Name(s)
Nebulization of Amikacin during NIV (RR: 25 cycles/minute)
Other Intervention Name(s)
Amikacine sulfate
Intervention Description
500 mg/4 mL Amikacin nebulized using vibrating mesh nebulizer associated with a single limb bilevel ventilator. The nebulizations are considered as finished when there is no visible evidence of nebulization for a period of 30 seconds. The NIV is set with an IPAP of 15 cmH2O and EPAP of 5 cmH2O. The lung model is simulated with a respiratory rate of 25 cycles/minute
Primary Outcome Measure Information:
Title
Inhaled dose
Description
The inhaled dose assessed by residual gravimetric method
Time Frame
after 24 hours
Secondary Outcome Measure Information:
Title
Expiratory wasted dose
Description
The dose expelled in the ambient air through the exhalation port assessed by residual gravimetric method
Time Frame
after 24 hours
Other Pre-specified Outcome Measures:
Title
Estimated lost dose
Description
The dose lost into the circuit (nebulizer included) assessed by residual gravimetric method
Time Frame
after 24 hours
10. Eligibility
Sex
All
Accepts Healthy Volunteers
Accepts Healthy Volunteers
Eligibility Criteria
Inclusion Criteria:
Not applicable (in vitro study)
Exclusion Criteria:
hypersensitivity (allergic) reactions to aminoglycosides
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Jean-Bernard Michotte, PhD student
Organizational Affiliation
University of Health Sciences (HESAV)
Official's Role
Principal Investigator
First Name & Middle Initial & Last Name & Degree
Jonathan Dugernier, PhD student
Organizational Affiliation
Cliniques universitaires Saint-Luc, service des soins intensifs
Official's Role
Principal Investigator
First Name & Middle Initial & Last Name & Degree
Enrico Staderini, PhD
Organizational Affiliation
School of Gestion and Engineering Vaud (HEIG-VD)
Official's Role
Principal Investigator
First Name & Middle Initial & Last Name & Degree
Rares Rusu, MSc
Organizational Affiliation
School of Gestion and Engineering Vaud (HEIG-VD)
Official's Role
Study Chair
First Name & Middle Initial & Last Name & Degree
Jean Roeseler, PhD
Organizational Affiliation
Cliniques universitaires Saint-Luc, service des soins intensifs
Official's Role
Study Chair
First Name & Middle Initial & Last Name & Degree
Giuseppe Liistro, MD PhD
Organizational Affiliation
Cliniques universitaires Saint-Luc, service de pneumologie
Official's Role
Study Chair
First Name & Middle Initial & Last Name & Degree
Gregory Reychler, PhD
Organizational Affiliation
Cliniques universitaires Saint-Luc, service de pneumologie
Official's Role
Study Director
Facility Information:
Facility Name
University Hospital St Luc
City
Brussels
ZIP/Postal Code
1200
Country
Belgium
12. IPD Sharing Statement
Learn more about this trial
In Vitro Assessment of a Breath-synchronized Vibrating Mesh Nebulizer During Non Invasive Ventilation
We'll reach out to this number within 24 hrs