Clinical Outcomes of MICROPURE and PodEye Monofocal Intraocular Lenses (PHY1803)

Primary Purpose

Cataract, Lens Opacities

Status
Completed
Phase
Not Applicable
Locations
Czechia
Study Type
Interventional
Intervention
Micropure 1.2.3.
PODEYE
Sponsored by
Beaver-Visitec International, Inc.
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Cataract focused on measuring Intraocular lens, Monofocal, Hydrophobic

Eligibility Criteria

45 Years - undefined (Adult, Older Adult)All SexesAccepts Healthy Volunteers

Inclusion Criteria:

  • Cataractous eyes with no comorbidity
  • Availability, willingness and sufficient cognitive awareness to comply with examination procedures
  • Clear intraocular media other than cataract;
  • Signed informed consent

Exclusion Criteria:

  • Age of patient < 45 years;
  • Irregular astigmatism;
  • Regular corneal astigmatism >0.75 dioptres by an automatic keratometer or biometer or >1.0 dioptres if the steep axis of cylinder is between 90° and 120°;
  • Difficulty for cooperation (distance from their home, general health condition);
  • Previous intraocular or corneal surgery;
  • Traumatic cataract;
  • Any ocular comorbidity;
  • Instability of keratometry or biometry measurements.

Sites / Locations

  • GEMINI Eye Clinic

Arms of the Study

Arm 1

Arm 2

Arm Type

Experimental

Active Comparator

Arm Label

Investigational Device Micropure 1.2.3.

Comparator PODEYE

Arm Description

Implantation of monofocal intraocular lens (IOL) "Micropure 1.2.3." in one of the eyes of the study subject

Implantation of monofocal intraocular lens (IOL) "PODEYE" in the contralateral eye of the study subject

Outcomes

Primary Outcome Measures

monocular Corrected Distance Visual Acuity (CDVA)
Statistically non-inferior visual acuity outcomes on monocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions compared to eyes implanted with the comparator PODEYE at the 6 months follow up visit.

Secondary Outcome Measures

Manifested refraction
The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2018.
Monocular Uncorrected Distance Visual Acuity (UDVA) under photopic light conditions
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is performed under photopic light conditions
Monocular Uncorrected Distance Visual Acuity (UDVA) under mesopic light conditions
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is performed under mesopic light conditions
Monocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is performed under photopic light conditions
Monocular Corrected Distance Visual Acuity (CDVA) under mesopic light conditions
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is performed under mesopic light conditions
Monocular Contrast Sensitivity under photopic light conditions
Contrast Sensitivity under photopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Monocular Contrast Sensitivity under mesopic light conditions
Contrast Sensitivity under mesopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Slitlamp examination - Corneal status
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Corneal status.
Slitlamp examination - Fundus
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Fundus.
Slitlamp examination - Signs of inflammation
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Signs of inflammation.
Slitlamp examination - Pupillary block
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Pupillary block.
Slitlamp examination - Retinal detachment
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Retinal detachment.
Slitlamp examination - Status of anterior and posterior capsule
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Status of anterior and posterior capsule.
Slitlamp examination - IOL decentration
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL decentration.
Slitlamp examination IOL tilt
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL tilt.
Slitlamp examination - IOL discoloration
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL discoloration.
Slitlamp examination - IOL opacity
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL opacity.

Full Information

First Posted
January 4, 2019
Last Updated
July 8, 2021
Sponsor
Beaver-Visitec International, Inc.
search

1. Study Identification

Unique Protocol Identification Number
NCT03796624
Brief Title
Clinical Outcomes of MICROPURE and PodEye Monofocal Intraocular Lenses
Acronym
PHY1803
Official Title
Comparison of Clinical Outcomes After Implantation of MICROPURE (PhysIOL) and PodEye (PhysIOL) Monofocal Intraocular Lenses Into the Two Eyes of the Same Patient
Study Type
Interventional

2. Study Status

Record Verification Date
July 2021
Overall Recruitment Status
Completed
Study Start Date
February 12, 2019 (Actual)
Primary Completion Date
June 17, 2021 (Actual)
Study Completion Date
June 17, 2021 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Beaver-Visitec International, Inc.

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
Prospective, non-randomised, open, controlled, single-center post-market clinical follow study about Micropure 1.2.3. and PODEYE intraocular lenses.
Detailed Description
This clinical investigation is a prospective, non-randomised, open, controlled, single-center post-market clinical follow whereby study patients undergoing routine cataract surgery will have implantation of monofocal intraocular lens Micropure 1.2.3. (PhysIOL, Liège, Belgium) in one eye and implantation of monofocal intraocular lens PODEYE (PhysIOL, Liège, Belgium) in the contralateral eye of the same patient. The study purpose is to obtain clinical data on visual acuity and contrast sensitivity on patients implanted with Micropure 1.2.3. and PODEYE The devices under investigation (Micropure 1.2.3. and PODEYE) are a monofocal glistening-free hydrophobic acrylic intraocular lenses (IOLs) manufactured by the sponsor of this study PhysIOL sa/nv. The optical properties of the lenses are very comparable. The main difference is the mechanical design of the haptics, that is not expected to have an influence on the clinical outcomes. The IOLs will be implanted as part of the routine cataract surgery on patients suffering from cataract development. In total 76 patients will be recruited for this clinical study and receive the implantation of Micropure 1.2.3. and POPDEYE intraocular lenses. Subjects participating in the trial will attend study visits over a period of 12 months.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Cataract, Lens Opacities
Keywords
Intraocular lens, Monofocal, Hydrophobic

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
Randomized
Enrollment
76 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Investigational Device Micropure 1.2.3.
Arm Type
Experimental
Arm Description
Implantation of monofocal intraocular lens (IOL) "Micropure 1.2.3." in one of the eyes of the study subject
Arm Title
Comparator PODEYE
Arm Type
Active Comparator
Arm Description
Implantation of monofocal intraocular lens (IOL) "PODEYE" in the contralateral eye of the study subject
Intervention Type
Device
Intervention Name(s)
Micropure 1.2.3.
Intervention Description
Implantation of intraocular lens (IOL). Name: "MicroPure 1.2.3." It is a monofocal intraocular lens consisting of hydrophobic acrylic material. One IOL per patient will be implanted
Intervention Type
Device
Intervention Name(s)
PODEYE
Intervention Description
Implantation of intraocular lens (IOL). Name: "PODEYE" It is a monofocal intraocular lens consisting of hydrophobic acrylic material. The IOL will be implanted in the contralateral eye of the same patient already implanted with Micropure 1.2.3.
Primary Outcome Measure Information:
Title
monocular Corrected Distance Visual Acuity (CDVA)
Description
Statistically non-inferior visual acuity outcomes on monocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions compared to eyes implanted with the comparator PODEYE at the 6 months follow up visit.
Time Frame
6 months postoperative
Secondary Outcome Measure Information:
Title
Manifested refraction
Description
The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2018.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Monocular Uncorrected Distance Visual Acuity (UDVA) under photopic light conditions
Description
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is performed under photopic light conditions
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Monocular Uncorrected Distance Visual Acuity (UDVA) under mesopic light conditions
Description
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is performed under mesopic light conditions
Time Frame
preoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Monocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions
Description
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is performed under photopic light conditions
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Monocular Corrected Distance Visual Acuity (CDVA) under mesopic light conditions
Description
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is performed under mesopic light conditions
Time Frame
preoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Monocular Contrast Sensitivity under photopic light conditions
Description
Contrast Sensitivity under photopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Time Frame
6 months postoperative
Title
Monocular Contrast Sensitivity under mesopic light conditions
Description
Contrast Sensitivity under mesopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Time Frame
6 months postoperative
Title
Slitlamp examination - Corneal status
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Corneal status.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - Fundus
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Fundus.
Time Frame
preoperative, 12 months postoperative
Title
Slitlamp examination - Signs of inflammation
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Signs of inflammation.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - Pupillary block
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Pupillary block.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - Retinal detachment
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Retinal detachment.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - Status of anterior and posterior capsule
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Status of anterior and posterior capsule.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - IOL decentration
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL decentration.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination IOL tilt
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL tilt.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - IOL discoloration
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL discoloration.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Slitlamp examination - IOL opacity
Description
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • IOL opacity.
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Other Pre-specified Outcome Measures:
Title
Intraocular pressure (IOP) measurement
Description
The IOP will be measured with non-contact tonometer as part of the routine follow up examinations
Time Frame
preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
Title
Questionnaire on IOL implantation
Description
A questionnaire will be handed out to the surgeon right after the surgery to document the ease of use and possible issues during IOL implantation. These data will be used to compare the outcomes to different surgery techniques or injectors. The questionnaire is not validated and the outcomes serve only for the sponsor to receive feedback if one or the other IOL is easier to implant.
Time Frame
peroperative
Title
Keratometry
Description
Keratometric measurements are performed to calculate the required IOL power
Time Frame
preoperative
Title
Biometry
Description
Biometry measurements are performed to calculate the required IOL power
Time Frame
preoperative

10. Eligibility

Sex
All
Minimum Age & Unit of Time
45 Years
Accepts Healthy Volunteers
Accepts Healthy Volunteers
Eligibility Criteria
Inclusion Criteria: Cataractous eyes with no comorbidity Availability, willingness and sufficient cognitive awareness to comply with examination procedures Clear intraocular media other than cataract; Signed informed consent Exclusion Criteria: Age of patient < 45 years; Irregular astigmatism; Regular corneal astigmatism >0.75 dioptres by an automatic keratometer or biometer or >1.0 dioptres if the steep axis of cylinder is between 90° and 120°; Difficulty for cooperation (distance from their home, general health condition); Previous intraocular or corneal surgery; Traumatic cataract; Any ocular comorbidity; Instability of keratometry or biometry measurements.
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Pavel Stodůlka, MD, PhD
Organizational Affiliation
Gemini Eye Clinic
Official's Role
Principal Investigator
Facility Information:
Facility Name
GEMINI Eye Clinic
City
Zlin
ZIP/Postal Code
76001
Country
Czechia

12. IPD Sharing Statement