Validation of Sleep Apnea Screening Device
Primary Purpose
Obstructive Sleep Apnea
Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
Zansors® sleep screening device
Sponsored by
About this trial
This is an interventional diagnostic trial for Obstructive Sleep Apnea focused on measuring obstructive sleep apnea screening device
Eligibility Criteria
Inclusion Criteria:
- Must be over 18 and referred by medical staff for an overnight assessment for suspected sleep apnea
Exclusion Criteria:
- Pregnancy
- Heart disease including congestive heart failure or a pacemaker
- Breathing disorder (emphysema or chronic obstructive breathing disorder)
- Neurological disorder such as Parkinson's Disease
- Restless leg syndrome or Periodic limb movement
- Allergies to metal
- Pre-existing skin conditions where sensor would be attached
Sites / Locations
- University of Michigan
Arms of the Study
Arm 1
Arm Type
Experimental
Arm Label
Zansors® sleep screening device
Arm Description
Zansors device compared to overnight polysomnography
Outcomes
Primary Outcome Measures
Specificity
Specificity of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Secondary Outcome Measures
Sensitivity
Sensitivity of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Positive Predictive Value (PPV)
PPV of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Negative Predictive Value (NPV)
NPV of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Full Information
NCT ID
NCT02814227
First Posted
October 25, 2015
Last Updated
January 25, 2023
Sponsor
University of Michigan
Collaborators
Zansors
1. Study Identification
Unique Protocol Identification Number
NCT02814227
Brief Title
Validation of Sleep Apnea Screening Device
Official Title
Wireless Sensor Patch for Reducing Barriers to In-home Sleep Apnea Screening
Study Type
Interventional
2. Study Status
Record Verification Date
January 2023
Overall Recruitment Status
Completed
Study Start Date
September 2014 (undefined)
Primary Completion Date
December 2015 (Actual)
Study Completion Date
December 2015 (Actual)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
University of Michigan
Collaborators
Zansors
4. Oversight
Data Monitoring Committee
No
5. Study Description
Brief Summary
Obstructive sleep apnea (OSA) is the most common type of sleep apnea. OSA affects an estimated 18-40 million adults and 0.7-3% of all children in the US. The marketplace currently does not have an affordable, easy-to-use, over-the-counter, home-based OSA screening device. An affordable, available, FDA-approved and easy-to-use over-the-counter OSA screening tool would allow greater screening of at-risk individuals, especially in underserved communities with low socioeconomic status, hopefully encouraging a greater proportion of such individuals to seek treatment for their condition. The specific goal of this project is to compare the Zansors® micro sleep sensor screening device against gold-standard polysomnography to establish the device's preliminary validity to screen for OSA accurately.
Detailed Description
Obstructive Sleep Apnea (OSA) results in tiredness, depression and fatigue, and has several associated common comorbidities. It is believed that over 80% of OSA remains undiagnosed. Poorer and disadvantaged communities are at higher risk for sleep disorders, including OSA, even after accounting for other risk factors. The gold-standard method for diagnosing OSA is in-laboratory polysomnography; however, this procedure is generally available only to patients with health insurance due to its high cost. As a result, communities at particularly high risk for OSA are less likely to have this serious sleep disorder appropriately diagnosed and treated.
An affordable, available, FDA-approved and easy-to-use over-the-counter OSA screening tool would allow greater screening of at-risk individuals, especially in underserved communities with low socioeconomic status, hopefully encouraging a greater proportion of such individuals to seek treatment for their condition. Zansors® has developed a bioengineered, semiconductor prototype that measures breathing and movement during sleep. It is a 1.5 x 2.5 x 0.2 inch wireless package using four key technologies:
an adjustable microphone;
a 3-axis accelerometer;
embedded algorithms to measure sleep events; and
acrylic adhesive.
The product will be designed as an FDA-approved over-the-counter device that the patient can wear during sleep and wake up with a color score of red, yellow, or green depending on their sleep apnea status. The features of this product include that it is home-based; easy to use; non-invasive; wireless; disposable; low cost ($20-50); and presents easily understandable test results. However, such a device must first be validated against gold-standard polysomnography.
The specific goal of this project is to compare the Zansors® micro sleep sensor screening device against gold-standard polysomnography to establish the device's preliminary validity to screen for OSA accurately.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Obstructive Sleep Apnea
Keywords
obstructive sleep apnea screening device
7. Study Design
Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
52 (Actual)
8. Arms, Groups, and Interventions
Arm Title
Zansors® sleep screening device
Arm Type
Experimental
Arm Description
Zansors device compared to overnight polysomnography
Intervention Type
Device
Intervention Name(s)
Zansors® sleep screening device
Intervention Description
Intervention is the validation of a sleep apnea screening device against the gold-standard assessment of in-laboratory polysomnography
Primary Outcome Measure Information:
Title
Specificity
Description
Specificity of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Time Frame
8 hours
Secondary Outcome Measure Information:
Title
Sensitivity
Description
Sensitivity of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Time Frame
8 hours
Title
Positive Predictive Value (PPV)
Description
PPV of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Time Frame
8 hours
Title
Negative Predictive Value (NPV)
Description
NPV of Zansors' device to detect apnea events compared to gold-standard polysomnography over an 8-hour period of sleep at 30-second intervals
Time Frame
8 hours
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
Must be over 18 and referred by medical staff for an overnight assessment for suspected sleep apnea
Exclusion Criteria:
Pregnancy
Heart disease including congestive heart failure or a pacemaker
Breathing disorder (emphysema or chronic obstructive breathing disorder)
Neurological disorder such as Parkinson's Disease
Restless leg syndrome or Periodic limb movement
Allergies to metal
Pre-existing skin conditions where sensor would be attached
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
J. Todd Arnedt, PhD
Organizational Affiliation
University of Michigan
Official's Role
Principal Investigator
Facility Information:
Facility Name
University of Michigan
City
Ann Arbor
State/Province
Michigan
ZIP/Postal Code
48109
Country
United States
12. IPD Sharing Statement
Learn more about this trial
Validation of Sleep Apnea Screening Device
We'll reach out to this number within 24 hrs