Flow Sensor and Colourimetric Capnometer in Verifying Tracheal Tube Positioning in Term and Preterm Infants (FlowNET)

Primary Purpose

Tracheal Intubation

Status
Recruiting
Phase
Not Applicable
Locations
Italy
Study Type
Interventional
Intervention
confirmation of endotracheal tube positioning
Sponsored by
Azienda Ospedaliera Universitaria Integrata Verona
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for Tracheal Intubation focused on measuring tracheal intubation, tracheal tube positioning, colorimetric capnometer, flow sensor, invasive mechanical ventilation, newborn, preterm infant

Eligibility Criteria

20 Minutes - 3 Months (Child)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

Term and preterm infants requiring tracheal intubation within the neonatal intensive care unit, both in elective and emergency conditions.

Exclusion Criteria:

  • intubation at birth
  • intubation during cardiopulmonary resuscitation
  • absence of informed consent by parents

Sites / Locations

  • Azienda Ospedaliera Universitaria Integrata VeronaRecruiting

Arms of the Study

Arm 1

Arm Type

Other

Arm Label

all patients undergoing tracheal intubation procedure

Arm Description

In current practice, the endotracheal intubation procedure includes the systematic use of the capnometer, which is immediately connected to the tube after its insertion by direct laringoscopy. Confirmation of the correct tube positioning is established by color change of the capnometer from violet to yellow. Then, the colorimetric capnometer is usually removed before connecting the endotracheal tube to the flow-sensor of the ventilator. For the purpose of our study, after the tube insertion the care provider will connect in line both the capnometer and the ventilator flow-sensor to the tracheal tube.

Outcomes

Primary Outcome Measures

Time taken by the flow-sensor to indicate the correct or not correct positioning of the tracheal tube, compared with time taken by the colorimetric capnometer.
To compare the time (expressed in seconds) taken by the flow sensor and the colorimetric capnometer, respectively, to indicate the correct or not correct positioning of the tube in newborns undergoing tracheal intubation procedure during neonatal intensive care unit admission. The aim is to evaluate if the flow sensor of the mechanical ventilator can discriminate with similar or faster times the correct positioning of the tube during the tracheal intubation procedure compared to the colorimetric capnometer.
Number of mechanical breaths necessary to indicate the correct or not correct positioning of the tracheal tube by evaluating the flow-sensor curves and the color change of the colorimetric capnometer.
To compare the minimal number of mechanical breaths provided by the ventilator to indicate the correct or not correct positioning of the tube in newborns undergoing tracheal intubation procedure, either evaluating the flow-sensor curves and the color change of the colorimetric capnometer, respectively. The aim is to evaluate if the flow sensor of the mechanical ventilator can indicate with lower or equal number of mechanical breaths the correct positioning of the tube during tracheal intubation compared to the colorimetric capnometer.

Secondary Outcome Measures

Full Information

First Posted
December 6, 2021
Last Updated
March 7, 2022
Sponsor
Azienda Ospedaliera Universitaria Integrata Verona
search

1. Study Identification

Unique Protocol Identification Number
NCT05162313
Brief Title
Flow Sensor and Colourimetric Capnometer in Verifying Tracheal Tube Positioning in Term and Preterm Infants
Acronym
FlowNET
Official Title
Comparison Between Flow Sensor and Colourimetric Capnometer in Verifying the Correct Positioning of the Endotracheal Tube in Term and Preterm Infants
Study Type
Interventional

2. Study Status

Record Verification Date
March 2022
Overall Recruitment Status
Recruiting
Study Start Date
February 10, 2022 (Actual)
Primary Completion Date
April 2023 (Anticipated)
Study Completion Date
June 2023 (Anticipated)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
Azienda Ospedaliera Universitaria Integrata Verona

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
This is a multicentre clinical trial with medical device. As currently recommended by international neonatal resuscitation guidelines, the most reliable method to verify the correct positioning of the endotracheal tube, in association with clinical signs, is the end-tidal capnometry obtained either by infrared spectroscopy or colorimetric method. The aim of the present study is to evaluate whether the flow sensor of a standard mechanical ventilator can discriminate with similar or faster times the correct positioning of the tube after a tracheal intubation attempt compared to the colorimetric capnometer, in newborns undergoing this procedure during hospitalization in the neonatal intensive care unit. Given the importance of defining a rapid and effective method to prevent possible adverse events of incorrect endotracheal intubation, this study aims to verify whether the ventilator flow sensor can allow a reliable assessment of the correct positioning of the endotracheal tube, with timing and success rates equivalent to or better than the colorimetric capnometer.
Detailed Description
Specifically, the neonatologist performing the procedure will follow the standard procedure for tracheal intubation, which includes using a colourimetric capnometer to confirm the correct tube positioning. However, he/she will remain blind and not aware of the flow sensor signal, while a study investigator will monitor that. The investigators will measure the percentage of responses of correct positioning of the endotracheal tube as well as the number of insufflations and the time elapsed before obtaining a signal of correct intubation by simultaneously comparing both methods and checking any differences based on the characteristics of the sample analyzed (e.g. premature infants versus term infants, elective versus emergency intubations).

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Tracheal Intubation
Keywords
tracheal intubation, tracheal tube positioning, colorimetric capnometer, flow sensor, invasive mechanical ventilation, newborn, preterm infant

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Model Description
Simultaneous assignment, as the two methods to confirm the endotracheal tube position, namely the colorimetric capnometer and the flow sensor signal, will be evaluated simultaneously in each recruited patient.
Masking
None (Open Label)
Masking Description
Participants are newborns. The care provider will be the attending physician performing the tracheal intubation procedure as per local protocols, including the colourimetric capnometry analysis. However, he/she will remain blind and not aware of the flow sensor data until the end of the procedure.
Allocation
N/A
Enrollment
105 (Anticipated)

8. Arms, Groups, and Interventions

Arm Title
all patients undergoing tracheal intubation procedure
Arm Type
Other
Arm Description
In current practice, the endotracheal intubation procedure includes the systematic use of the capnometer, which is immediately connected to the tube after its insertion by direct laringoscopy. Confirmation of the correct tube positioning is established by color change of the capnometer from violet to yellow. Then, the colorimetric capnometer is usually removed before connecting the endotracheal tube to the flow-sensor of the ventilator. For the purpose of our study, after the tube insertion the care provider will connect in line both the capnometer and the ventilator flow-sensor to the tracheal tube.
Intervention Type
Device
Intervention Name(s)
confirmation of endotracheal tube positioning
Intervention Description
After performing the tube insertion by direct laryngoscopy, videolaringoscopy or fiberoptic technique, the attending neonatologist will connect both the colorimetric capnometer and the proximal flow-sensor to the tracheal tube. If the flow sensor is not proximal but incorporated into the mechanical ventilator, the care provider will connect the capnometer only. During the procedure, the information given by the capnometer will be evaluated by the care provider performing the intubation, while the flow sensor information will be monitored only by a study investigator, who will not reveal any details to the care provider. In particular, the yellow color change by the capnometer and the presence of exhalation flow curves on the ventilator screen will be considered as confirmation of correct tracheal tube positioning, respectively.
Primary Outcome Measure Information:
Title
Time taken by the flow-sensor to indicate the correct or not correct positioning of the tracheal tube, compared with time taken by the colorimetric capnometer.
Description
To compare the time (expressed in seconds) taken by the flow sensor and the colorimetric capnometer, respectively, to indicate the correct or not correct positioning of the tube in newborns undergoing tracheal intubation procedure during neonatal intensive care unit admission. The aim is to evaluate if the flow sensor of the mechanical ventilator can discriminate with similar or faster times the correct positioning of the tube during the tracheal intubation procedure compared to the colorimetric capnometer.
Time Frame
Immediately after the tracheal intubation attempt
Title
Number of mechanical breaths necessary to indicate the correct or not correct positioning of the tracheal tube by evaluating the flow-sensor curves and the color change of the colorimetric capnometer.
Description
To compare the minimal number of mechanical breaths provided by the ventilator to indicate the correct or not correct positioning of the tube in newborns undergoing tracheal intubation procedure, either evaluating the flow-sensor curves and the color change of the colorimetric capnometer, respectively. The aim is to evaluate if the flow sensor of the mechanical ventilator can indicate with lower or equal number of mechanical breaths the correct positioning of the tube during tracheal intubation compared to the colorimetric capnometer.
Time Frame
Immediately after the tracheal intubation attempt

10. Eligibility

Sex
All
Minimum Age & Unit of Time
20 Minutes
Maximum Age & Unit of Time
3 Months
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Term and preterm infants requiring tracheal intubation within the neonatal intensive care unit, both in elective and emergency conditions. Exclusion Criteria: intubation at birth intubation during cardiopulmonary resuscitation absence of informed consent by parents
Central Contact Person:
First Name & Middle Initial & Last Name or Official Title & Degree
Paolo Biban, MD
Phone
+00390458122041
Email
PAOLO.BIBAN@AOVR.VENETO.IT
First Name & Middle Initial & Last Name or Official Title & Degree
Irene Sibona, MD
Phone
00390458121825
Email
irene.sibona@aovr.veneto.it
Facility Information:
Facility Name
Azienda Ospedaliera Universitaria Integrata Verona
City
Verona
ZIP/Postal Code
37126
Country
Italy
Individual Site Status
Recruiting
Facility Contact:
First Name & Middle Initial & Last Name & Degree
Paolo Biban, MD
Email
paolo.biban@aovr.veneto.it

12. IPD Sharing Statement

Learn more about this trial

Flow Sensor and Colourimetric Capnometer in Verifying Tracheal Tube Positioning in Term and Preterm Infants

We'll reach out to this number within 24 hrs