Clinical Evaluation of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use Device and the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test Device

Primary Purpose

COVID-19, Influenza A, Influenza Type B

Status
Recruiting
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
Panbio™
Sponsored by
Abbott Rapid Dx
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for COVID-19

Eligibility Criteria

undefined - undefined (Child, Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria: Subjects from all age groups, able and willing to provide written informed consent or assent, suspected of having a respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms, and presenting at least two of the following symptoms: fever, headache, tiredness, dry cough, sore throat, runny or stuffy nose, muscle aches, loss of smell, loss of taste or shortness of breath, will be prospectively enrolled. Exclusion Criteria: Subject has had a nasal or a NP swab taken within the last 4 hours and is not available for further testing after 4 hours has elapsed. Subject has active nose bleeds or acute facial injuries/trauma. Subject has received a nasal vaccine (i.e., FluMist®) within the previous fourteen (14) days. Subject is currently taking or has taken an antiviral medication- e.g., Amantadine, Rimantadine, Relenza® (Zanamivir), Tamiflu® (Oseltamivir Phosphate), and Flumadine®-for influenza or COVID-19 within the previous thirty (30) days. Subject is currently enrolled in a study to evaluate an investigational drug. Self-tester or caregiver has prior medical or laboratory training. Self-tester or caregiver has a visual impairment that cannot be restored using glasses or contact lenses or lay user is unable to read the Instructions for Use. Subject has a condition deemed unfit to safely perform the test. Subject is unwilling or unable to provide informed consent. Vulnerable populations as deemed inappropriate for the study by the site's PI.

Sites / Locations

  • UCCT @ AFC Urgent Care- EasleyRecruiting

Arms of the Study

Arm 1

Arm Type

Experimental

Arm Label

Nasal sampling/Testing (Self-Test) + Nasopharyngeal sampling/Testing (Professional Use)

Arm Description

Lay users (self-tester or caregiver) will be provided with a Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test kit. Each lay user will collect one mid-turbinate nasal swab from both nostrils, perform and interpret the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test. All procedures for testing and result interpretation, including sample collection and extraction will be conducted by the lay user following the Instructions for Use provided in the kit. The first NP swab sample will be used to conduct a Panbio™ COVID- 19/FluA&B Rapid Panel Professional Use test. The second NP swab sample will be eluted in Universal Transport Medium (UTM) provided by Abbott/the Core Laboratory, labelled and stored, according to the laboratory manual. UTM samples will be shipped to the Core Laboratory for testing with RT-PCR protocols for Flu A, Flu B and SARS-CoV-2, according to the laboratory manual.

Outcomes

Primary Outcome Measures

Sensitivity and specificity of the Panbio™ COVID-19/Flu A&B Rapid Panel device.
To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID-19/Flu A&B Rapid Panel in fresh collected mid-turbinate nasal swab specimens from individuals suspected of respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms in a Near Patient Testing setting, compared to Flu and SARS-CoV-2 RT-PCR tested using NP swab samples. Concordance analyisis of investigational device and reference method. Acceptance criterion: Sensitivity >80%, Specificity >95% To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID- 19/Flu A&B Rapid Panel Self-Test device, as performed by lay users (self-testers or caregivers) using fresh collected mid-turbinate nasal self- collected samples, when compared to Flu and SARS-CoV-2 RT- PCR using NP swab samples.

Secondary Outcome Measures

Concordance of self-test and professional use device
To assess concordance between the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device using mid-turbinate nasal self-collected samples and the Professional Use device using NP samples collected by a trained health worker.
Usability of the professional use device
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel professional use device when used by trained healthcare workers. Usability is assessed by a questionnaire.
Usability of the self test device
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device, as performed by lay users with no laboratory or clinical experience. Usability testing will also comprise of the interpretation of mock devices. User and observer questionnaires will also be utilized to assess usability.

Full Information

First Posted
January 9, 2023
Last Updated
February 3, 2023
Sponsor
Abbott Rapid Dx
search

1. Study Identification

Unique Protocol Identification Number
NCT05715515
Brief Title
Clinical Evaluation of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use Device and the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test Device
Official Title
Clinical Evaluation of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use Device and the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test Device
Study Type
Interventional

2. Study Status

Record Verification Date
January 2023
Overall Recruitment Status
Recruiting
Study Start Date
March 1, 2023 (Anticipated)
Primary Completion Date
January 1, 2024 (Anticipated)
Study Completion Date
January 1, 2024 (Anticipated)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Abbott Rapid Dx

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
This study is designed as an international prospective, multicentric, clinical study to investigate the performance and usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use and Self- Test devices for the qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in human nasopharyngeal (NP) and mid-turbinate nasal swabs, respectively. This study is part of the performance evaluation to support the CE conformity assessment procedures.
Detailed Description
In total, a minimum of 1531 male and female subjects will be prospectively enrolled at multiple clinical sites. A minimum of 100 lay users will also be asked to interpret randomly assigned mock devices. After obtaining written informed consent, a study-specific subject identification number (ID) will be assigned to the subject. All eligible subjects will be assigned to both Professional Use arm and the Self-Test arm of the study. Subject demographics and a brief medical history will be collected. Lay users (self-tester or caregiver) will first collect one mid-turbinate nasal swab from both nostrils, and perform and interpret the Panbio™ Rapid Panel ST device, under the observation of a trained observer. All procedures for testing and result interpretation, including sample collection and extraction, will be conducted following the IFU provided in the kit. Each Panbio™ Rapid Panel ST result will be interpreted firstly by the lay user and followed immediately by the observer who will also photograph the test at the time of interpretation. Each lay user will fill out a user evaluation questionnaire, and a study staff member will complete an observer questionnaire for each lay user. A minimum of 100 lay users will also read mock device tests presented by a study staff member. Mock devices (simulated to represent testing results) will be used for usability testing (result interpretation). A study staff member will then collect two NP swabs (one swab from each nostril) from all subjects. The first NP swab sample will be used to conduct a Panbio™ Rapid Panel PU test, following the IFU. The test result will be photographed at the time of result interpretation. Each study staff member who performs a Panbio™ Rapid Panel PU test will fill out a user evaluation questionnaire once, to assess the usability of the device. The second NP swab sample will be eluted in Universal Transport Medium (UTM) provided by Abbott/the Core Laboratory, labelled and stored, according to the laboratory manual. UTM samples will be shipped to the Core Laboratory for testing with RT-PCR protocols for Flu A, Flu B and SARS-CoV-2, according to the laboratory manual. Each subject's demographic information, education, symptomology data, test results and questionnaires, where relevant, will be recorded on source documentation and transferred to an Electronic Data Capture system.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
COVID-19, Influenza A, Influenza Type B

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
1531 (Anticipated)

8. Arms, Groups, and Interventions

Arm Title
Nasal sampling/Testing (Self-Test) + Nasopharyngeal sampling/Testing (Professional Use)
Arm Type
Experimental
Arm Description
Lay users (self-tester or caregiver) will be provided with a Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test kit. Each lay user will collect one mid-turbinate nasal swab from both nostrils, perform and interpret the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test. All procedures for testing and result interpretation, including sample collection and extraction will be conducted by the lay user following the Instructions for Use provided in the kit. The first NP swab sample will be used to conduct a Panbio™ COVID- 19/FluA&B Rapid Panel Professional Use test. The second NP swab sample will be eluted in Universal Transport Medium (UTM) provided by Abbott/the Core Laboratory, labelled and stored, according to the laboratory manual. UTM samples will be shipped to the Core Laboratory for testing with RT-PCR protocols for Flu A, Flu B and SARS-CoV-2, according to the laboratory manual.
Intervention Type
Diagnostic Test
Intervention Name(s)
Panbio™
Other Intervention Name(s)
Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use Product, Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test Device
Intervention Description
The Panbio™ COVID-19/Flu A&B Rapid Panel Professional use device is a visual lateral flow assay, for rapid immunochromatographic qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in nasopharyngeal swabs, collected by a trained health worker. The Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device is a visual lateral flow assay, for rapid immunochromatographic qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in self-collected mid-turbinate nasal swabs. The test is designed to be used by lay users
Primary Outcome Measure Information:
Title
Sensitivity and specificity of the Panbio™ COVID-19/Flu A&B Rapid Panel device.
Description
To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID-19/Flu A&B Rapid Panel in fresh collected mid-turbinate nasal swab specimens from individuals suspected of respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms in a Near Patient Testing setting, compared to Flu and SARS-CoV-2 RT-PCR tested using NP swab samples. Concordance analyisis of investigational device and reference method. Acceptance criterion: Sensitivity >80%, Specificity >95% To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID- 19/Flu A&B Rapid Panel Self-Test device, as performed by lay users (self-testers or caregivers) using fresh collected mid-turbinate nasal self- collected samples, when compared to Flu and SARS-CoV-2 RT- PCR using NP swab samples.
Time Frame
"up to 24 months"
Secondary Outcome Measure Information:
Title
Concordance of self-test and professional use device
Description
To assess concordance between the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device using mid-turbinate nasal self-collected samples and the Professional Use device using NP samples collected by a trained health worker.
Time Frame
"up to 24 months"
Title
Usability of the professional use device
Description
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel professional use device when used by trained healthcare workers. Usability is assessed by a questionnaire.
Time Frame
"up to 24 months"
Title
Usability of the self test device
Description
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device, as performed by lay users with no laboratory or clinical experience. Usability testing will also comprise of the interpretation of mock devices. User and observer questionnaires will also be utilized to assess usability.
Time Frame
"up to 24 months"

10. Eligibility

Sex
All
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Subjects from all age groups, able and willing to provide written informed consent or assent, suspected of having a respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms, and presenting at least two of the following symptoms: fever, headache, tiredness, dry cough, sore throat, runny or stuffy nose, muscle aches, loss of smell, loss of taste or shortness of breath, will be prospectively enrolled. Exclusion Criteria: Subject has had a nasal or a NP swab taken within the last 4 hours and is not available for further testing after 4 hours has elapsed. Subject has active nose bleeds or acute facial injuries/trauma. Subject has received a nasal vaccine (i.e., FluMist®) within the previous fourteen (14) days. Subject is currently taking or has taken an antiviral medication- e.g., Amantadine, Rimantadine, Relenza® (Zanamivir), Tamiflu® (Oseltamivir Phosphate), and Flumadine®-for influenza or COVID-19 within the previous thirty (30) days. Subject is currently enrolled in a study to evaluate an investigational drug. Self-tester or caregiver has prior medical or laboratory training. Self-tester or caregiver has a visual impairment that cannot be restored using glasses or contact lenses or lay user is unable to read the Instructions for Use. Subject has a condition deemed unfit to safely perform the test. Subject is unwilling or unable to provide informed consent. Vulnerable populations as deemed inappropriate for the study by the site's PI.
Central Contact Person:
First Name & Middle Initial & Last Name or Official Title & Degree
Simon Kordowich
Phone
0000000
Email
simon.kordowich@abbott.com
First Name & Middle Initial & Last Name or Official Title & Degree
Peter Kelly, PhD
Phone
00000000
Email
peter.kelly@abbott.com
Facility Information:
Facility Name
UCCT @ AFC Urgent Care- Easley
City
Easley
State/Province
South Carolina
ZIP/Postal Code
29640
Country
United States
Individual Site Status
Recruiting
Facility Contact:
First Name & Middle Initial & Last Name & Degree
Krystal Reese