Measurement Algorithm Control and Optimization With Subsequent Performance Evaluation of Sanmina Non-invasive Biosensors During a Standardized Meal Test in Patients With Diabetes Mellitus Type 1 and Type 2

Primary Purpose

Diabetes Mellitus

Status
Completed
Phase
Not Applicable
Locations
Germany
Study Type
Interventional
Intervention
Sanmina Biosensor
Sponsored by
Sciema UG
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional other trial for Diabetes Mellitus

Eligibility Criteria

18 Years - 100 Years (Adult, Older Adult)All SexesAccepts Healthy Volunteers

Inclusion Criteria:

  • Type 1 or Type 2 diabetic subjects or healthy probands
  • Subjects who are able to complete informed consent form (by him/herself or by his/her guardian);
  • 18 years old and above;
  • Anatomically suitable finger in discretion of the investigator

Exclusion Criteria:

  • Does not meet inclusion criteria;
  • Any conditions that may hamper good visual contact between the finger or wrist and sensor, such as raised birthmarks, scars, tattoos;
  • Pregnancy;
  • Nursing mothers;
  • Any skin scratch(es), damage, over dry, long nails on the measured finger;
  • Unsuitable finger with the device might be excluded if recognized during the trial;
  • Medication containing nitrates

Sites / Locations

  • Pfuetzner Science & Health Institute GmbH

Arms of the Study

Arm 1

Arm 2

Arm 3

Arm Type

Experimental

Experimental

Other

Arm Label

Optimization and control 1

Optimization and control 2

Precision and accuracy

Arm Description

The first study part/arm includes 12 patients and will be used for optimization and control of the measurement algorithms of the Sanmina biosensors.

The second study part/arm includes 12 patients and will be used for optimization and control of the measurement algorithms of the Sanmina biosensors.

During the third study part with enrolment of 36 patients, the precision and accuracy of the final Sanmina biosensor algorithm will be demonstrated.

Outcomes

Primary Outcome Measures

Performance and accuracy of glucose levels
performance and accuracy of the Sanmina biosensors (wearable and finger clip) in monitoring glucose levels (mg/dL) during a standardized meal experiment. Measurements (during the meal experiment) of the Sanmina biosensor will be compared to the capillary blood glucose reference method YSI STAT 2300 (glucose, in mg/dL).
Performance and accuracy of pO2
performance and accuracy of the Sanmina biosensors (wearable and finger clip) in monitoring pO2 (%) during a standardized meal experiment. PO2-measurements (during the meal experiment) of the Sanmina biosensor will be compared to the patient monitor (%).
Performance and accuracy of heart rate
performance and accuracy of the Sanmina biosensors (wearable and finger clip) in monitoring heart rate (pulse/min) during a standardized meal experiment. Heart rate measurements (during the meal experiment) of the Sanmina biosensor will be compared to the patient monitor heart rate(pulse/min).

Secondary Outcome Measures

Full Information

First Posted
April 29, 2020
Last Updated
May 4, 2020
Sponsor
Sciema UG
Collaborators
Sanmina
search

1. Study Identification

Unique Protocol Identification Number
NCT04379427
Brief Title
Measurement Algorithm Control and Optimization With Subsequent Performance Evaluation of Sanmina Non-invasive Biosensors During a Standardized Meal Test in Patients With Diabetes Mellitus Type 1 and Type 2
Official Title
Measurement Algorithm Control and Optimization With Subsequent Performance Evaluation of Sanmina Non-invasive Biosensors During a Standardized Meal Test in Patients With Diabetes Mellitus Type 1 and Type 2
Study Type
Interventional

2. Study Status

Record Verification Date
April 2020
Overall Recruitment Status
Completed
Study Start Date
March 25, 2019 (Actual)
Primary Completion Date
November 7, 2019 (Actual)
Study Completion Date
April 1, 2020 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Sciema UG
Collaborators
Sanmina

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
Measurement algorithm control and optimization with subsequent performance evaluation of Sanmina biosensors in monitoring of glucose, heart rate (HR), and SpO2 in patients with type 1 and type 2 diabetes during a standardized meal experiment.
Detailed Description
This study is a combined, open label, prospective, comparative single-center study. The study is separated in 3 consecutive study parts. The first 2 study parts include 12 patients each. In the third study part 36 patients will be enrolled. Study part 1 and 2 will be used for optimization and control of the measurement algorithms of the Sanmina biosensors. During the third study part with enrolment of 36 patients, the precision and accuracy of the final Sanmina biosensor algorithm will be demonstrated. The study conduct for all 60 participants of all 3 study parts will be exactly the same. In all groups of participants, the Sanmina finger clip and wearable non-invasive monitoring biosensor will be individually introduced and assigned. During the study visit at the study site, a standardized meal will be given to the participants. At eleven time points before and after the standardized meal, glucose, heart rate, and pO2 will be measured using the Sanmina finger clip and wearable non-invasive monitoring biosensor. The measurement time points are -30, 0, 15, 30, 45, 60, 75, 90, 120, 150 and 180 min. The measurements will be compared to capillary blood glucose measurements by YSI Stat 2300 and StatStrip (glucose), and parallel measurements of the heart rate and the pO2 using a patient monitor. Further the exact skin colour will be evaluated using a skin colour card.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Diabetes Mellitus

7. Study Design

Primary Purpose
Other
Study Phase
Not Applicable
Interventional Study Model
Sequential Assignment
Model Description
The first 2 study parts include 12 patients each. In the third study part 36 patients will be enrolled. Study part 1 and 2 will be used for optimization and control of the measurement algorithms of the Sanmina biosensors. During the third study part with enrolment of 36 patients, the precision and accuracy of the final Sanmina biosensor algorithm will be demonstrated.
Masking
None (Open Label)
Allocation
Non-Randomized
Enrollment
60 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Optimization and control 1
Arm Type
Experimental
Arm Description
The first study part/arm includes 12 patients and will be used for optimization and control of the measurement algorithms of the Sanmina biosensors.
Arm Title
Optimization and control 2
Arm Type
Experimental
Arm Description
The second study part/arm includes 12 patients and will be used for optimization and control of the measurement algorithms of the Sanmina biosensors.
Arm Title
Precision and accuracy
Arm Type
Other
Arm Description
During the third study part with enrolment of 36 patients, the precision and accuracy of the final Sanmina biosensor algorithm will be demonstrated.
Intervention Type
Device
Intervention Name(s)
Sanmina Biosensor
Intervention Description
non-invasive blood glucose measurement using the Sanmina Biosensor referenced with the invasive reference blood glucose device YSI 2400 Plus.
Primary Outcome Measure Information:
Title
Performance and accuracy of glucose levels
Description
performance and accuracy of the Sanmina biosensors (wearable and finger clip) in monitoring glucose levels (mg/dL) during a standardized meal experiment. Measurements (during the meal experiment) of the Sanmina biosensor will be compared to the capillary blood glucose reference method YSI STAT 2300 (glucose, in mg/dL).
Time Frame
approx. 240 minutes
Title
Performance and accuracy of pO2
Description
performance and accuracy of the Sanmina biosensors (wearable and finger clip) in monitoring pO2 (%) during a standardized meal experiment. PO2-measurements (during the meal experiment) of the Sanmina biosensor will be compared to the patient monitor (%).
Time Frame
approx. 240 minutes
Title
Performance and accuracy of heart rate
Description
performance and accuracy of the Sanmina biosensors (wearable and finger clip) in monitoring heart rate (pulse/min) during a standardized meal experiment. Heart rate measurements (during the meal experiment) of the Sanmina biosensor will be compared to the patient monitor heart rate(pulse/min).
Time Frame
approx. 240 minutes

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Maximum Age & Unit of Time
100 Years
Accepts Healthy Volunteers
Accepts Healthy Volunteers
Eligibility Criteria
Inclusion Criteria: Type 1 or Type 2 diabetic subjects or healthy probands Subjects who are able to complete informed consent form (by him/herself or by his/her guardian); 18 years old and above; Anatomically suitable finger in discretion of the investigator Exclusion Criteria: Does not meet inclusion criteria; Any conditions that may hamper good visual contact between the finger or wrist and sensor, such as raised birthmarks, scars, tattoos; Pregnancy; Nursing mothers; Any skin scratch(es), damage, over dry, long nails on the measured finger; Unsuitable finger with the device might be excluded if recognized during the trial; Medication containing nitrates
Facility Information:
Facility Name
Pfuetzner Science & Health Institute GmbH
City
Mainz
State/Province
Rhineland-Palatinate
ZIP/Postal Code
55128
Country
Germany

12. IPD Sharing Statement

Plan to Share IPD
No

Learn more about this trial

Measurement Algorithm Control and Optimization With Subsequent Performance Evaluation of Sanmina Non-invasive Biosensors During a Standardized Meal Test in Patients With Diabetes Mellitus Type 1 and Type 2