Validation of Diagnostics to Identify Glucose-6-phosphate Dehydrogenase Activity in the US

Primary Purpose

G6PD Deficiency

Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
SD Biosensor G6PD Analyzer
Pointe Scientific Test Kit
HemoCue System
Sponsored by
PATH
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for G6PD Deficiency focused on measuring G6PD deficiency

Eligibility Criteria

18 Years - 65 Years (Adult, Older Adult)All SexesAccepts Healthy Volunteers

Inclusion Criteria:

-Willingness to provide consent

Exclusion Criteria:

-Blood transfusion in the past 90 days by self-report

Sites / Locations

  • Fred Hutchison Prevention Center

Arms of the Study

Arm 1

Arm Type

Other

Arm Label

G6PD Diagnostic Testing

Arm Description

Participants provided whole blood samples as well as fingerstick capillary blood samples. At the clinic site lab, study staff conducted the SD Biosensor point-of-care G6PD test and the point-of-care HemoCue Hb test on both finger stick blood and whole blood samples. At the reference laboratory, G6PD activity was measured from whole blood samples using the Pointe Scientific G6PD reference assay and hemoglobin was measured using a hematology analyzer.

Outcomes

Primary Outcome Measures

Sensitivity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals
For the purposes of this study, an individual was considered G6PD deficient if they tested positive in the Pointe Scientific assay. A true positive was defined as a participant with ≤ 30% of normal G6PD activity in circulating venous blood determined by the Pointe Scientific test. Diagnostic sensitivity of the SD Biosensor G6PD test is defined as the percentage of participants who tested positive for G6PD deficiency on the reference test who were identified by the test assay as positive, calculated as: Number of participants who were both test and true positive / (Number of participants who were test and true positive + Number of participants who were test negative but true positive [ie false negative]) * 100%. Sensitivity of the SD Biosensor POC G6PD test was calculated from both venous and capillary blood samples.
Sensitivity of SD Biosensor POC G6PD Test for Identifying Women With Intermediate G6PD Activity
To investigate the performance of the SD Biosensor G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, assay sensitivity was determined at a G6PD activity threshold of 70%. A true positive for intermediate G6PD activity in females was defined as G6PD activity between 30 and 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Sensitivity of the SD Biosensor POC G6PD test is the percentage of women who tested positive for intermediate G6PD activity on the reference test who were identified by the test assay as positive, calculated as: Number of women who were both test and true positive / (Number of women who were test and true positive + Number of women who were test negative but true positive [ie false negative]) * 100%. Sensitivity of the SD Biosensor G6PD test for identifying females with intermediate G6PD activity was calculated from both venous and capillary blood samples.
Specificity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals
For the purposes of this study, an individual was considered G6PD deficient if they tested positive by the Pointe Scientific assay. A true negative was defined as > 30% of normal G6PD activity in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative [ie false positive]) * 100%. Specificity of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.
Specificity of SD Biosensor G6PD Test for Identifying Women With Intermediate G6PD Activity
To investigate the performance of the SD Biosensor G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, the specificity was determined at a G6PD activity threshold of 70%. A true negative for intermediate G6PD activity in females was defined as G6PD activity > 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative [ie false positive]) * 100%. Specificity of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.

Secondary Outcome Measures

Accuracy Between the SD Biosensor POC G6PD Test Assay and the Pointe Scientific Test Kit
Accuracy between the SD Biosensor POC G6PD test assay and the Pointe Scientific G6PD reference assay was calculated as the percent agreement between both G6PD tests, ie, the percentage of participants with the same diagnosis according to both tests (either deficient, intermediate, or normal G6PD levels). Accuracy of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.
Accuracy Between the SD Biosensor POC G6PD Test Measure of Hemoglobin and the Reference Hemoglobin Test
Accuracy between the SD Biosensor POC G6PD test measure of hemoglobin and the hemoglobin reference assay was calculated as the percentage agreement between both tests, ie, the percentage of participants with the same diagnosis according to both tests (either non/mild anemia, moderate anemia or severe anemia).
Median G6PD Values Measured by the SD Biosensor G6PD Test for Venous and Capillary Blood Samples

Full Information

First Posted
May 21, 2019
Last Updated
April 21, 2022
Sponsor
PATH
Collaborators
Fred Hutchinson Cancer Center
search

1. Study Identification

Unique Protocol Identification Number
NCT04010695
Brief Title
Validation of Diagnostics to Identify Glucose-6-phosphate Dehydrogenase Activity in the US
Official Title
Validation of Diagnostics to Identify Glucose-6-phosphate Dehydrogenase Activity in the US
Study Type
Interventional

2. Study Status

Record Verification Date
June 2021
Overall Recruitment Status
Completed
Study Start Date
May 13, 2019 (Actual)
Primary Completion Date
September 19, 2019 (Actual)
Study Completion Date
September 19, 2019 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
PATH
Collaborators
Fred Hutchinson Cancer Center

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
Yes
Device Product Not Approved or Cleared by U.S. FDA
Yes
Product Manufactured in and Exported from the U.S.
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
The primary objective of this study is to assess the accuracy of the SD Biosensor STANDARD G6PD Analyzer in measuring G6PD activity when used by trained health care workers.
Detailed Description
This is a cross-sectional diagnostic accuracy study with up to 250 participants with a goal of obtaining 20 deficient and 20 intermediate samples. The clinic will recruit and consent study participants. Clinic staff will draw 3 whole blood samples and obtain finger stick capillary blood. G6PD activity is reported in terms of grams of hemoglobin (Hb), hence the hemoglobin concentration must be measured. Clinic staff will perform the investigational Standard Diagnostics (SD) Biosensor point-of-care (POC) test for glucose-6-phosphate dehydrogenase (G6PD) deficiency and a HemoCue® hemoglobin test on finger stick capillary blood and on the venous blood samples. Another venous blood sample will be sent to a clinical laboratory improvement amendments (CLIA)-certified laboratory for reference testing by the gold standard assays: G6PD measurement by spectrophotometry using the Pointe Scientific G6PD reference assay hemoglobin measurement by a hematology analyzer Individuals identified as G6PD deficient with the reference test will be notified of their results by the clinic and referred to their physician for follow-up.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
G6PD Deficiency
Keywords
G6PD deficiency

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
250 (Actual)

8. Arms, Groups, and Interventions

Arm Title
G6PD Diagnostic Testing
Arm Type
Other
Arm Description
Participants provided whole blood samples as well as fingerstick capillary blood samples. At the clinic site lab, study staff conducted the SD Biosensor point-of-care G6PD test and the point-of-care HemoCue Hb test on both finger stick blood and whole blood samples. At the reference laboratory, G6PD activity was measured from whole blood samples using the Pointe Scientific G6PD reference assay and hemoglobin was measured using a hematology analyzer.
Intervention Type
Diagnostic Test
Intervention Name(s)
SD Biosensor G6PD Analyzer
Intervention Description
The SD Biosensor G6PD Analyzer is designed to measure the quantitative determination of total hemoglobin concentration and G6PD enzymatic activity in fresh human whole blood specimens based on reflectometry assays in a point-of-care setting. The test is intended to aid in the identification of people with G6PD deficiency. The test is currently not licensed for use in the US and is considered an investigational product.
Intervention Type
Diagnostic Test
Intervention Name(s)
Pointe Scientific Test Kit
Intervention Description
The Pointe Scientific test kit will serve as the reference assay to assess G6PD activity. Its intended use is for the quantitative, kinetic determination of G6PD in blood at 340 nm.
Intervention Type
Diagnostic Test
Intervention Name(s)
HemoCue System
Intervention Description
The HemoCue hemoglobin (Hb) 201+ system is designed for quantitative point-of-care whole blood hemoglobin determination in primary care using a specially designed analyzer, the HemoCue Hb 201+ Analyzer, and specially designed microcuvettes, the HemoCue Hb 201+Microcuvettes.
Primary Outcome Measure Information:
Title
Sensitivity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals
Description
For the purposes of this study, an individual was considered G6PD deficient if they tested positive in the Pointe Scientific assay. A true positive was defined as a participant with ≤ 30% of normal G6PD activity in circulating venous blood determined by the Pointe Scientific test. Diagnostic sensitivity of the SD Biosensor G6PD test is defined as the percentage of participants who tested positive for G6PD deficiency on the reference test who were identified by the test assay as positive, calculated as: Number of participants who were both test and true positive / (Number of participants who were test and true positive + Number of participants who were test negative but true positive [ie false negative]) * 100%. Sensitivity of the SD Biosensor POC G6PD test was calculated from both venous and capillary blood samples.
Time Frame
All samples were collected on study day 1
Title
Sensitivity of SD Biosensor POC G6PD Test for Identifying Women With Intermediate G6PD Activity
Description
To investigate the performance of the SD Biosensor G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, assay sensitivity was determined at a G6PD activity threshold of 70%. A true positive for intermediate G6PD activity in females was defined as G6PD activity between 30 and 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Sensitivity of the SD Biosensor POC G6PD test is the percentage of women who tested positive for intermediate G6PD activity on the reference test who were identified by the test assay as positive, calculated as: Number of women who were both test and true positive / (Number of women who were test and true positive + Number of women who were test negative but true positive [ie false negative]) * 100%. Sensitivity of the SD Biosensor G6PD test for identifying females with intermediate G6PD activity was calculated from both venous and capillary blood samples.
Time Frame
All samples were collected on study day 1
Title
Specificity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals
Description
For the purposes of this study, an individual was considered G6PD deficient if they tested positive by the Pointe Scientific assay. A true negative was defined as > 30% of normal G6PD activity in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative [ie false positive]) * 100%. Specificity of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.
Time Frame
All samples were collected on study day 1
Title
Specificity of SD Biosensor G6PD Test for Identifying Women With Intermediate G6PD Activity
Description
To investigate the performance of the SD Biosensor G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, the specificity was determined at a G6PD activity threshold of 70%. A true negative for intermediate G6PD activity in females was defined as G6PD activity > 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative [ie false positive]) * 100%. Specificity of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.
Time Frame
All samples were collected on study day 1
Secondary Outcome Measure Information:
Title
Accuracy Between the SD Biosensor POC G6PD Test Assay and the Pointe Scientific Test Kit
Description
Accuracy between the SD Biosensor POC G6PD test assay and the Pointe Scientific G6PD reference assay was calculated as the percent agreement between both G6PD tests, ie, the percentage of participants with the same diagnosis according to both tests (either deficient, intermediate, or normal G6PD levels). Accuracy of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.
Time Frame
All samples were collected on study day 1
Title
Accuracy Between the SD Biosensor POC G6PD Test Measure of Hemoglobin and the Reference Hemoglobin Test
Description
Accuracy between the SD Biosensor POC G6PD test measure of hemoglobin and the hemoglobin reference assay was calculated as the percentage agreement between both tests, ie, the percentage of participants with the same diagnosis according to both tests (either non/mild anemia, moderate anemia or severe anemia).
Time Frame
All samples were collected on study day 1
Title
Median G6PD Values Measured by the SD Biosensor G6PD Test for Venous and Capillary Blood Samples
Time Frame
All samples were collected on study day 1

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Maximum Age & Unit of Time
65 Years
Accepts Healthy Volunteers
Accepts Healthy Volunteers
Eligibility Criteria
Inclusion Criteria: -Willingness to provide consent Exclusion Criteria: -Blood transfusion in the past 90 days by self-report
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
James Kublin, MD, MPH
Organizational Affiliation
Fred Hutchinson Cancer Center
Official's Role
Principal Investigator
Facility Information:
Facility Name
Fred Hutchison Prevention Center
City
Seattle
State/Province
Washington
ZIP/Postal Code
98109
Country
United States

12. IPD Sharing Statement

Plan to Share IPD
No
IPD Sharing Plan Description
Not planning to share IPD to other researchers
Citations:
PubMed Identifier
34543346
Citation
Pal S, Myburgh J, Bansil P, Hann A, Robertson L, Gerth-Guyette E, Ambler G, Bizilj G, Kahn M, Zobrist S, Manis MR, Styke NA, Allan V, Ansbro R, Akingbade T, Bryan A, Murphy SC, Kublin JG, Layton M, Domingo GJ. Reference and point-of-care testing for G6PD deficiency: Blood disorder interference, contrived specimens, and fingerstick equivalence and precision. PLoS One. 2021 Sep 20;16(9):e0257560. doi: 10.1371/journal.pone.0257560. eCollection 2021.
Results Reference
background

Learn more about this trial

Validation of Diagnostics to Identify Glucose-6-phosphate Dehydrogenase Activity in the US