Automated Insulin Delivery in Elderly With Type 1 Diabetes (AIDE T1D) (AIDE T1D)
Primary Purpose
Type 1 Diabetes Mellitus
Status
Active
Phase
Phase 4
Locations
United States
Study Type
Interventional
Intervention
Tandom t:slim X2 with HCL or PLGS
Sponsored by
About this trial
This is an interventional treatment trial for Type 1 Diabetes Mellitus
Eligibility Criteria
Inclusion Criteria:
- Clinical diagnosis of type 1 diabetes
- Age ≥ 65 years old
- T1D Duration of at least 1 year
- HbA1c < 10.0% from point of care or local lab within the past 6 months
- Insulin regimen involves basal/bolus insulin via insulin pump or multiple daily injections
- Most recent GFR ≥ 30 ml/min/m^2 from local lab within the past 6 months
- Willingness to use a rapid acting insulin compatible with the Tandem t:slim X2 pump (currently aspart and lispro; other rapid acting insulins likely to be approved for pump use prior to study initiation such as Fiasp)
- Familiarity with and willingness to use a carbohydrate ratio for meal boluses
- Willing to use study devices and automated insulin delivery features
- Ability to download study devices at home or if not able to download at home willing to come into clinic to bring devices for download of data at visits and as needed for safety
- Participant is independently managing his/her diabetes with respect to insulin administration and glucose monitoring (may include assistance from spouse or other caregiver)
- Participant understands the study protocol, agrees to comply with it and is able to successfully pass the consent understanding assessment with no more than 2 attempts
- Participant comprehends written and spoken English
- At least 240 hours of CGM readings available during the end of run-in assessment
- At least 1.5% of time with CGM glucose levels < 70 mg/dL prior to SAP initiation
- Active prescription for glucagon and willing and able to have glucagon available
Exclusion Criteria:
- Use of PLGS technology or HCL insulin delivery in the past 1 month
- History of 1 or more Diabetic Ketoacidosis episodes in the previous 6 months
- Clinical diagnosis by a primary care provider, neurologist or psychiatrist of dementia, in the investigator's opinion a suspected severe cognitive impairment such that it would preclude ability to understand the study or use devices, or a score of 6 or less out of 15 on the 5 min MoCA (5-min T MoCA Version 2.1) (mild cognitive impairment is not an exclusion)
- A condition, which in the opinion of the investigator or designee, would put the participant or study at risk, including severe vision or hearing impairment and any contraindication to the use of any of the study devices per FDA labeling
- Known adhesive allergy or skin reaction during the run-in pre-randomization phase or previous difficulty with pump and CGM insertions that would preclude participation in the randomized trial
- Concurrent use of any non-insulin glucose-lowering agent other than metformin (including GLP-1 agonists, Symlin, DPP-4 inhibitors, SGLT-2 inhibitors, sulfonylureas)
- Stage 4 or 5 renal disease
- The presence of a significant medical or psychiatric condition or use of a medication that in the judgment of the investigator may affect completion of any aspect of the protocol, or is likely to be associated with life expectancy of <1 year
Sites / Locations
- AdventHealth Diabetes Institute
- Mayo Clinic
- SUNY Upstate
- University of Pennsylvania
Arms of the Study
Arm 1
Arm 2
Arm 3
Arm Type
Active Comparator
Active Comparator
No Intervention
Arm Label
Hybrid Closed Loop Control (HCL)
Predictive Low-Glucose Insulin Suspension (PLGS)
Sensor-Augmented Pump (SAP)
Arm Description
The HCL intervention arm will utilize the Tandom t:slim X2 with Control-IQ Technology and Dexcom G6 CGM
The PLGS intervention arm will utilize the Tandom t:slim X2 with Basal-IQ Technology and Dexcom G6 CGM
The SAP arm will utilize the Tandem t:slim X2 without HCL or PLGS features turned on and Dexcom G6 CGM
Outcomes
Primary Outcome Measures
CGM Measured Time <70 mg/dL
Percentage of sensor glucose values <70 mg/dL
Secondary Outcome Measures
Hypoglycemia
Percentage of sensor glucose values <54 mg/dL
Hypoglycemia
Frequency of CGM-measured hypoglycemic events
Glucose Control
Mean glucose
Glucose Control
Percentage of sensor glucose values 70 to 180 mg/dL
Glucose Control
Coefficient of variation
Hyperglycemia
Percentage of values >180 mg/dL
Hyperglycemia
Percentage of values >250 mg/dL
HbA1c
HbA1c
Hypoglycemia Unawareness
Gold survey
Full Information
NCT ID
NCT04016662
First Posted
July 9, 2019
Last Updated
February 13, 2023
Sponsor
Jaeb Center for Health Research
Collaborators
National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK), DexCom, Inc., Tandem Diabetes Care, Inc., University of Minnesota - Advanced Research and Diagnostic Laboratory, Mayo Clinic, University of Pennsylvania, AdventHealth Diabetes Institute, Washington State University
1. Study Identification
Unique Protocol Identification Number
NCT04016662
Brief Title
Automated Insulin Delivery in Elderly With Type 1 Diabetes (AIDE T1D)
Acronym
AIDE T1D
Official Title
A Randomized Cross-over Trial Evaluating Automated Insulin Delivery Technologies on Hypoglycemia and Quality of Life in Elderly Adults With Type 1 Diabetes
Study Type
Interventional
2. Study Status
Record Verification Date
February 2023
Overall Recruitment Status
Active, not recruiting
Study Start Date
September 28, 2020 (Actual)
Primary Completion Date
August 31, 2023 (Anticipated)
Study Completion Date
December 31, 2023 (Anticipated)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Jaeb Center for Health Research
Collaborators
National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK), DexCom, Inc., Tandem Diabetes Care, Inc., University of Minnesota - Advanced Research and Diagnostic Laboratory, Mayo Clinic, University of Pennsylvania, AdventHealth Diabetes Institute, Washington State University
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
Yes
Data Monitoring Committee
Yes
5. Study Description
Brief Summary
A multi-center, randomized, crossover trial consisting of three sequential 12-week periods, with the HCL feature used during one period, the PLGS feature used during one period and SAP therapy (control) during one period. The crossover trial will be preceded by a run-in phase in which participants will receive training using the study devices (Dexcom G6 and Tandem t:slim X2 pump). After the last crossover period, participants will be given the opportunity to use study devices for an additional 12 weeks to assess preference of system use (PLGS, HCL or SAP) and associated characteristics, durability and safety in a more real-world setting with less frequent study contact.
Detailed Description
Automated insulin delivery (AID) technologies hold the promise of optimizing glycemic control and reducing the burden of diabetes care for patients with Type 1 Diabetes (T1D). However, clinical trials of lower burden AID technologies have not included older adults in sufficient numbers to allow for focused evaluation of efficacy and quality of life (QOL) impacts that may differ from those observed in younger age groups. Most notably, primary endpoints have focused on reducing hyperglycemia, while avoidance of hypoglycemia is of upmost concern for older adults with T1D. T1D Exchange clinic registry data have shown severe hypoglycemia (SH) occurs more commonly in older adults with longstanding T1D than in younger individuals with events occurring just as often with HbA1c levels >8.0% as with HbA1c levels <7.0%. These data do not support the strategy of "raising the HbA1c" as being an effective approach for hypoglycemia prevention in older adults with T1D. In addition to acutely altered mental status, hypoglycemia is associated with an increased risk for falls leading to fractures, car accidents, emergency room (ER) visits, hospitalizations, and mortality resulting in substantial societal costs. The occurrence of hypoglycemia, hypoglycemia unawareness and fear of hypoglycemia have adverse effects on overall QOL of both individuals with T1D and their families.
While continuous glucose monitoring (CGM) technology alone has the potential to be beneficial in reducing hypoglycemia in older patients, our preliminary data from the Wireless Innovations for Seniors with Diabetes Mellitus (WISDM) trial shows a majority of patients still have frequent hypoglycemia even when using CGM. Thus, knowledge of CGM alone may not be sufficient to avoid hypoglycemia in this population. Predictive low-glucose suspend algorithms have particular promise when the primary goal is hypoglycemia avoidance rather than glucose reduction. Whether the added complexity of closed loop systems provides additional glycemic benefit is not known. There is a critical need to determine whether automated insulin delivery can reduce hypoglycemia in the older adult population with T1D.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Type 1 Diabetes Mellitus
7. Study Design
Primary Purpose
Treatment
Study Phase
Phase 4
Interventional Study Model
Crossover Assignment
Model Description
Randomized crossover trial with three 12-week crossover periods (HCL during one period, PLGS during one period, and SAP therapy (control) during one period) preceded by a run-in phase.
Masking
None (Open Label)
Allocation
Randomized
Enrollment
90 (Anticipated)
8. Arms, Groups, and Interventions
Arm Title
Hybrid Closed Loop Control (HCL)
Arm Type
Active Comparator
Arm Description
The HCL intervention arm will utilize the Tandom t:slim X2 with Control-IQ Technology and Dexcom G6 CGM
Arm Title
Predictive Low-Glucose Insulin Suspension (PLGS)
Arm Type
Active Comparator
Arm Description
The PLGS intervention arm will utilize the Tandom t:slim X2 with Basal-IQ Technology and Dexcom G6 CGM
Arm Title
Sensor-Augmented Pump (SAP)
Arm Type
No Intervention
Arm Description
The SAP arm will utilize the Tandem t:slim X2 without HCL or PLGS features turned on and Dexcom G6 CGM
Intervention Type
Device
Intervention Name(s)
Tandom t:slim X2 with HCL or PLGS
Other Intervention Name(s)
Tandom t:slim X2 with Control-IQ Technology, Tandom t:slim X2 with Basal-IQ Technology
Intervention Description
The system components include the t:slim X2 with Control-IQ Technology and the Dexcom CGM G6. The modular control algorithm has a safety supervision module that limits insulin delivery to prevent hypoglycemia at all times. The algorithm gradually decreases hyperglycemia from bedtime to reach a target of 120 mg/dL by waking time. During awake hours, the control algorithm attempts to maintain glucose within a target range (112.5 to 160 mg/dL) with meal time insulin boluses delivered based on usual bolus procedures undertaken by patients on an insulin pump (Hybrid closed loop).
The system components include the t:slim X2 with with Basal-IQ Technology and the Dexcom CGM G6. The PLGS System is able to stop and resume basal insulin delivery automatically in response to predicted or low sensor glucose values, thereby reducing the incidence and duration of hypoglycemic episodes. The pump includes the hypoglycemia minimization strategy that will issue insulin delivery commands.
Primary Outcome Measure Information:
Title
CGM Measured Time <70 mg/dL
Description
Percentage of sensor glucose values <70 mg/dL
Time Frame
12 weeks for each arm of the crossover
Secondary Outcome Measure Information:
Title
Hypoglycemia
Description
Percentage of sensor glucose values <54 mg/dL
Time Frame
12 weeks for each arm of the crossover
Title
Hypoglycemia
Description
Frequency of CGM-measured hypoglycemic events
Time Frame
12 weeks for each arm of the crossover
Title
Glucose Control
Description
Mean glucose
Time Frame
12 weeks for each arm of the crossover
Title
Glucose Control
Description
Percentage of sensor glucose values 70 to 180 mg/dL
Time Frame
12 weeks for each arm of the crossover
Title
Glucose Control
Description
Coefficient of variation
Time Frame
12 weeks for each arm of the crossover
Title
Hyperglycemia
Description
Percentage of values >180 mg/dL
Time Frame
12 weeks for each arm of the crossover
Title
Hyperglycemia
Description
Percentage of values >250 mg/dL
Time Frame
12 weeks for each arm of the crossover
Title
HbA1c
Description
HbA1c
Time Frame
12 weeks for each arm of the crossover
Title
Hypoglycemia Unawareness
Description
Gold survey
Time Frame
12 weeks for each arm of the crossover
Other Pre-specified Outcome Measures:
Title
Patient Reported Questionnaires
Description
Hypoglycemia Fear Survey
Time Frame
12 weeks for each arm of the crossover
Title
Patient Reported Questionnaires
Description
Hypoglycemia Confidence
Time Frame
12 weeks for each arm of the crossover
Title
Patient Reported Questionnaires
Description
Diabetes Distress Scale. 6-point scale from not a problem to a very serious problem. A high total DD score may indicate overall severity.
Time Frame
12 weeks for each arm of the crossover
Title
Patient Reported Questionnaires
Description
AIDE Technology Acceptance is based on the Technology Acceptance Model. 5-point scale from strongly disagree to strongly agree. A high score may indicate acceptance of technology.
Time Frame
12 weeks for each arm of the crossover
Title
Patient Reported Questionnaires
Description
System Usability. 5-point scale from strongly disagree to strongly agree. A high score may indicate ease of system usability.
Time Frame
12 weeks for each arm of the crossover
10. Eligibility
Sex
All
Minimum Age & Unit of Time
65 Years
Accepts Healthy Volunteers
Accepts Healthy Volunteers
Eligibility Criteria
Inclusion Criteria:
Clinical diagnosis of type 1 diabetes
Age ≥ 65 years old
T1D Duration of at least 1 year
HbA1c < 10.0% from point of care or local lab within the past 6 months
Insulin regimen involves basal/bolus insulin via insulin pump or multiple daily injections
Most recent GFR ≥ 30 ml/min/m^2 from local lab within the past 6 months
Willingness to use a rapid acting insulin compatible with the Tandem t:slim X2 pump (currently aspart and lispro; other rapid acting insulins likely to be approved for pump use prior to study initiation such as Fiasp)
Familiarity with and willingness to use a carbohydrate ratio for meal boluses
Willing to use study devices and automated insulin delivery features
Ability to download study devices at home or if not able to download at home willing to come into clinic to bring devices for download of data at visits and as needed for safety
Participant is independently managing his/her diabetes with respect to insulin administration and glucose monitoring (may include assistance from spouse or other caregiver)
Participant understands the study protocol, agrees to comply with it and is able to successfully pass the consent understanding assessment with no more than 2 attempts
Participant comprehends written and spoken English
At least 240 hours of CGM readings available during the end of run-in assessment
At least 1.5% of time with CGM glucose levels < 70 mg/dL prior to SAP initiation
Active prescription for glucagon and willing and able to have glucagon available
Exclusion Criteria:
Use of PLGS technology or HCL insulin delivery in the past 1 month
History of 1 or more Diabetic Ketoacidosis episodes in the previous 6 months
Clinical diagnosis by a primary care provider, neurologist or psychiatrist of dementia, in the investigator's opinion a suspected severe cognitive impairment such that it would preclude ability to understand the study or use devices, or a score of 6 or less out of 15 on the 5 min MoCA (5-min T MoCA Version 2.1) (mild cognitive impairment is not an exclusion)
A condition, which in the opinion of the investigator or designee, would put the participant or study at risk, including severe vision or hearing impairment and any contraindication to the use of any of the study devices per FDA labeling
Known adhesive allergy or skin reaction during the run-in pre-randomization phase or previous difficulty with pump and CGM insertions that would preclude participation in the randomized trial
Concurrent use of any non-insulin glucose-lowering agent other than metformin (including GLP-1 agonists, Symlin, DPP-4 inhibitors, SGLT-2 inhibitors, sulfonylureas)
Stage 4 or 5 renal disease
The presence of a significant medical or psychiatric condition or use of a medication that in the judgment of the investigator may affect completion of any aspect of the protocol, or is likely to be associated with life expectancy of <1 year
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Robert Henderson, MS
Organizational Affiliation
Jaeb Center for Health Research
Official's Role
Principal Investigator
Facility Information:
Facility Name
AdventHealth Diabetes Institute
City
Orlando
State/Province
Florida
ZIP/Postal Code
32803
Country
United States
Facility Name
Mayo Clinic
City
Rochester
State/Province
Minnesota
ZIP/Postal Code
55905
Country
United States
Facility Name
SUNY Upstate
City
Syracuse
State/Province
New York
ZIP/Postal Code
13214
Country
United States
Facility Name
University of Pennsylvania
City
Philadelphia
State/Province
Pennsylvania
ZIP/Postal Code
19104
Country
United States