Correlation Between FlowMet™ and Other Gold Standard Assessments in the Management of Critical Limb Ischemia (CLI)
Primary Purpose
Critical Limb Ischemia
Status
Withdrawn
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
FlowMet
Sponsored by

About this trial
This is an interventional diagnostic trial for Critical Limb Ischemia focused on measuring peripheral arterial disease
Eligibility Criteria
Inclusion Criteria:
- Subject is willing and able to provide an informed consent.
- Subject is willing and able to comply with the study procedures.
- Subject is able to understand the study procedures.
- Subject is being seen at the clinic for the evaluation/treatment of CLI.
- Subject has a Rutherford score of V or VI.
Exclusion Criteria:
- Subject has any medical condition, which in the judgment of the Investigator and/or designee makes the subject a poor candidate for the investigational procedure
- Anyone with a latex allergy
- Must be over 18 years old
- Must not be pregnant during time of study
Sites / Locations
- Cleveland Clinic
Arms of the Study
Arm 1
Arm Type
Experimental
Arm Label
FlowMet device
Arm Description
This is a feasibility study
Outcomes
Primary Outcome Measures
Correlation between FlowMet perfusion values and ankle-brachial index and toe-brachial index in 34 patients who present to outpatient clinic.
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to ABI and TBI. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal values of well established testing. Normal values of ABI (ratios of > 0.9 to < 1.3) and TBI (ratios of > 0.7).
Correlation between FlowMet perfusion values and skin perfusion pressure in 34 patients who present to outpatient clinic.
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to skin perfusion pressure testing. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal values of well established testing. Normal values of SPP are 40-50 mmHg.
Correlation between FlowMet perfusion values and transcutaneous oxygen tension in 34 patients who present to outpatient clinic.
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to transcutaneous oxygen testing. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal values of well established testing. Normal value at the foot is 40-50 mmHg.
Correlation between FlowMet perfusion values and angiography in 34 patients who present to outpatient clinic.
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to angiography. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal angiographic images as defined by severity of stenosis.
Secondary Outcome Measures
Correlation between FlowMet perfusion values and Rutherford score
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to Rutherford score of patient at time of evaluation.
Correlation between FlowMet perfusion values and time of surgical (re)admission
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude to determine timing of surgical (re)admission (days).
Correlation between FlowMet perfusion values and speed of wound healing
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude to determine speed of wound healing (days).
Full Information
1. Study Identification
Unique Protocol Identification Number
NCT03094559
Brief Title
Correlation Between FlowMet™ and Other Gold Standard Assessments in the Management of Critical Limb Ischemia (CLI)
Official Title
Correlation Between FlowMet™, ABI, SPP, TBI, TcPO2 and Angiography
Study Type
Interventional
2. Study Status
Record Verification Date
April 2017
Overall Recruitment Status
Withdrawn
Why Stopped
Investigator leaving institution
Study Start Date
March 7, 2017 (Actual)
Primary Completion Date
January 2018 (Anticipated)
Study Completion Date
June 2018 (Anticipated)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Sponsor
Name of the Sponsor
The Cleveland Clinic
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
Yes
5. Study Description
Brief Summary
The purpose of this study is to demonstrate the utility of the FlowMet™ device in the assessment of perfusion in subjects suspected of or being treated for critical limb ischemia, in comparison to other gold standard technologies, such as ankle brachial pulse index (ABI), skin perfusion pressure (SPP), toe brachial pulse index (TBI), transcutaneous oxygen tension (TcP02), and angiography currently in use in the clinic.
Detailed Description
TRIAL OBJECTIVES
1.0
To demonstrate that the FlowMet is effective for the assessment of tissue perfusion compared to gold standard technologies in patients suspected of or being treated for CLI.
2.0 STUDY DESIGN
This is a single center, cross-sectional study of subjects who are scheduled for CLI assessment with a Rutherford score of V and VI.
The subjects will be evaluated at baseline for determination if they meet the inclusion/exclusion criteria of the protocol. If the subject is eligible to enroll in the trial and they have signed an informed consent form (ICF) and are scheduled for evaluation, they will be enrolled. Subjects will undergo FlowMet, ABI, SPP, TBI, TcPO2, and angiography measurements (if not performed previously). The correlation between the aforementioned standard clinical measurements and FlowMet measurements will be assessed.
As ABI, SPP, TBI, TcPO2, Rutherford score, time of surgical (re)admission, speed of wound healing, and angiography measurements will likely change over a period of one year following initial FlowMet measurements, future measurements up to one year post FlowMet measurement (and performed as a standard of care) may also be correlated to initial FlowMet measurements to assess the whether FlowMet data is correlated to long-term patient outcomes.
3.0 STUDY PROCEDURES
3.1 Randomization, Blinding and Subject Identification
This is a cross-sectional study of subjects being treated for CLI with a Rutherford score of V or VI. No randomization scheme is required.
3.2 Product Equipment (FlowMet)
All parts of the FlowMet were manufactured in facilities with ISO 9001:2008 certification. Instruments are constructed in an ISO 9 cleanroom. All materials intended to be in contact with the patient's skin are constructed of polydimethylsiloxane (silicone) rubber which was sterilized via autoclave prior to device construction. Polydimethylsiloxane is widely used in existing medical devices and has been shown to be safe for both short-term contact, as well as long-term implantation (The AMS Sphincter 800™ Urinary Prosthesis, PMA P000053). Between uses, all polydimethylsiloxane components, in addition to all components which can accidentally contact a patient's skin during use, can be sanitized using alcohol wipes or germicidal wipes, such as alcohoPSI Inc. Sani-Cloths, which are commonly used in hospitals and which are bactericidal, tuberculocidal, and virucidal. The FlowMet has been used on 37 patients without adverse events at the University of California, Irvine in a study of tissue optical properties approved and overseen by the University of California, Irvine IRB. The FlowMet has not been evaluated, cleared or approved for use by the FDA.
3.3 Details of FlowMet usage
Extremity perfusion measurements will be performed using the FlowMet peripheral perfusion device. All extremity blood flow measurements will be collected during room air inhalation.
3.3.1
i. The FlowMet device will be placed on the index finger of the right hand. Signal fidelity will be assured by confirming the presence of a pulse waveform if one is expected and by assuring that an appropriate amount of light is be detected by the FlowMet device. The following light intensity values are acceptable: maximum intensity greater than 25 but less than 255 and average subtracted intensity greater than 20 but less than 175.
ii. Following placement of the FlowMet device, a period of 10 seconds will be allowed to elapse to allow transient changes in perfusion due to device placement to diminish.
iii. FlowMet blood flow data will be recorded for 20 seconds using the FlowMet data acquisition software.
iv. Steps i-iii will be repeated for digits 1 and 2 on the limb with diagnosed or suspected CLI, or on both limbs if both are diagnosed or suspected of having CLI. All data sets will be saved, and blood flow in each digit will be computed as the average of the collected data over the 20 second data collection period. If alternative digits are used for data acquisition due to differences in physiology, this will be recorded. If subject is missing the aforementioned toes, the tester will choose a different set of digits and this change will be recorded.
4.0 STUDY PROCEDURES
4.1 Baseline and Assessment
The Cleveland Clinic's standard of care will be followed for all participating study subjects. The following procedures will be conducted as part of a baseline assessment:
Informed consent
Complete medical history
Demographics
Vital signs (heart rate and blood pressure), height and weight
Record angiographic data for study (if available)
The following procedures are standard of care and will be conducted to assess perfusion:
Ankle brachial pulse index (ABI)
Skin Perfusion Pressure (SPP)
Toe brachial pulse index (TBI)
Transcutaneous oxygen tension (TcP02)
Angiography (if not performed previously)
The following procedures are not standard of care and will be conducted to assess perfusion:
FlowMet measurement on subject's hand and foot/feet
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Critical Limb Ischemia
Keywords
peripheral arterial disease
7. Study Design
Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
0 (Actual)
8. Arms, Groups, and Interventions
Arm Title
FlowMet device
Arm Type
Experimental
Arm Description
This is a feasibility study
Intervention Type
Device
Intervention Name(s)
FlowMet
Intervention Description
The FlowMet device will be placed on the index finger of the right hand. Signal fidelity will be assured by confirming the presence of a pulse waveform if one is expected and by assuring that an appropriate amount of light is be detected by the FlowMet device. Following placement of the FlowMet device, a period of 10 seconds will be allowed to elapse to allow transient changes in perfusion due to device placement to diminish. FlowMet blood flow data will be recorded for 20 seconds using the FlowMet data acquisition software. These steps will be repeated for digits 1 and 2 on the limb with diagnosed or suspected CLI, or on both limbs if both are diagnosed or suspected of having CLI.
Primary Outcome Measure Information:
Title
Correlation between FlowMet perfusion values and ankle-brachial index and toe-brachial index in 34 patients who present to outpatient clinic.
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to ABI and TBI. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal values of well established testing. Normal values of ABI (ratios of > 0.9 to < 1.3) and TBI (ratios of > 0.7).
Time Frame
One year
Title
Correlation between FlowMet perfusion values and skin perfusion pressure in 34 patients who present to outpatient clinic.
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to skin perfusion pressure testing. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal values of well established testing. Normal values of SPP are 40-50 mmHg.
Time Frame
One year
Title
Correlation between FlowMet perfusion values and transcutaneous oxygen tension in 34 patients who present to outpatient clinic.
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to transcutaneous oxygen testing. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal values of well established testing. Normal value at the foot is 40-50 mmHg.
Time Frame
One year
Title
Correlation between FlowMet perfusion values and angiography in 34 patients who present to outpatient clinic.
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to angiography. This is a feasibility study to determine the degree of correlation between this novel device and current standard of care testing to see how the numerical values obtained correlate with both normal and abnormal angiographic images as defined by severity of stenosis.
Time Frame
One year
Secondary Outcome Measure Information:
Title
Correlation between FlowMet perfusion values and Rutherford score
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude as compared to Rutherford score of patient at time of evaluation.
Time Frame
One year
Title
Correlation between FlowMet perfusion values and time of surgical (re)admission
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude to determine timing of surgical (re)admission (days).
Time Frame
One year
Title
Correlation between FlowMet perfusion values and speed of wound healing
Description
Correlation between FlowMet perfusion values evaluating the median systolic, median diastolic, and overall flow and amplitude to determine speed of wound healing (days).
Time Frame
One year
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
Subject is willing and able to provide an informed consent.
Subject is willing and able to comply with the study procedures.
Subject is able to understand the study procedures.
Subject is being seen at the clinic for the evaluation/treatment of CLI.
Subject has a Rutherford score of V or VI.
Exclusion Criteria:
Subject has any medical condition, which in the judgment of the Investigator and/or designee makes the subject a poor candidate for the investigational procedure
Anyone with a latex allergy
Must be over 18 years old
Must not be pregnant during time of study
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Mehdi Shishehbor
Organizational Affiliation
The Cleveland Clinic
Official's Role
Principal Investigator
Facility Information:
Facility Name
Cleveland Clinic
City
Cleveland
State/Province
Ohio
ZIP/Postal Code
44195
Country
United States
12. IPD Sharing Statement
Plan to Share IPD
No
IPD Sharing Plan Description
No plan to share IPD
Citations:
PubMed Identifier
7252245
Citation
Anderson RR, Parrish JA. The optics of human skin. J Invest Dermatol. 1981 Jul;77(1):13-9. doi: 10.1111/1523-1747.ep12479191.
Results Reference
background
PubMed Identifier
6656695
Citation
Wilson BC, Adam G. A Monte Carlo model for the absorption and flux distributions of light in tissue. Med Phys. 1983 Nov-Dec;10(6):824-30. doi: 10.1118/1.595361.
Results Reference
background
PubMed Identifier
9805653
Citation
Got I. [Transcutaneous oxygen pressure (TcPO2): advantages and limitations]. Diabetes Metab. 1998 Sep;24(4):379-84. French.
Results Reference
background
PubMed Identifier
12236021
Citation
Zwicky S, Mahler F, Baumgartner I. Evaluation of clinical tests to assess perfusion in chronic critical limb ischemia. Vasa. 2002 Aug;31(3):173-8. doi: 10.1024/0301-1526.31.3.173.
Results Reference
background
PubMed Identifier
8412384
Citation
Jakobsson A, Nilsson GE. Prediction of sampling depth and photon pathlength in laser Doppler flowmetry. Med Biol Eng Comput. 1993 May;31(3):301-7. doi: 10.1007/BF02458050.
Results Reference
background
PubMed Identifier
20598482
Citation
Sprengers RW, Teraa M, Moll FL, de Wit GA, van der Graaf Y, Verhaar MC; JUVENTAS Study Group; SMART Study Group. Quality of life in patients with no-option critical limb ischemia underlines the need for new effective treatment. J Vasc Surg. 2010 Oct;52(4):843-9, 849.e1. doi: 10.1016/j.jvs.2010.04.057.
Results Reference
background
PubMed Identifier
20117502
Citation
Varu VN, Hogg ME, Kibbe MR. Critical limb ischemia. J Vasc Surg. 2010 Jan;51(1):230-41. doi: 10.1016/j.jvs.2009.08.073.
Results Reference
background
PubMed Identifier
18687757
Citation
Hirsch AT, Hartman L, Town RJ, Virnig BA. National health care costs of peripheral arterial disease in the Medicare population. Vasc Med. 2008 Aug;13(3):209-15. doi: 10.1177/1358863X08089277.
Results Reference
background
PubMed Identifier
17140820
Citation
Norgren L, Hiatt WR, Dormandy JA, Nehler MR, Harris KA, Fowkes FG; TASC II Working Group; Bell K, Caporusso J, Durand-Zaleski I, Komori K, Lammer J, Liapis C, Novo S, Razavi M, Robbs J, Schaper N, Shigematsu H, Sapoval M, White C, White J, Clement D, Creager M, Jaff M, Mohler E 3rd, Rutherford RB, Sheehan P, Sillesen H, Rosenfield K. Inter-Society Consensus for the Management of Peripheral Arterial Disease (TASC II). Eur J Vasc Endovasc Surg. 2007;33 Suppl 1:S1-75. doi: 10.1016/j.ejvs.2006.09.024. Epub 2006 Nov 29. No abstract available.
Results Reference
background
PubMed Identifier
16291068
Citation
Falanga V. Wound healing and its impairment in the diabetic foot. Lancet. 2005 Nov 12;366(9498):1736-43. doi: 10.1016/S0140-6736(05)67700-8.
Results Reference
background
PubMed Identifier
11675083
Citation
Ouriel K. Peripheral arterial disease. Lancet. 2001 Oct 13;358(9289):1257-64. doi: 10.1016/S0140-6736(01)06351-6.
Results Reference
background
PubMed Identifier
12631001
Citation
Cronberg CN, Sjoberg S, Albrechtsson U, Leander P, Lindh M, Norgren L, Danielsson P, Sonesson B, Larsson EM. Peripheral arterial disease. Contrast-enhanced 3D MR angiography of the lower leg and foot compared with conventional angiography. Acta Radiol. 2003 Jan;44(1):59-66.
Results Reference
background
PubMed Identifier
9777971
Citation
Stadelmann WK, Digenis AG, Tobin GR. Impediments to wound healing. Am J Surg. 1998 Aug;176(2A Suppl):39S-47S. doi: 10.1016/s0002-9610(98)00184-6.
Results Reference
background
PubMed Identifier
914099
Citation
Mosely LH, Finseth F. Cigarette smoking: impairment of digital blood flow and wound healing in the hand. Hand. 1977 Jun;9(2):97-101. doi: 10.1016/s0072-968x(77)80001-6.
Results Reference
background
Citation
Prahl, S. A., Keijzer, M., Jacques, S. L. & Welch, A. J., SPIE Proceedings of Dosimetry of Laser Radiation in Medicine and Biology. 102-111 (Proc. SPIE IS 5, 1989).
Results Reference
background
PubMed Identifier
16545667
Citation
Hirsch AT, Haskal ZJ, Hertzer NR, Bakal CW, Creager MA, Halperin JL, Hiratzka LF, Murphy WR, Olin JW, Puschett JB, Rosenfield KA, Sacks D, Stanley JC, Taylor LM Jr, White CJ, White J, White RA, Antman EM, Smith SC Jr, Adams CD, Anderson JL, Faxon DP, Fuster V, Gibbons RJ, Halperin JL, Hiratzka LF, Hunt SA, Jacobs AK, Nishimura R, Ornato JP, Page RL, Riegel B; American Association for Vascular Surgery; Society for Vascular Surgery; Society for Cardiovascular Angiography and Interventions; Society for Vascular Medicine and Biology; Society of Interventional Radiology; ACC/AHA Task Force on Practice Guidelines; American Association of Cardiovascular and Pulmonary Rehabilitation; National Heart, Lung, and Blood Institute; Society for Vascular Nursing; TransAtlantic Inter-Society Consensus; Vascular Disease Foundation. ACC/AHA 2005 guidelines for the management of patients with peripheral arterial disease (lower extremity, renal, mesenteric, and abdominal aortic): executive summary a collaborative report from the American Association for Vascular Surgery/Society for Vascular Surgery, Society for Cardiovascular Angiography and Interventions, Society for Vascular Medicine and Biology, Society of Interventional Radiology, and the ACC/AHA Task Force on Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Peripheral Arterial Disease) endorsed by the American Association of Cardiovascular and Pulmonary Rehabilitation; National Heart, Lung, and Blood Institute; Society for Vascular Nursing; TransAtlantic Inter-Society Consensus; and Vascular Disease Foundation. J Am Coll Cardiol. 2006 Mar 21;47(6):1239-312. doi: 10.1016/j.jacc.2005.10.009. No abstract available.
Results Reference
background