Pilot Clinical Study on a Low-power Electromagnetic Wave Breast Imaging Device for Cancer Screening Purposes.

Primary Purpose

Breast Cancer, Breast Cyst

Status
Completed
Phase
Not Applicable
Locations
Ireland
Study Type
Interventional
Intervention
Low-power microwave breast imaging system
Sponsored by
MVG Industries SAS
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for Breast Cancer

Eligibility Criteria

18 Years - undefined (Adult, Older Adult)FemaleDoes not accept healthy volunteers

Inclusion Criteria:

  1. Subjects must be able and willing to give written informed consent and to comply with the requirements of this study protocol
  2. Subject must have attended the Symptomatic Breast Unit with a palpable breast lump
  3. Subjects must have had a mammogram in the clinical assessment period (≤ 6 weeks before the microwave breast investigation)
  4. Subjects must be able to comfortably lie reasonably still in a prone position for up to 15 minutes
  5. Subjects with bra size larger than 32B and cup size larger or equal to B.
  6. Subjects whose breast size is adapted to the cylindrical container of the MBI system with sufficient margin to allow the presence of transition liquid around the breast. Final decision to be taken by the physicians based on their judgement.

Exclusion Criteria:

  1. Subjects unable to provide written informed consent
  2. Subjects who are pregnant or breast-feeding
  3. Subjects who have had previous surgery to the breast
  4. Subjects who have previously received chemotherapy or radiotherapy to the breast
  5. Subjects who have had a breast biopsy less than two weeks prior to imaging
  6. Subjects with any active or metallic implant (e.g. cardiac pacemaker, stents, internal cardiac defibrillator, cardiac resynchronisation device, nerve stimulator…), or subjects bearing any non-removable metallic object (e.g. piercing) on their torso
  7. Post-biopsy patients whose breast tissue is not healed sufficiently for the imaging procedure, in the opinion of the investigator
  8. Patients who have had or plan to have a breast cyst aspiration before MBI.
  9. Subjects with significant co-morbidities which, in the opinion of the investigator, may influence the result of the study
  10. Subjects with prior or concurrent malignancy
  11. Subjects under the age of 18 years old
  12. Subjects with evidence of inflammation and/or erythema of the breast as well as any subjects who have a break in the skin which would be in contact with the coupling fluid
  13. Subjects who would be unsuitable for an MBI scan, unlikely to attend a follow up visit, or would otherwise be unsuitable for such an investigation, in the opinion of the Investigator

Sites / Locations

  • HRB Clinical Research Facility Galway

Arms of the Study

Arm 1

Arm 2

Arm 3

Arm Type

Experimental

Experimental

Experimental

Arm Label

Pre-diagnosed breast cancer

Pre-diagnosed breast cyst

Pre-diagnosed benign lesion

Arm Description

Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation

Low-power microwave breast imaging system. No prior biopsy

Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation

Outcomes

Primary Outcome Measures

Number of Serious Adverse Events, Directly Related to Normal Functioning of the Device
To provide early safety information for the proposed investigational medical imaging device.

Secondary Outcome Measures

Full Information

First Posted
March 12, 2018
Last Updated
May 11, 2021
Sponsor
MVG Industries SAS
Collaborators
HRB Clinical Research Facility, Galway (CRFG)
search

1. Study Identification

Unique Protocol Identification Number
NCT03475992
Brief Title
Pilot Clinical Study on a Low-power Electromagnetic Wave Breast Imaging Device for Cancer Screening Purposes.
Official Title
Pilot Clinical Evaluation of a Microwave Imaging System for Early Breast Cancer Detection
Study Type
Interventional

2. Study Status

Record Verification Date
May 2021
Overall Recruitment Status
Completed
Study Start Date
October 10, 2018 (Actual)
Primary Completion Date
May 7, 2020 (Actual)
Study Completion Date
May 7, 2020 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
MVG Industries SAS
Collaborators
HRB Clinical Research Facility, Galway (CRFG)

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
The trialed investigational medical imaging device is a low-power microwave breast imaging system for cancer screening purposes. It is an active device which uses non-ionizing radiation. Microwave imaging is an emerging imaging modality for the early detection of the breast cancer. The physical basis of microwave imaging is the dielectric contrast between healthy and cancerous breast tissues at microwave frequencies. Microwave imaging can potentially be used for monitoring neoadjuvant chemotherapy treatment, breast health monitoring, and for routine screening and diagnosis of the breast cancer at the early-stage. The non-invasive and the non-ionizing characteristics of microwaves should allow for frequent scans of the breast using microwave imaging, unlike X-ray mammography. In addition to safety, microwave imaging does not require uncomfortable breast compression and it is potentially a lower-cost modality. This is a first-in-human clinical test of the investigational device, which has been so far tested only with experimental phantoms modelling the human female breast. The clinical data that will be collected in the context of this study is intended to provide early safety information for the investigational medical imaging device. In addition, this exploratory data will guide the refinement of the device hardware and the imaging algorithm design, before decision to proceed (or not) with further clinical tests. Furthermore, this study will be used to guide sample size calculation for a subsequent study designed to evaluate efficacy should that appear warranted once this study is completed.
Detailed Description
This is a single-site early-phase pilot clinical study, taking place at Galway University Hospital/HRB Clinical Research Facility, Galway. Patients will have a conventional history and breast examinations (Mammogram and / or Ultrasound and Clinical assessment) performed by the physician, as per normal practice in the Symptomatic Breast Unit of University Hospital Galway. If patients are interested in the study, provide written informed consent and are deemed eligible, they will attend the Clinical Research Facility Galway for a Microwave Breast Imaging (MBI) scan. The data from the scan will be collected and stored securely. Patients will be followed up 7-21 days after the microwave breast investigation or before surgery, whichever comes first. Patients will be assessed for their experience of the microwave breast investigation, and for any adverse events. The investigational medical imaging device consists of two subsystems, both performing a non-invasive examination: the Microwave Breast Imaging subsystem (MBI subsystem); the Optical Breast Contour Detection subsystem (OBCD subsystem); The MBI subsystem is an active device which uses non-ionizing radiation. It illuminates the breast with low-power electromagnetic waves in the microwave frequency spectrum, which penetrate the breast under examination. The subsystem collects the scattered electromagnetic waves and recovers useful information on the breast tissue consistency, given the dielectric contrast of these tissues. Multi-static radar detection technology is employed in the MBI subsystem for breast image formation.The well-established Microwave Vision Group (MVG) technology for fast antenna measurement, using multiple sensors in a vertical arch configuration, has been transposed to a horizontal arch of sensors. In addition, vertical translation of the horizontal arch has been enabled, such that 3D multi-static short-range radar imaging is possible. The sensors are in contact with a cylindrical container filled with a liquid; the so-called coupling liquid has been designed to have electromagnetic (EM) properties appropriately selected such that the EM wave penetration in the breast is maximized. During the MBI scan, the patient is lying in a face down position on a special bed, integrated with the MBI subsystem. The breast under examination is immersed in the coupling liquid, through a dedicated circular opening of the bed. The breast is then scanned. This vertical scan takes approximately 10-15 minutes to complete, depending on the size of each breast. In order to compute the required volume of coupling liquid, such that the container of the MBI subsystem is optimally filled after immersion of the breast, a simple process for optical assessment of the total volume of the breast takes place just before starting the MBI scan. The Optical Breast Contour Detection (OBCD) subsystem serves to provide the total volume of the breast, and also its external contour, as a priori information to the MBI subsystem. The OBCD subsystem consists of a 3D optical camera placed below the examination table, at a distance of several tens of centimetres below the breast. An azimuthal scan of the 3D camera permits to reconstruct the external surface of the breast. In order for the breast contour to be a useful a priori information for the MBI subsystem, it is important that the patient is lying in the same face down position during both the MBI and OBCD scans. Thus, an identical examination table, as the one integrated with the MBI subsystem, is also integrated with the OBCD subsystem. During the OBCD scan, the patient is lying on the examination table, with her breast under examination inserted in the circular opening of the examination table. For this scan, there is no coupling liquid; the breast is in the air, hanging below the examination table. Both the breast bearing the palpable lump and the contralateral breast will be scanned. The contralateral breast scan will serve as control in the data analysis. The MBI scan duration of each breast should not exceed 15 minutes. The recorded data will be checked and in the case of non-optimal positioning of the patient, the scan may be repeated. It is expected that the duration of the microwave breast investigation (including OBCD scan of both breasts for breast volume assessment, preparation of the MBI subsystem, MBI scan of both breasts, and data verification) should not exceed 1 hour. If the Data Quality Check suggests repetition of the MBI scan, and upon agreement by the patient, the total duration of the microwave breast investigation will be extended. The written radiology reports from conventional imaging will be acquired and used to evaluate the performance of the MBI subsystem in terms of detecting and estimating the size and consistency of the breast lump. MBI data analysis will be performed off-line, at MVG premises (France), and is expected to be completed few months after the end-date of data collection. Following the assessment of the MBI results by the medical experts involved in the study, a decision will be made regarding the potential of this emerging imaging modality and the interest in proceeding with clinical studies involving larger sample sizes. Based on these results, MVG will decide and plan actions for refinement of the system, towards an upgraded version to be used for potential future tests with a more significant impact.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Breast Cancer, Breast Cyst

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Parallel Assignment
Model Description
The study is performed on 3 groups of patients: Group 1: 15 patients with pre-diagnosed breast cancer (core needle biopsy performed ≥14 days before the microwave breast investigation). Group 2: 10 patients with breast cyst. No prior biopsy Group 3: 5 patients with pre-diagnosed benign lesion (core needle biopsy performed ≥14 days before the microwave breast investigation).
Masking
None (Open Label)
Allocation
Non-Randomized
Enrollment
25 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Pre-diagnosed breast cancer
Arm Type
Experimental
Arm Description
Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation
Arm Title
Pre-diagnosed breast cyst
Arm Type
Experimental
Arm Description
Low-power microwave breast imaging system. No prior biopsy
Arm Title
Pre-diagnosed benign lesion
Arm Type
Experimental
Arm Description
Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation
Intervention Type
Device
Intervention Name(s)
Low-power microwave breast imaging system
Intervention Description
Investigate the capacity of microwave imaging to detect and characterise diagnosed palpable breast lump
Primary Outcome Measure Information:
Title
Number of Serious Adverse Events, Directly Related to Normal Functioning of the Device
Description
To provide early safety information for the proposed investigational medical imaging device.
Time Frame
Through study completion, up to 21 days
Other Pre-specified Outcome Measures:
Title
Capacity of the Low-power Microwave Breast Imaging (MBI) System to Detect Pre-diagnosed Palpable Breast Lesions. Detection Rate for Malignant and Benign Breast Lesions.
Description
Percentage of total number of biopsy-confirmed malignant breast lesions being detected with MBI; Percentage of total number of pre-diagnosed breast cysts being detected with MBI; Percentage of total number of biopsy-confirmed benign breast lesions being detected with MBI.
Time Frame
Duration of study - 19 months

10. Eligibility

Sex
Female
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Subjects must be able and willing to give written informed consent and to comply with the requirements of this study protocol Subject must have attended the Symptomatic Breast Unit with a palpable breast lump Subjects must have had a mammogram in the clinical assessment period (≤ 6 weeks before the microwave breast investigation) Subjects must be able to comfortably lie reasonably still in a prone position for up to 15 minutes Subjects with bra size larger than 32B and cup size larger or equal to B. Subjects whose breast size is adapted to the cylindrical container of the MBI system with sufficient margin to allow the presence of transition liquid around the breast. Final decision to be taken by the physicians based on their judgement. Exclusion Criteria: Subjects unable to provide written informed consent Subjects who are pregnant or breast-feeding Subjects who have had previous surgery to the breast Subjects who have previously received chemotherapy or radiotherapy to the breast Subjects who have had a breast biopsy less than two weeks prior to imaging Subjects with any active or metallic implant (e.g. cardiac pacemaker, stents, internal cardiac defibrillator, cardiac resynchronisation device, nerve stimulator…), or subjects bearing any non-removable metallic object (e.g. piercing) on their torso Post-biopsy patients whose breast tissue is not healed sufficiently for the imaging procedure, in the opinion of the investigator Patients who have had or plan to have a breast cyst aspiration before MBI. Subjects with significant co-morbidities which, in the opinion of the investigator, may influence the result of the study Subjects with prior or concurrent malignancy Subjects under the age of 18 years old Subjects with evidence of inflammation and/or erythema of the breast as well as any subjects who have a break in the skin which would be in contact with the coupling fluid Subjects who would be unsuitable for an MBI scan, unlikely to attend a follow up visit, or would otherwise be unsuitable for such an investigation, in the opinion of the Investigator
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Michael Kerin, Prof.
Organizational Affiliation
Galway University Hospital
Official's Role
Principal Investigator
Facility Information:
Facility Name
HRB Clinical Research Facility Galway
City
Galway
Country
Ireland

12. IPD Sharing Statement

Plan to Share IPD
No
Links:
URL
http://www.mvg-world.com/en
Description
The manufacturer of the trialed investigational medical device and sponsor of the clinical investigation
URL
http://www.nuigalway.ie/hrb_crfg/
Description
The investigational site
URL
http://www.nuigalway.ie/our-research/people/michaelkerin/
Description
The principal investigator

Learn more about this trial

Pilot Clinical Study on a Low-power Electromagnetic Wave Breast Imaging Device for Cancer Screening Purposes.

We'll reach out to this number within 24 hrs