ApneaLink Air Home Sleep Testing (HST) Device Validation Study
Primary Purpose
Sleep-disordered Breathing
Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
ApneaLink Air
Sponsored by
About this trial
This is an interventional diagnostic trial for Sleep-disordered Breathing
Eligibility Criteria
Inclusion Criteria:
- Participant is 18 years of age or older
- Participant is willing to provide informed consent
- Participant is willing to participate in all study related procedures
Exclusion Criteria:
- Unable to cease positive airway pressure (PAP) therapy during PSG (if currently using)
- Requires use of oxygen therapy during sleep
- Diagnosis of uncontrolled clinically relevant sleep disorder (e.g., untreated insomnia or restless leg syndrome)
- Pregnant
- Participant is unsuitable to participate in the study in the opinion of the investigator
Sites / Locations
- University of Florida
- University of Miami
Arms of the Study
Arm 1
Arm Type
Experimental
Arm Label
Participants
Arm Description
ApneaLink Air
Outcomes
Primary Outcome Measures
Diagnostic Sensitivity of Apnealink (AL) Compared With Polysomnography (PSG)
The Respiratory Event Index (REI) provided by the AL will be compared with the REI provided by the PSG test. REI will be scored according to American Academy of Sleep Medicine (AASM) 2012 guidelines.
The REI provided by the AL for each participant is divided by the REI provided by the PSG for the same participant to get the diagnostic sensitivity of the Apnealink.
Secondary Outcome Measures
Diagnostic Sensitivity of Apnealink (AL) Compared With Polysomnography (PSG) to Identify Obstructive Events
Total number of Obstructive events reported by the ApneaLink (AL) compared with total number of Obstructive Events reported by polysomnography (PSG). The Obstructive events of the AL are divided by the Obstructive events of the PSG to get the diagnostic sensitivity. This is then compared with the threshold of 0.825
Diagnostic Sensitivity of the AL to Determine Central Events Compared With PSG
Total number of Central events reported by the ApneaLink (AL) compared with total number of Central Events reported by polysomnography (PSG). The Central events of the AL are divided by the Central events of the PSG to get the diagnostic sensitivity. This is then compared with the threshold of 0.825
Positive Likelihood Ratio of Apnealink Air
The positive likelihood ratio (the estimated sensitivity divided by 1 minus the estimated specificity) was calculated along with the associated one-sided 97.5% confidence interval. The positive likelihood ratio was compared with the threshold of 5 (AASM criteria for out-of-center testing devices12) using a Wald test.
Full Information
1. Study Identification
Unique Protocol Identification Number
NCT03470493
Brief Title
ApneaLink Air Home Sleep Testing (HST) Device Validation Study
Official Title
Comparison of the ApneaLink Air Home Sleep Testing Device to Polysomnography
Study Type
Interventional
2. Study Status
Record Verification Date
April 2021
Overall Recruitment Status
Completed
Study Start Date
April 11, 2018 (Actual)
Primary Completion Date
March 6, 2019 (Actual)
Study Completion Date
July 31, 2019 (Actual)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Sponsor
Name of the Sponsor
ResMed
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
Yes
Product Manufactured in and Exported from the U.S.
No
Data Monitoring Committee
No
5. Study Description
Brief Summary
This study compares the diagnostic capability of a home sleep apnea testing device to polysomnography.
Detailed Description
The current gold standard for sleep disordered breathing (SDB) diagnosis is in-laboratory polysomnography (PSG). A barrier of acceptance of Home Sleep Apnea Testing (HSAT) devices as a diagnostic test is their inability to accurately measure total sleep time (TST). A novel algorithm developed by ResMed, Ltd. allows the AL device to accurately calculate TST, however, this algorithm has not yet been validated.
The ApneaLink Air (AL) device is a type III HSAT device. The device is capable of recording up to four channels of data including: flow and snore via a nasal cannula attached to a pressure transducer, a respiratory effort belt, a pulse oximeter to measure pulse and oxygen saturation, and an actigraphy monitor to measure TST along with flow. The AL device has been validated against PSG for AHI, and Cheyne-Stoke respiration detection . Further validation of the effort belt is necessary to determine the accuracy of the AL ability to differentiate between obstructive and central apneic events.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Sleep-disordered Breathing
7. Study Design
Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
130 (Actual)
8. Arms, Groups, and Interventions
Arm Title
Participants
Arm Type
Experimental
Arm Description
ApneaLink Air
Intervention Type
Diagnostic Test
Intervention Name(s)
ApneaLink Air
Intervention Description
ApneaLink Air to be used on each participant undergoing PSG
Primary Outcome Measure Information:
Title
Diagnostic Sensitivity of Apnealink (AL) Compared With Polysomnography (PSG)
Description
The Respiratory Event Index (REI) provided by the AL will be compared with the REI provided by the PSG test. REI will be scored according to American Academy of Sleep Medicine (AASM) 2012 guidelines.
The REI provided by the AL for each participant is divided by the REI provided by the PSG for the same participant to get the diagnostic sensitivity of the Apnealink.
Time Frame
Day 1
Secondary Outcome Measure Information:
Title
Diagnostic Sensitivity of Apnealink (AL) Compared With Polysomnography (PSG) to Identify Obstructive Events
Description
Total number of Obstructive events reported by the ApneaLink (AL) compared with total number of Obstructive Events reported by polysomnography (PSG). The Obstructive events of the AL are divided by the Obstructive events of the PSG to get the diagnostic sensitivity. This is then compared with the threshold of 0.825
Time Frame
Day 1
Title
Diagnostic Sensitivity of the AL to Determine Central Events Compared With PSG
Description
Total number of Central events reported by the ApneaLink (AL) compared with total number of Central Events reported by polysomnography (PSG). The Central events of the AL are divided by the Central events of the PSG to get the diagnostic sensitivity. This is then compared with the threshold of 0.825
Time Frame
Day 1
Title
Positive Likelihood Ratio of Apnealink Air
Description
The positive likelihood ratio (the estimated sensitivity divided by 1 minus the estimated specificity) was calculated along with the associated one-sided 97.5% confidence interval. The positive likelihood ratio was compared with the threshold of 5 (AASM criteria for out-of-center testing devices12) using a Wald test.
Time Frame
Day 1
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
Participant is 18 years of age or older
Participant is willing to provide informed consent
Participant is willing to participate in all study related procedures
Exclusion Criteria:
Unable to cease positive airway pressure (PAP) therapy during PSG (if currently using)
Requires use of oxygen therapy during sleep
Diagnosis of uncontrolled clinically relevant sleep disorder (e.g., untreated insomnia or restless leg syndrome)
Pregnant
Participant is unsuitable to participate in the study in the opinion of the investigator
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Richard Berry, MD
Organizational Affiliation
University of Florida
Official's Role
Principal Investigator
Facility Information:
Facility Name
University of Florida
City
Gainesville
State/Province
Florida
ZIP/Postal Code
32611
Country
United States
Facility Name
University of Miami
City
Miami
State/Province
Florida
ZIP/Postal Code
33136
Country
United States
12. IPD Sharing Statement
Plan to Share IPD
No
Learn more about this trial
ApneaLink Air Home Sleep Testing (HST) Device Validation Study