search
Back to results

Clinical Assessment of a Closed-loop System With Glucagon, Exercise and Mixed Meals

Primary Purpose

Type 1 Diabetes

Status
Completed
Phase
Not Applicable
Locations
United Kingdom
Study Type
Interventional
Intervention
Bi-hormonal closed loop pump
Standard opened loop pump
Insulin closed loop
Sponsored by
Imperial College London
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Type 1 Diabetes focused on measuring Type 1 diabetes, Hypoglycaemia, Closed loop insulin pump, Artificial pancreas, Exercise, Mixed meals, Bihormonal pump

Eligibility Criteria

18 Years - undefined (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • Adults over 18 years of age
  • Type 1 diabetes confirmed on the basis of clinical features and a fasting c-peptide <200 pmol/L
  • Type 1 diabetes for greater than 1 year
  • Continuous subcutaneous insulin infusion for greater than 6 months
  • HbA1c < 10% (86mmol/mol)

Exclusion Criteria:

  • Recurrent severe hypoglycaemia and hypoglycaemia unawareness
  • Pregnant or planning pregnancy
  • Breastfeeding
  • Enrolled in other clinical trials
  • Have active malignancy or under investigation for malignancy
  • Allergic to lactose
  • Allergic to glucagon

Sites / Locations

  • Imperial College

Arms of the Study

Arm 1

Arm 2

Arm 3

Arm Type

Experimental

Active Comparator

Placebo Comparator

Arm Label

Bihormonal closed loop

Standard opened loop pump

Insulin only

Arm Description

Bi-hormonal closed loop pump will be used in the exercise and mixed meal substudies

Standard opened loop pump will be used in the exercise and mixed meal substudies

Insulin only closed loop

Outcomes

Primary Outcome Measures

Percentage of Time in Target Range Over 24 Hour
The primary outcome from the studies will be time spent with a glucose concentration in the target range (3.9-10.0mmol/l).

Secondary Outcome Measures

Full Information

First Posted
March 5, 2015
Last Updated
October 31, 2019
Sponsor
Imperial College London
search

1. Study Identification

Unique Protocol Identification Number
NCT02397265
Brief Title
Clinical Assessment of a Closed-loop System With Glucagon, Exercise and Mixed Meals
Official Title
Clinical Assessment of a Closed-loop System With Glucagon, Exercise and Mixed Meals
Study Type
Interventional

2. Study Status

Record Verification Date
October 2019
Overall Recruitment Status
Completed
Study Start Date
December 3, 2014 (Actual)
Primary Completion Date
July 22, 2015 (Actual)
Study Completion Date
July 22, 2015 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Imperial College London

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
Yes

5. Study Description

Brief Summary
The diabetes technology group at Imperial College have developed a bio-inspired artificial pancreas (BiAP) system which uses a control algorithm based on a mathematical model of beta-cell physiology. The algorithm is implemented on a miniature silicon microchip within a portable handheld device, which interfaces the components of the artificial pancreas. Development of closed-loop insulin delivery devices to intensify control without hypoglycaemia has been extensively reviewed and have shown encouraging results . However, they have not yet proven to be robust when challenged with uncertainty and the external challenges (such as mixed meal contents, physical exercise, physiological stress and intercurrent illness) that people with Type 1 Diabetes Mellitus (T1DM) may be exposed to outside the clinical environment. The principal research objective is to assess the safety and efficacy of a closed-loop system for T1DM compared to standard insulin pump therapy (open-loop). The primary outcome from the studies will be % time spent with a glucose concentration in the target range (3.9-10.0mmol/l). This outcome incorporates safety as it ensures subjects do not have low or high glucose excursions and is the principal measure of efficacy for closed-loop insulin delivery systems in the scientific literature. Other measured outcomes will be % time spent in euglycaemia (3.9-7.8mmol/l), % time spent in hypoglycaemia (<3.9mmol/l), % time spent in hyperglycaemia (>10mmol/l), mean venous blood and sensor glucose, glycaemic variability as measured by standard metrics (Standard Deviation, Continuous Overlapping Net Glycaemic Action, Lability Index, J-Index, Glycaemic Risk Assessment Diabetes Equation, Mean Of Daily Differences, Mean Amplitude of Glucose Excursion, Average Daily Risk Range, M-VALUE, Mean Average Glucose), glycaemic risk as measured by Low Blood Glucose Index (LBGI) and High Blood Glucose Index (HBGI), closed-loop error grid analysis, glucose area under the curve. All measures have been previously published and validated. This clinical trial protocol assesses the artificial pancreas system in three separate sub-studies: In a bi-hormonal (insulin and glucagon) configuration During and after exercise with bi-hormonal closed loop, and standard insulin opened loop During and after meals of mixed composition with bi-hormonal closed loop, and standard insulin opened loop
Detailed Description
Methodology Randomised controlled cross-over open label study Sub study 1 (n=10) In sub-study 1 which includes the first bio-inspired artificial pancreas study using bi-hormonal (insulin and glucagon) control a fasting 6-hour bi-hormonal closed-loop study will be conducted to assess proof of concept and safety prior to the 25-hour randomised controlled crossover study Each subject will then be randomised to attend either a closed-loop visit or an open-loop visit first. Once either visit is completed each subject will crossover and attend the remaining visit. Sub-study 2 (n=20) The aim of sub-study 2 is to challenge the bi-hormonal pump during a moderate exercise. Subjects will be initially randomised to either the bi-hormonal or the standard opened loop and then cross over. Each study will last for 25 hours • Exercise protocol: Subjects will be connected to the gas analyser and heart rate/ECG monitor as per COSMED manufacturer instructions. The oxygen consumption (VO2)/Carbon dioxide production (VCO2) and heart rate/ECG trace will be displayed continuously in real time while the subject is exercising. Warm-up: 3 minutes of cycling at low resistance (20-30 watts) while maintaining a speed of 60-80 revolutions per minute After the warm-up the subjects will undergo a 30 minutes moderate intensity exercise session on the bike aiming to maintain their VO2 between 10-20% above their 1st ventilatory anaerobic threshold. If their VO2 falls below or goes above this level then the resistance will be increased or decreased accordingly. The target VO2 and the starting resistance (watts) will be individually set based on their baseline exercise test outcome. The estimated time needed to complete the exercise test is 35-40 minutes. Real-time continuous glucose monitoring alarms will be set at 4mmol/L and 15mmol/L and will be audible by the subject and the research team. The glucagon solution will be replaced with a freshly reconstituted glucagon solution every 8 hours throughout the closed-loop study If the venous blood glucose concentrations fall below 3.5mmol/L or if the subject experiences hypoglycaemia symptoms then hypoglycaemia will be confirmed by an additional venous blood plasma glucose sample and will be treated according to Imperial College Hospitals National Health Service Trust Guidelines. After 25 hours of closed-loop, at 12:00 the next day, the subject's own insulin pump will be primed, reconnected and started as per the subject's usual insulin regime. Once running, the closed loop system will be disconnected. The subject can then eat and drink freely and may be discharged after 2 hours, or when glucose concentrations are stable. The same protocol will be used when subjects cross over for the standard opened loop 25 hour study for the remaining visit. Sub-study 3: Bi-hormonal closed-loop control during- and after mixed meals (n=20) A high fat/high carbohydrate content dinner (45g fat, 80g CHO) will be given at 19:00, a high glycaemic index breakfast (40g CHO) at 07:00 and a high protein/low carbohydrate lunch ( 30g protein,10g CHO) at 12:00. Subjects will be free to gently mobilise around the clinical research unit, smoking will not be permitted. Subjects may drink unlimited water throughout the visit. Real-time continuous glucose monitoring alarms will be set at 4mmol/L and 15mmol/L and will be audible by the subject and the research team. The glucagon solution will be replaced with a freshly reconstituted glucagon solution every 8 hours throughout the closed-loop studies. If the venous blood glucose concentrations fall below 3.5mmol/L or if the subject experiences hypoglycaemia symptoms then hypoglycaemia will be confirmed by an additional venous blood plasma glucose sample and will be treated according to Imperial College Hospitals National Health Service Trust Guidelines. After 25 hours of closed-loop, at 18:00 the next day, the subject's own insulin pump will be primed, reconnected and started as per the subject's usual insulin regime. Once running, the closed-loop system will be disconnected. The subject can then eat and drink freely and may be discharged after 2 hours, or when glucose concentrations are stable. The same protocol will be applied after cross over but with a standard opened loop insulin pump for the remaining visit.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Type 1 Diabetes
Keywords
Type 1 diabetes, Hypoglycaemia, Closed loop insulin pump, Artificial pancreas, Exercise, Mixed meals, Bihormonal pump

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Crossover Assignment
Masking
None (Open Label)
Allocation
Randomized
Enrollment
24 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Bihormonal closed loop
Arm Type
Experimental
Arm Description
Bi-hormonal closed loop pump will be used in the exercise and mixed meal substudies
Arm Title
Standard opened loop pump
Arm Type
Active Comparator
Arm Description
Standard opened loop pump will be used in the exercise and mixed meal substudies
Arm Title
Insulin only
Arm Type
Placebo Comparator
Arm Description
Insulin only closed loop
Intervention Type
Device
Intervention Name(s)
Bi-hormonal closed loop pump
Intervention Description
Using a bio-inspired artificial pancreas consisting of a bi-hormonal closed loop pump
Intervention Type
Device
Intervention Name(s)
Standard opened loop pump
Intervention Description
Using a standard opened loop pump
Intervention Type
Device
Intervention Name(s)
Insulin closed loop
Intervention Description
Using a closed loop
Primary Outcome Measure Information:
Title
Percentage of Time in Target Range Over 24 Hour
Description
The primary outcome from the studies will be time spent with a glucose concentration in the target range (3.9-10.0mmol/l).
Time Frame
24 hours

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Adults over 18 years of age Type 1 diabetes confirmed on the basis of clinical features and a fasting c-peptide <200 pmol/L Type 1 diabetes for greater than 1 year Continuous subcutaneous insulin infusion for greater than 6 months HbA1c < 10% (86mmol/mol) Exclusion Criteria: Recurrent severe hypoglycaemia and hypoglycaemia unawareness Pregnant or planning pregnancy Breastfeeding Enrolled in other clinical trials Have active malignancy or under investigation for malignancy Allergic to lactose Allergic to glucagon
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Nick Oliver, MRCP
Organizational Affiliation
Imperial College London
Official's Role
Principal Investigator
Facility Information:
Facility Name
Imperial College
City
London
Country
United Kingdom

12. IPD Sharing Statement

Plan to Share IPD
No
Citations:
Citation
Nathan D, Cleary P, Backlund J, Genuth S, Lachin J, Orchard T, Raskin R, Zinman B. Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Study Research Group. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes,
Results Reference
background
PubMed Identifier
18694975
Citation
Buckingham B, Wilson DM, Lecher T, Hanas R, Kaiserman K, Cameron F. Duration of nocturnal hypoglycemia before seizures. Diabetes Care. 2008 Nov;31(11):2110-2. doi: 10.2337/dc08-0863. Epub 2008 Aug 11.
Results Reference
background
PubMed Identifier
10097898
Citation
Sovik O, Thordarson H. Dead-in-bed syndrome in young diabetic patients. Diabetes Care. 1999 Mar;22 Suppl 2:B40-2. Erratum In: Diabetes Care 1999 Aug;22(8):1389.
Results Reference
background
PubMed Identifier
21343892
Citation
Hovorka R. Closed-loop insulin delivery: from bench to clinical practice. Nat Rev Endocrinol. 2011 Feb 22;7(7):385-95. doi: 10.1038/nrendo.2011.32.
Results Reference
background
PubMed Identifier
22025773
Citation
Cobelli C, Renard E, Kovatchev B. Artificial pancreas: past, present, future. Diabetes. 2011 Nov;60(11):2672-82. doi: 10.2337/db11-0654. No abstract available.
Results Reference
background
PubMed Identifier
21129325
Citation
Ward WK, Massoud RG, Szybala CJ, Engle JM, El Youssef J, Carroll JM, Roberts CT Jr, DiMarchi RD. In vitro and in vivo evaluation of native glucagon and glucagon analog (MAR-D28) during aging: lack of cytotoxicity and preservation of hyperglycemic effect. J Diabetes Sci Technol. 2010 Nov 1;4(6):1311-21. doi: 10.1177/193229681000400604.
Results Reference
background
PubMed Identifier
20144397
Citation
Oliver N, Georgiou P, Johnston D, Toumazou C. A benchtop closed-loop system controlled by a bio-inspired silicon implementation of the pancreatic beta cell. J Diabetes Sci Technol. 2009 Nov 1;3(6):1419-24. doi: 10.1177/193229680900300623.
Results Reference
background
PubMed Identifier
22768892
Citation
Herrero P, Georgiou P, Oliver N, Johnston DG, Toumazou C. A bio-inspired glucose controller based on pancreatic beta-cell physiology. J Diabetes Sci Technol. 2012 May 1;6(3):606-16. doi: 10.1177/193229681200600316.
Results Reference
background
PubMed Identifier
23911175
Citation
Herrero P, Georgiou P, Oliver N, Reddy M, Johnston D, Toumazou C. A composite model of glucagon-glucose dynamics for in silico testing of bihormonal glucose controllers. J Diabetes Sci Technol. 2013 Jul 1;7(4):941-51. doi: 10.1177/193229681300700416.
Results Reference
background

Learn more about this trial

Clinical Assessment of a Closed-loop System With Glucagon, Exercise and Mixed Meals

We'll reach out to this number within 24 hrs