Evaluation of Synthesized 18-Lead ECG in the Emergency Department
Primary Purpose
Chest Pain
Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
Two consecutive 15-lead ECGs
Sponsored by
About this trial
This is an interventional diagnostic trial for Chest Pain
Eligibility Criteria
Inclusion Criteria:
- Subjects with a chief complaint of chest pain or chest discomfort or chest pressure
- Troponin test has been ordered (unknown positive at time of recruitment)
- Subjects are capable of providing informed consent
- English speaking
Exclusion Criteria:
- STEMI patients
- Patients who have not had a troponin test ordered
- Pregnant women
- Patients under 18 years of age
- Prisoners
Sites / Locations
- North Shore University Hospital
- Long Island Jewish Medical Center
Arms of the Study
Arm 1
Arm Type
Other
Arm Label
Test
Arm Description
Obtain two consecutive 15-lead ECGs; The first 15-lead ECG will have an additional three electrodes/stickers on the right side of the chest and the second 15-lead ECG will have an additional three electrodes/stickers on the posterior side.
Outcomes
Primary Outcome Measures
Sensitivity and Specificity Agreement Between Synthesized and Actual 18-lead ECGs in the Identification of ST Elevation, ST Depression, and T Wave Inversion
Sensitivity, Specificity, Positive Predictive Value and Negative Predictive Value for comparison of actual 18-lead ECG vs synthesized ECG leads. All measures at 95% confidence Index.
Kappa Agreement Between Synthesized and Actual 18-lead ECGs in the Identification of ST Elevation, ST Depression, and T Wave Inversion
Kappa value for comparison of actual 18-lead ECG vs synthesized ECG leads. All measures at 95% confidence Index.
Secondary Outcome Measures
Full Information
NCT ID
NCT03426436
First Posted
February 1, 2018
Last Updated
November 22, 2019
Sponsor
Nihon Kohden
Collaborators
Feinstein Institute for Medical Research
1. Study Identification
Unique Protocol Identification Number
NCT03426436
Brief Title
Evaluation of Synthesized 18-Lead ECG in the Emergency Department
Official Title
Evaluation of Synthesized 18-Lead ECG in the Emergency Department
Study Type
Interventional
2. Study Status
Record Verification Date
October 2019
Overall Recruitment Status
Completed
Study Start Date
February 14, 2018 (Actual)
Primary Completion Date
August 16, 2018 (Actual)
Study Completion Date
August 15, 2019 (Actual)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Nihon Kohden
Collaborators
Feinstein Institute for Medical Research
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
Yes
Device Product Not Approved or Cleared by U.S. FDA
Yes
Product Manufactured in and Exported from the U.S.
No
Data Monitoring Committee
No
5. Study Description
Brief Summary
This study, sponsored by Nihon Kohden, aims to determine the sensitivity and specificity of synthesized 18-lead electrocardiogram (ECG) in the diagnosis of posterior-lateral and/or right-ventricular ischemia, using actual 18-lead ECG as the gold standard. The synthesized 18-lead ECG calculates six extra leads of information from the standard 12-lead ECG done in the Emergency Department (ED). Nihon Kohden has created synECi18 Technology, which can mathematically synthesize and display the extra leads.
The patient population (n=300) will consist of patients who present to the North Shore University Hospital or Long Island Jewish Medical Center ED with a chief complaint of chest pain, chest pressure, or chest discomfort. These patients will be receiving a standard 12-lead ECG as part of their routine clinical care. Patients with ST-Elevation Myocardial Infarction (STEMI) and a known positive troponin value will be excluded from the study.
Trained ED study technicians will use ECG machines provided by Nihon Kohden to obtain two consecutive 15-lead ECGs. The two 15-lead ECGs are then saved onto the ECG machine.The saved information will later be input into an algorithm to calculate the synthesized 18-leads. The actual 18-leads (composed of the two 15-leads conducted in the ED) will be compared with the synthesized 18-leads produced by the synECi18 technology. A study cardiologists will evaluate the actual 18-lead ECGs and synthesized 18-lead ECGs and determine whether there is posterior-lateral and/or right-sided ventricular ischemia. The cardiologists will be blinded to the type of waveform they are analyzing (synthesized versus actual).
The study will not interfere with patient care or treatment, however, the two 15-lead ECGs done in the ED will be shown to the physician who can determine whether or not to order an official 18-lead ECG in the hospital's electronic medical record system.
It is hypothesized that the synECi18 synthesized 18-lead ECGs will provide high sensitivity and specificity for diagnosing posterior-lateral and/or right-ventricular ischemia, with the actual 18-lead ECGs being used as the reference for comparison.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Chest Pain
7. Study Design
Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Masking Description
The cardiologist will be blinded to the type of waveform they are analyzing (synthesized versus actual).
Allocation
N/A
Enrollment
314 (Actual)
8. Arms, Groups, and Interventions
Arm Title
Test
Arm Type
Other
Arm Description
Obtain two consecutive 15-lead ECGs; The first 15-lead ECG will have an additional three electrodes/stickers on the right side of the chest and the second 15-lead ECG will have an additional three electrodes/stickers on the posterior side.
Intervention Type
Diagnostic Test
Intervention Name(s)
Two consecutive 15-lead ECGs
Intervention Description
The 15-lead ECG consists of a traditional 12-lead ECG and an additional three leads on the right side of the body. The first 15-lead ECG will have an additional three electrodes/stickers on the right side of the chest and the second 15-lead ECG will have an additional three electrodes/stickers on the posterior side.
Primary Outcome Measure Information:
Title
Sensitivity and Specificity Agreement Between Synthesized and Actual 18-lead ECGs in the Identification of ST Elevation, ST Depression, and T Wave Inversion
Description
Sensitivity, Specificity, Positive Predictive Value and Negative Predictive Value for comparison of actual 18-lead ECG vs synthesized ECG leads. All measures at 95% confidence Index.
Time Frame
30 minutes
Title
Kappa Agreement Between Synthesized and Actual 18-lead ECGs in the Identification of ST Elevation, ST Depression, and T Wave Inversion
Description
Kappa value for comparison of actual 18-lead ECG vs synthesized ECG leads. All measures at 95% confidence Index.
Time Frame
30 minutes
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
Subjects with a chief complaint of chest pain or chest discomfort or chest pressure
Troponin test has been ordered (unknown positive at time of recruitment)
Subjects are capable of providing informed consent
English speaking
Exclusion Criteria:
STEMI patients
Patients who have not had a troponin test ordered
Pregnant women
Patients under 18 years of age
Prisoners
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Timmy Li, PhD
Organizational Affiliation
Northwell Health
Official's Role
Principal Investigator
Facility Information:
Facility Name
North Shore University Hospital
City
Manhasset
State/Province
New York
ZIP/Postal Code
11030
Country
United States
Facility Name
Long Island Jewish Medical Center
City
New Hyde Park
State/Province
New York
ZIP/Postal Code
11040
Country
United States
12. IPD Sharing Statement
Learn more about this trial
Evaluation of Synthesized 18-Lead ECG in the Emergency Department