search
Back to results

Servo Controlled Oxygen Targeting Study (SCO2T)

Primary Purpose

Premature Infant, Oxygen Therapy, Hypoxia

Status
Completed
Phase
Not Applicable
Locations
United Kingdom
Study Type
Interventional
Intervention
Servo control
Sponsored by
University of Edinburgh
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Premature Infant focused on measuring Neonatology, Servo-control, Transcutaneous oxygen, Saturation monitoring

Eligibility Criteria

undefined - 1 Month (Child)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  1. Infants born at less than 29 weeks gestation
  2. Infants greater than 48 hours of age
  3. Infants who are receiving supplementary oxygen
  4. Person with parental responsibility able to give consent

Exclusion Criteria:

  1. Congenital anomalies that would prevent targeting SpO2 to 90-95% (e.g. cardiac defects)
  2. Clinical condition of an infant would impair accurateTcPO2 measurement (e.g. impaired perfusion or requirement of inotropic or vasopressor support)

Sites / Locations

  • The Simpson Centre for Reproductive Health, Royal Infirmary Edinburgh

Arms of the Study

Arm 1

Arm 2

Arm Type

Experimental

No Intervention

Arm Label

Servo control

Manual control

Arm Description

Automated control of oxygen. The oxygen saturation target range will be set to 93%. Automated oxygen control can be over-ridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets.

Standard practice. Oxygen adjustments will be made by clinical/nursing staff to maintain a target oxygen range of 90%-95%.

Outcomes

Primary Outcome Measures

Incidence of hyperoxia and hypoxia on transcutaneous monitoring
To discover the percentage time spent within a TcPO2 range of 50mmHg (6.7kPa) - 80mmHg (10.7kPa) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.

Secondary Outcome Measures

Transcutaneous oxygen variability
To discover the variability in TcPO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Incidence of hyperoxia and hypoxia on saturation monitoring
To discover the percentage time spent within target SpO2 range of 90-95% when infants are targeted using automated (servo) versus manual control.
Saturation variability
To discover the variability in SpO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Fraction of inspired oxygen variability
To discover the variability in FiO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Pooled frequency histogram of TcPO2
To generate a pooled frequency histogram of percentage time at a TcPO2 of below 30mmHg, 30-39.9mmHg, 40-49.9mmHg, 50-59.9mmHg, 60-69.9mmHg, 70-79.9mmHg, and 80mmHg and above for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Pooled frequency histogram of SpO2
To generate a pooled frequency histogram of percentage time at each SpO2 point between 80 - 100% for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Pooled frequency histogram of FiO2
To generate a pooled frequency histogram of the cumulative frequency at a FiO2 of 0.21-0.3, 0.31-0.4, 0.41-0.5, 0.51-0.6, 0.61-0.7, 0.81-0.9 and 0.91-1.0 for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Desaturations
To discover the frequency, duration and depth of desaturations and the area (change in PO2 versus time) above and below the set PO2 threshold for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.

Full Information

First Posted
November 22, 2019
Last Updated
March 11, 2022
Sponsor
University of Edinburgh
Collaborators
NHS Lothian
search

1. Study Identification

Unique Protocol Identification Number
NCT04177992
Brief Title
Servo Controlled Oxygen Targeting Study
Acronym
SCO2T
Official Title
Randomised Cross-over Study of Servo-controlled Oxygen Targeting for Premature Infants
Study Type
Interventional

2. Study Status

Record Verification Date
January 2021
Overall Recruitment Status
Completed
Study Start Date
January 16, 2020 (Actual)
Primary Completion Date
January 6, 2021 (Actual)
Study Completion Date
January 6, 2021 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
University of Edinburgh
Collaborators
NHS Lothian

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
Most premature babies require oxygen therapy. There is uncertainty about what oxygen levels are the best. The oxygen levels in the blood are measured using a monitor called a saturation monitor and the oxygen the baby breathes is adjusted to keep the level in a target range. Although there is evidence that lower oxygen levels maybe harmful, it is not known how high they need to be for maximum benefit. Very high levels are also harmful. Saturation monitors are not very good for checking for high oxygen levels. For this a different kind of monitor, called a transcutaneous monitor, is better. Keeping oxygen levels stable is usually done by nurses adjusting the oxygen levels by hand (manual control). There is also equipment available that can do this automatically (servo control). It is not known which is best. Studies of automated control have shown that infants spend more time within their intended target oxygen saturation range. These have not included measurements of transcutaneous oxygen. The investigators aim to show the transcutaneous oxygen levels as well as the oxygen saturation levels when babies have their oxygen adjusted manually or automatically.
Detailed Description
Presently oxygen is titrated against saturation (SpO2) by manual adjustment. Automated or servo-control systems have been developed that result in tighter control of SpO2 and more time spent in the intended target range. These systems are already in clinical use. Automated systems produce quite large fluctuations in fraction of inspired oxygen (FiO2) in order to keep SpO2 in range. It is possible that this could result in short periods of high or low oxygen tension (PO2) that are undetectable using saturation monitoring. Studies to date have examined the effects of manual and automated (servo) oxygen targeting on SpO2 but not on transcutaneous oxygen tension (TcPO2). There is a need to determine the achieved SpO2 and TcPO2 distributions associated with the use of manual and automated control as a first step in planning trials comparing these approaches over many weeks. When this is measured over a small number of hours it is not anticipated that this would have an influence on clinical outcome. This study is a prospective, single centre, randomised crossover trial of automated (servo) control versus manual oxygen titration. Each infant will act as their own control. Infants born at less than 29 weeks gestation, greater than 48 hour of age and receiving supplementary oxygen will be eligible for inclusion. The study will be undertaken in the Neonatal Unit at the Simpson Centre for Reproductive Health at the Royal Infirmary of Edinburgh. Total study time is 12 hours for each infant. Infants will be randomised to commence on either automated (servo) control or manual mode. SpO2 (range 90-95%) will be continuously monitored as per normal standard of care. A second pulse oximetry probe will be place for servo control input. Additional monitoring will be carried out as shown below: TcPO2 monitoring FiO2 monitoring Heart rate monitoring (used to validate SpO2 readings) Arterial gas sampling (only if conducted by the direct care team as part of the routine care of the infant; no extra blood samples will be taken as part of the study) In manual mode, all oxygen adjustments will be made by clinical/nursing staff. In automated mode, oxygen will be adjusted by the respiratory support device. In automated mode (servo control), oxygen adjustments will be made by one of two devices (depending on the clinical requirements of the baby) - the IntellO2 device (IntellO2, Vapotherm) or Leoni plus CLAC (Closed-Loop Automated oxygen Control) ventilator (Leoni plus, Löwenstein Medical). SpO2 and TcPO2 readings will be downloaded directly from the multiparameter patient monitor. SpO2 will be measured using a Phillips MX500 multiparameter monitor. TcPO2 will be measured using a SenTec Digital Monitoring System with OxiVent sensor. TcPO2 is calculated by dynamic fluorescence quenching which measures oxygen molecules present in the vicinity of a fluorescent dye incorporated within the sensor surface. The sensor is operated at a constant temperature of 43 degrees Celsius. Control of sensor temperature and application duration are designed to meet all applicable standards and this monitoring device is used routinely in many neonatal units.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Premature Infant, Oxygen Therapy, Hypoxia, Hyperoxia, Obstetric Labor, Premature
Keywords
Neonatology, Servo-control, Transcutaneous oxygen, Saturation monitoring

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Crossover Assignment
Model Description
This is a randomised cross-over study of servo-controlled oxygen targeting in premature infants, with infants acting as their own controls.
Masking
None (Open Label)
Masking Description
The study is randomised but not blinded. Infants will be randomised to commence on either automated (servo) control or manual mode, and then cross-over to the alternative range after 6 hours of monitoring (total study time of 12 hours). SpO2 (range 90-95%) will be continuously monitored as per normal standard of care.
Allocation
Randomized
Enrollment
22 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Servo control
Arm Type
Experimental
Arm Description
Automated control of oxygen. The oxygen saturation target range will be set to 93%. Automated oxygen control can be over-ridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets.
Arm Title
Manual control
Arm Type
No Intervention
Arm Description
Standard practice. Oxygen adjustments will be made by clinical/nursing staff to maintain a target oxygen range of 90%-95%.
Intervention Type
Other
Intervention Name(s)
Servo control
Intervention Description
FiO2 adjustments will be made by one of two respiratory devices depending on the clinical requirements and current ventilatory therapy of the infant
Primary Outcome Measure Information:
Title
Incidence of hyperoxia and hypoxia on transcutaneous monitoring
Description
To discover the percentage time spent within a TcPO2 range of 50mmHg (6.7kPa) - 80mmHg (10.7kPa) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Secondary Outcome Measure Information:
Title
Transcutaneous oxygen variability
Description
To discover the variability in TcPO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Title
Incidence of hyperoxia and hypoxia on saturation monitoring
Description
To discover the percentage time spent within target SpO2 range of 90-95% when infants are targeted using automated (servo) versus manual control.
Time Frame
12 hours
Title
Saturation variability
Description
To discover the variability in SpO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Title
Fraction of inspired oxygen variability
Description
To discover the variability in FiO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Title
Pooled frequency histogram of TcPO2
Description
To generate a pooled frequency histogram of percentage time at a TcPO2 of below 30mmHg, 30-39.9mmHg, 40-49.9mmHg, 50-59.9mmHg, 60-69.9mmHg, 70-79.9mmHg, and 80mmHg and above for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Title
Pooled frequency histogram of SpO2
Description
To generate a pooled frequency histogram of percentage time at each SpO2 point between 80 - 100% for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Title
Pooled frequency histogram of FiO2
Description
To generate a pooled frequency histogram of the cumulative frequency at a FiO2 of 0.21-0.3, 0.31-0.4, 0.41-0.5, 0.51-0.6, 0.61-0.7, 0.81-0.9 and 0.91-1.0 for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours
Title
Desaturations
Description
To discover the frequency, duration and depth of desaturations and the area (change in PO2 versus time) above and below the set PO2 threshold for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
Time Frame
12 hours

10. Eligibility

Sex
All
Maximum Age & Unit of Time
1 Month
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Infants born at less than 29 weeks gestation Infants greater than 48 hours of age Infants who are receiving supplementary oxygen Person with parental responsibility able to give consent Exclusion Criteria: Congenital anomalies that would prevent targeting SpO2 to 90-95% (e.g. cardiac defects) Clinical condition of an infant would impair accurateTcPO2 measurement (e.g. impaired perfusion or requirement of inotropic or vasopressor support)
Facility Information:
Facility Name
The Simpson Centre for Reproductive Health, Royal Infirmary Edinburgh
City
Edinburgh
ZIP/Postal Code
EH16 4SA
Country
United Kingdom

12. IPD Sharing Statement

Plan to Share IPD
No

Learn more about this trial

Servo Controlled Oxygen Targeting Study

We'll reach out to this number within 24 hrs