Respiratory Adaptive Computed Tomography: Feasibility Study on Real-Time Gated 4DCT for Lung Cancer Radiotherapy (REACT)
Primary Purpose
Lung Cancer
Status
Not yet recruiting
Phase
Not Applicable
Locations
Australia
Study Type
Interventional
Intervention
Real-time Gated 4DCT
Conventional 4DCT
Sponsored by

About this trial
This is an interventional other trial for Lung Cancer focused on measuring Radiotherapy, 4DCT, Real-time gated 4DCT, Imaging artifacts, Computed tomography
Eligibility Criteria
Inclusion Criteria:
- 18 years or older
- Have the ability to give informed consent
- A diagnosis of lung cancer with an indication for radiation therapy
- Radiation therapy treatment involving the acquisition of a 4DCT scan for treatment planning
Exclusion Criteria:
- Pregnant women
- Patients <18 years
- Patients who in the opinion of the treating physician could not tolerate the extra time on the CT couch for an extra scan
Sites / Locations
- Blacktown Hospital
Arms of the Study
Arm 1
Arm Type
Experimental
Arm Label
Real-Time Gated 4DCT and Conventional 4DCT
Arm Description
During the simulation (radiotherapy planning) session, the participant will undergo both the experimental 'Real-Time Gated 4DCT' and standard 'Conventional 4DCT'.
Outcomes
Primary Outcome Measures
The feasibility of using Real-Time Gated 4DCT instead of Conventional 4DCT for planning radiotherapy for lung cancer.
A change in the anticipated rate of image artifacts in images acquired from Conventional of 60% to 35% or less in images acquired from Real-Time Gated 4DCT.
Secondary Outcome Measures
Patient tolerance to the experimental scan (Real-Time Gated 4DCT)
The radiation therapist conducting the radiation therapy study scan will complete a survey consisting of two yes/no questions after the planning session to assess patient tolerance (1. whether the patient finishes the scan without getting off the couch. 2. If the patient voiced any complaints during the scan and seem uncomfortable.)
clinician confidence in acquiring a Real-Time Gated 4DCT scan
The radiation therapist conducting the radiation therapy study scan will complete an in-house survey of 10 questions after the treatment session to evaluate their experience with REACT software and its usability.
clinician confidence in delineation on a Real-Time Gated 4DCT scan
The radiation oncologist performing the delineation on the radiation therapy planning scan images and study scan images will complete an in-house survey of 10 questions to evaluate his experience with REACT and his confidence with using REACT for delineation.
Effect on Real-Time Gated 4DCT image quality of patient characteristics: breathing type, breathing period, BMI, tumour location, breathing amplitude, breathing regularity, pulmonary function, ECOG status.
Correlation between image quality (the number of artifacts greater than 4mm in the images acquired from Real-Time Gated 4DCT) and participant characteristics known to affect image quality in conventional scans: Breathing period ( seconds) and breathing amplitude will be measured using the Varian real-time position management system; body mass index (weight(kg) x height(m)2); tumour location;, breathing regularity (root mean square error (RMSE) of 6 breaths/min was obtained), pulmonary function, Performance status (Eastern Cooperative Oncology Group (ECOG) performance scale).
Difference in scanning time between Real-Time Gated 4DCT compared with Conventional 4DCT for planning of radiotherapy for lung cancer
Time in minutes from start to end of scan.
Quality of treatment plan using images acquired during Real-Time Gated 4DCT compared with images acquired during Conventional 4DCT
Contouring and planning using the Real-Time Gated 4DCT image set to develop a treatment plan of prescribed radiation therapy. Treatment plans will be assessed using the 3%/3mm and 2%/2mm gamma pass index.
Full Information
NCT ID
NCT05030207
First Posted
June 24, 2021
Last Updated
October 15, 2023
Sponsor
University of Sydney
Collaborators
Western Sydney Local Health District
1. Study Identification
Unique Protocol Identification Number
NCT05030207
Brief Title
Respiratory Adaptive Computed Tomography: Feasibility Study on Real-Time Gated 4DCT for Lung Cancer Radiotherapy
Acronym
REACT
Official Title
Respiratory Adaptive Computed Tomography: A Pilot Feasibility Study on the Use of Real-Time Gated 4DCT for Lung Cancer Radiation Therapy
Study Type
Interventional
2. Study Status
Record Verification Date
October 2023
Overall Recruitment Status
Not yet recruiting
Study Start Date
November 2023 (Anticipated)
Primary Completion Date
November 2024 (Anticipated)
Study Completion Date
November 2025 (Anticipated)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Sponsor
Name of the Sponsor
University of Sydney
Collaborators
Western Sydney Local Health District
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No
5. Study Description
Brief Summary
This study does not involve a therapeutic intervention as standard radiation therapy treatment will be prescribed. This study involves one additional 4DCT scan (i.e. the Real-Time Gated 4DCT scan) acquired immediately before or following the conventional 4DCT scan. This will take place on the day of the patient's treatment simulation, as per the current standard of care. The scanning sequence (i.e. conventional first versus gated first) will be randomised.
The Real-Time Gated 4DCT is anticipated to take longer than the conventional 4DCT scan, due to its gated (beam-pause) nature. However, upper limits for timing will be implemented in the software, and the scan aborted for highly erratic breathing traces that would not benefit from a Real-Time Gated 4DCT scan.
Detailed Description
Four-dimensional computed tomography (4DCT) is considered the standard of care for modern high-precision stereotactic ablative body radiotherapy (SABR) techniques. Clinical 4DCT works by acquiring CT slices and breathing data synchronously as the patient moves through the rotating X-ray imager. The CT slices are sorted into 5-10 'phase images' (mid-exhale, peak-exhale, mid-inhale, peak-inhale, etc.) and stitched together to form a 3D+breathing induced tumour/organ motion or '4D' representation of the breathing anatomy. Current best practice in radiation therapy is for clinicians to perform manual delineation or 'contouring' of lung tumours on one or more 4DCT phase images to ensure proper coverage by the treatment beam despite breathing motion. In the presence of irregular breathing, mismatches arise between CT slices that are in the same breathing phase but imaged at different couch positions. As a result, the images produced suffer from various kinds of discontinuities including truncation, duplication or overlap. Yamamoto found that irregular breathing caused at least one error >4mm in 90% (45 out of 50) abdominal and thoracic patient scans. The main consequence of 4DCT image errors is that they can introduce tumour volume and position uncertainties as large as 30% between different 4D phase bins or different observers.
This can potentially reduce the likelihood of tumour control by up to one-third and lead to unnecessary irradiation of healthy lung tissue, contributing to major dose-limiting side effects like radiation pneumonitis, which is symptomatic in up to 30% of patients and fatal in 2%. The proposed solution is Real-Time Gated 4DCT using prospective regularity gating, which will be implemented for the first time on lung cancer patients.
The Real-Time Gated 4DCT method detects and then pauses the CT beam during irregular breathing events. We will analyse the patient's breathing pattern to prospectively gate acquisition in real-time. Real-time gated 4DCT ensures that the X-ray imager is switched on only when the breathing is regular; the scan gets paused otherwise. The benefits for the patient of gating a 4DCT scan have been demonstrated in previous studies:
a reduction of breathing motion errors by 50%
a reduction of imaging dose. Delineation of tumours and organs in CT scans is the first step in a lung cancer patient's radiation therapy journey. Accurate delineation is pivotal to patient outcomes because errors propagate throughout treatment and can lead to radiation dose missing parts of the tumour or unnecessarily irradiating healthy tissue or organs. Unfortunately, the most important imaging modality used for the delineation of lung cancers, 4DCT, is error prone. Radiation Oncologists (ROs) are faced with an enormous challenge and are required to delineate tumours when 90% of clinical scans suffer severe imaging errors. These errors obfuscate the true tumour position/shape and reduce the chance of tumour control by up to one third.
The main culprit for imaging errors is irregular breathing: i.e. natural breathing variations, which are heightened for lung cancer patients but not accounted for by clinical scanners. A solution to this problem was proposed, which is to gate the CT scanner (i.e. pausing the beam) whenever a breathing condition occurs that is likely to produce an imaging artefact.
Research Question The primary endpoint is that the number of image artifacts using the Real-time gated scan will produce fewer artefacts than the conventional scan.
Secondary endpoints involve examining image quality and the impact on treatment planning with the benefits to the patient being:
Consistent/predictable image quality: Consistent image quality will assist radiation oncologists in identifying tumour and organ interfaces and therefore reduce delineation errors.
Improvements in treatment plans: More accurate delineation will lead to improved treatment plans and therefore improved patient outcomes.
This clinical trial is a first-in-humans pilot feasibility study and the aim is therefore to prove feasibility of Real-Time Gated 4DCT, with the aim of leading to an efficacy trial. As such, patient scans will be acquired across a broad range of patient breathing conditions to optimise the Real-Time Gated technique and to plan a larger, hypothesis-driven phase II clinical trial to follow. Automatic cut-offs for these erratic breathing traces will be built into the Real-Time Gated 4DCT software.
Rationale for Current Study In-silico studies suggest that the Real-Time Gated 4DCT approach, can reduce imaging errors by up to 50% whilst also reducing imaging dose by >20%. It has been estimated that the current rate of artifacts using conventional 4DCT is at least 60% and possibly up to 95%. A reduction to at most 35% (25% of images with fewer artifacts) would be clinically worthwhile.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Lung Cancer
Keywords
Radiotherapy, 4DCT, Real-time gated 4DCT, Imaging artifacts, Computed tomography
7. Study Design
Primary Purpose
Other
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Model Description
Single site, 1 arm unblinded, pilot feasibility clinical trial to investigate the feasibility of acquiring Real-Time Gated 4DCT scans for lung cancer patients undergoing radiation therapy and to determine the incidence of imaging artifacts in comparison to conventional 4DCT
Masking
None (Open Label)
Allocation
N/A
Enrollment
30 (Anticipated)
8. Arms, Groups, and Interventions
Arm Title
Real-Time Gated 4DCT and Conventional 4DCT
Arm Type
Experimental
Arm Description
During the simulation (radiotherapy planning) session, the participant will undergo both the experimental 'Real-Time Gated 4DCT' and standard 'Conventional 4DCT'.
Intervention Type
Diagnostic Test
Intervention Name(s)
Real-time Gated 4DCT
Intervention Description
Real-time gated four-dimensional cone beam computed tomography (4DCT) will be conducted at the simulation session. The Real-Time Gated 4DCT will be conducted while the patient moves through the rotating X-ray imager. This method detects and then pauses the CT beam during irregular breathing events.
Intervention Type
Diagnostic Test
Intervention Name(s)
Conventional 4DCT
Intervention Description
Four-dimensional computed tomography (4DCT), the standard of care for modern high-precision stereotactic ablative body radiotherapy (SABR), will be conducted at the simulation session. It will run continuously while the patient moves through the rotating X-ray imager.
Primary Outcome Measure Information:
Title
The feasibility of using Real-Time Gated 4DCT instead of Conventional 4DCT for planning radiotherapy for lung cancer.
Description
A change in the anticipated rate of image artifacts in images acquired from Conventional of 60% to 35% or less in images acquired from Real-Time Gated 4DCT.
Time Frame
30 minutes
Secondary Outcome Measure Information:
Title
Patient tolerance to the experimental scan (Real-Time Gated 4DCT)
Description
The radiation therapist conducting the radiation therapy study scan will complete a survey consisting of two yes/no questions after the planning session to assess patient tolerance (1. whether the patient finishes the scan without getting off the couch. 2. If the patient voiced any complaints during the scan and seem uncomfortable.)
Time Frame
1 minute at end of planning session
Title
clinician confidence in acquiring a Real-Time Gated 4DCT scan
Description
The radiation therapist conducting the radiation therapy study scan will complete an in-house survey of 10 questions after the treatment session to evaluate their experience with REACT software and its usability.
Time Frame
5 minutes at end of planning session
Title
clinician confidence in delineation on a Real-Time Gated 4DCT scan
Description
The radiation oncologist performing the delineation on the radiation therapy planning scan images and study scan images will complete an in-house survey of 10 questions to evaluate his experience with REACT and his confidence with using REACT for delineation.
Time Frame
5-10 minutes, following delineation of standard care and study scans.
Title
Effect on Real-Time Gated 4DCT image quality of patient characteristics: breathing type, breathing period, BMI, tumour location, breathing amplitude, breathing regularity, pulmonary function, ECOG status.
Description
Correlation between image quality (the number of artifacts greater than 4mm in the images acquired from Real-Time Gated 4DCT) and participant characteristics known to affect image quality in conventional scans: Breathing period ( seconds) and breathing amplitude will be measured using the Varian real-time position management system; body mass index (weight(kg) x height(m)2); tumour location;, breathing regularity (root mean square error (RMSE) of 6 breaths/min was obtained), pulmonary function, Performance status (Eastern Cooperative Oncology Group (ECOG) performance scale).
Time Frame
40 minutes, prior to and during the planning session
Title
Difference in scanning time between Real-Time Gated 4DCT compared with Conventional 4DCT for planning of radiotherapy for lung cancer
Description
Time in minutes from start to end of scan.
Time Frame
30 minutes
Title
Quality of treatment plan using images acquired during Real-Time Gated 4DCT compared with images acquired during Conventional 4DCT
Description
Contouring and planning using the Real-Time Gated 4DCT image set to develop a treatment plan of prescribed radiation therapy. Treatment plans will be assessed using the 3%/3mm and 2%/2mm gamma pass index.
Time Frame
1 week
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
18 years or older
Have the ability to give informed consent
A diagnosis of lung cancer with an indication for radiation therapy
Radiation therapy treatment involving the acquisition of a 4DCT scan for treatment planning
Exclusion Criteria:
Pregnant women
Patients <18 years
Patients who in the opinion of the treating physician could not tolerate the extra time on the CT couch for an extra scan
Central Contact Person:
First Name & Middle Initial & Last Name or Official Title & Degree
Shona J Silvester, MMedSc
Phone
+61286271185
Email
Shona.Silvester@sydney.edu.au
First Name & Middle Initial & Last Name or Official Title & Degree
Natalie A Plant, MHSc
Phone
+61286271133
Email
Natalie.Plant@sydney.edu.au
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Ricky O'Brien, PhD
Organizational Affiliation
University of Sydney
Official's Role
Principal Investigator
Facility Information:
Facility Name
Blacktown Hospital
City
Blacktown
State/Province
New South Wales
ZIP/Postal Code
2148
Country
Australia
12. IPD Sharing Statement
Learn more about this trial
Respiratory Adaptive Computed Tomography: Feasibility Study on Real-Time Gated 4DCT for Lung Cancer Radiotherapy
We'll reach out to this number within 24 hrs