Micra Atrial TRacking Using A Ventricular AccELerometer 2 (MARVEL2)

Primary Purpose

Atrioventricular Conduction Block

Status
Completed
Phase
Not Applicable
Locations
International
Study Type
Interventional
Intervention
MARVEL 2 Algorithm Monitor Mode then Adaptive Mode
Sponsored by
Medtronic Cardiac Rhythm and Heart Failure
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Atrioventricular Conduction Block

Eligibility Criteria

18 Years - undefined (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • Subject has been implanted with a Micra TPS (Model MC1VR01) with remaining device longevity of 6 years or more or is expected to be implanted with a Micra TPS.
  • Subject has history of AV block*
  • Subject is ≥ 18 years old and as per required local law.
  • Subject (and/or witness as applicable per local regulations) provides signed and dated authorization and/or consent per institution and local requirements.
  • Subject is willing and able to comply with the protocol. *This includes subjects with normal sinus function and persistent 3rd degree AV block and subjects with other forms of AV block.

Exclusion Criteria:

  • Subject is currently enrolled or planning to participate in a potentially confounding drug or device trial during the study. Co-enrollment in concurrent trials is only allowed when documented pre-approval is obtained from the Medtronic Clinical Research Specialist.
  • Subject is pregnant (if required by local law, women of child-bearing potential must undergo a pregnancy test within seven days prior to MARVEL 2 study procedures).
  • Subject meets any exclusion criteria required by local law (age or other).

Sites / Locations

  • Baptist Medical Center Jacksonville
  • Baptist Hospital
  • North Shore University Hospital- Northwell
  • NYU Langone Medical Center
  • Duke University Medical Center
  • Kepler Universitätsklinikum Med Campus III
  • UZ Leuven - Campus Gasthuisberg
  • Odense Universitetshospital
  • Hôpital Haut-Lévêque - CHU de Bordeaux
  • Prince of Wales Hospital
  • Institut Jantung Negara
  • Hospital Universitari Clínic de Barcelona

Arms of the Study

Arm 1

Arm Type

Experimental

Arm Label

MARVEL 2 Algorithm Monitor Mode, Then MARVEL 2 Adaptive Mode

Arm Description

Participants first received MARVEL 2 algorithm monitor mode which provides standard VVI pacing for approximately 20 minutes followed by MARVEL 2 algorithm adaptive mode for approximately 2 hours which provides VDD pacing.

Outcomes

Primary Outcome Measures

Atrial Ventricular Pacing Synchrony Success During Rest
A subject will meet the primary endpoint if a paced or sensed ventricular beat is within 300 ms following an ECG confirmed P-wave for at least 70% of the ECG confirmed P-waves. The primary endpoint will be evaluated during the MARVEL 2 setup phase during MARVEL 2 Algorithm monitor mode (VDI pacing which is effectively VVI pacing) and during the 20-minute resting period in which the MARVEL 2 Algorithm is programmed to Adaptive mode (VDD pacing)
Number of Participants Free From Inappropriate Pacemaker Function Events
A participant will meet the primary endpoint if they are free from the following MARVEL 2 software related events during the entire Holter monitoring period where the MARVEL 2 features are enabled. Inappropriate pacemaker function events are defined as: Cardiac pauses lasting longer than 2 cardiac cycles (where cardiac cycle length is defined by the programmed lower rate interval of the pacemaker), OR Pacemaker oversensing induced tachycardia exceeding 3-minutes, defined as oversensing of the pacemaker's accelerometer signal leading to a heart rate exceeding 100 beats per minute.

Secondary Outcome Measures

Left Ventricular Outflow Tract Velocity Time Integral
The secondary endpoint is left ventricular outflow tract (LVOT) velocity time integral (VTI) as obtained from echocardiogram while the MARVEL 2 features are in adaptive mode (VDD pacing) and while the MARVEL 2 features are in monitor mode (VDI pacing which is effectively VVI pacing). This was measured by the echo core laboratory that was blinded to the study participant and MARVEL 2 algorithm mode.

Full Information

First Posted
November 14, 2018
Last Updated
May 29, 2020
Sponsor
Medtronic Cardiac Rhythm and Heart Failure
search

1. Study Identification

Unique Protocol Identification Number
NCT03752151
Brief Title
Micra Atrial TRacking Using A Ventricular AccELerometer 2
Acronym
MARVEL2
Official Title
Micra Atrial TRacking Using A Ventricular AccELerometer 2
Study Type
Interventional

2. Study Status

Record Verification Date
May 2020
Overall Recruitment Status
Completed
Study Start Date
January 22, 2019 (Actual)
Primary Completion Date
August 7, 2019 (Actual)
Study Completion Date
August 7, 2019 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Medtronic Cardiac Rhythm and Heart Failure

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
Yes
Device Product Not Approved or Cleared by U.S. FDA
Yes
Data Monitoring Committee
No

5. Study Description

Brief Summary
The purpose of the Micra Atrial TRacking Using A Ventricular AccELerometer 2 (MARVEL 2) study is to demonstrate safe and effective operation of the MARVEL 2 features for providing AV synchronous pacing in patients with normal sinus node function and AV block
Detailed Description
The MARVEL 2 study is an acute, prospective, global, multi-center, software-download clinical study. The study is planned to be conducted in US, Europe, and Asia. The study is expected to be conducted at approximately 15-20 centers in up to 100 subjects to obtain at least 70 usable Holter datasets to meet the objectives of the study. The expected total study duration (from first subject enrollment to the exit of the last subject) is approximately 6-months; this represents the time necessary to enroll the target sample size of at least 70 subjects with usable Holter datasets. Software will be downloaded into patients implanted with a Micra device to allow the new algorithm performance to be measured in patients where a Micra device has already been chosen as most appropriate for the patient. Most enrolled subjects will complete the study procedures during a single study visit. However, subjects enrolling in the study at the time of their Micra implant (anticipated to be approximately 10 subjects) will have the investigational algorithm downloaded following Micra implant, prior to hospital discharge, and approximately 1-month post-implant. This subset of subjects with de novo Micra implants will allow the MARVEL 2 features to be tested at multiple points in the device life cycle. Since the download algorithm running in a Micra device significantly increases current drain, a 2-4 hour acute study is used to limit the reduction in device longevity.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Atrioventricular Conduction Block

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Sequential Assignment
Model Description
Subjects enrolled in the MARVEL 2 study. Enrolled subjects will have the MARVEL 2 algorithm downloaded into their implanted market released Micra device. Each subject will be tested during when the MARVEL 2 algorithm is in monitor mode (VVI pacing) and in adaptive mode (VDD pacing).
Masking
None (Open Label)
Masking Description
Echocardiogram core lab will be masked to the individual patient subjects and MARVEL 2 algorithm mode.
Enrollment
77 (Actual)

8. Arms, Groups, and Interventions

Arm Title
MARVEL 2 Algorithm Monitor Mode, Then MARVEL 2 Adaptive Mode
Arm Type
Experimental
Arm Description
Participants first received MARVEL 2 algorithm monitor mode which provides standard VVI pacing for approximately 20 minutes followed by MARVEL 2 algorithm adaptive mode for approximately 2 hours which provides VDD pacing.
Intervention Type
Device
Intervention Name(s)
MARVEL 2 Algorithm Monitor Mode then Adaptive Mode
Intervention Description
Software download into implanted Micra device and programmed to MARVEL 2 Monitor Mode then MARVEL 2 Adaptive Mode
Primary Outcome Measure Information:
Title
Atrial Ventricular Pacing Synchrony Success During Rest
Description
A subject will meet the primary endpoint if a paced or sensed ventricular beat is within 300 ms following an ECG confirmed P-wave for at least 70% of the ECG confirmed P-waves. The primary endpoint will be evaluated during the MARVEL 2 setup phase during MARVEL 2 Algorithm monitor mode (VDI pacing which is effectively VVI pacing) and during the 20-minute resting period in which the MARVEL 2 Algorithm is programmed to Adaptive mode (VDD pacing)
Time Frame
40 Minutes
Title
Number of Participants Free From Inappropriate Pacemaker Function Events
Description
A participant will meet the primary endpoint if they are free from the following MARVEL 2 software related events during the entire Holter monitoring period where the MARVEL 2 features are enabled. Inappropriate pacemaker function events are defined as: Cardiac pauses lasting longer than 2 cardiac cycles (where cardiac cycle length is defined by the programmed lower rate interval of the pacemaker), OR Pacemaker oversensing induced tachycardia exceeding 3-minutes, defined as oversensing of the pacemaker's accelerometer signal leading to a heart rate exceeding 100 beats per minute.
Time Frame
Through study completion an average of one day
Secondary Outcome Measure Information:
Title
Left Ventricular Outflow Tract Velocity Time Integral
Description
The secondary endpoint is left ventricular outflow tract (LVOT) velocity time integral (VTI) as obtained from echocardiogram while the MARVEL 2 features are in adaptive mode (VDD pacing) and while the MARVEL 2 features are in monitor mode (VDI pacing which is effectively VVI pacing). This was measured by the echo core laboratory that was blinded to the study participant and MARVEL 2 algorithm mode.
Time Frame
12 cardiac cycles

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Subject has been implanted with a Micra TPS (Model MC1VR01) with remaining device longevity of 6 years or more or is expected to be implanted with a Micra TPS. Subject has history of AV block* Subject is ≥ 18 years old and as per required local law. Subject (and/or witness as applicable per local regulations) provides signed and dated authorization and/or consent per institution and local requirements. Subject is willing and able to comply with the protocol. *This includes subjects with normal sinus function and persistent 3rd degree AV block and subjects with other forms of AV block. Exclusion Criteria: Subject is currently enrolled or planning to participate in a potentially confounding drug or device trial during the study. Co-enrollment in concurrent trials is only allowed when documented pre-approval is obtained from the Medtronic Clinical Research Specialist. Subject is pregnant (if required by local law, women of child-bearing potential must undergo a pregnancy test within seven days prior to MARVEL 2 study procedures). Subject meets any exclusion criteria required by local law (age or other).
Facility Information:
Facility Name
Baptist Medical Center Jacksonville
City
Jacksonville
State/Province
Florida
ZIP/Postal Code
32207
Country
United States
Facility Name
Baptist Hospital
City
Miami
State/Province
Florida
ZIP/Postal Code
33176
Country
United States
Facility Name
North Shore University Hospital- Northwell
City
Manhasset
State/Province
New York
ZIP/Postal Code
11030
Country
United States
Facility Name
NYU Langone Medical Center
City
New York
State/Province
New York
ZIP/Postal Code
10016
Country
United States
Facility Name
Duke University Medical Center
City
Durham
State/Province
North Carolina
ZIP/Postal Code
27710
Country
United States
Facility Name
Kepler Universitätsklinikum Med Campus III
City
Linz
ZIP/Postal Code
4020
Country
Austria
Facility Name
UZ Leuven - Campus Gasthuisberg
City
Leuven
ZIP/Postal Code
3000
Country
Belgium
Facility Name
Odense Universitetshospital
City
Odense C
Country
Denmark
Facility Name
Hôpital Haut-Lévêque - CHU de Bordeaux
City
Pessac
ZIP/Postal Code
33604
Country
France
Facility Name
Prince of Wales Hospital
City
Hong Kong
Country
Hong Kong
Facility Name
Institut Jantung Negara
City
Kuala Lumpur
ZIP/Postal Code
50400
Country
Malaysia
Facility Name
Hospital Universitari Clínic de Barcelona
City
Barcelona
ZIP/Postal Code
08036
Country
Spain

12. IPD Sharing Statement

Citations:
PubMed Identifier
33928713
Citation
Garweg C, Khelae SK, Chan JYS, Chinitz L, Ritter P, Johansen JB, Sagi V, Epstein LM, Piccini JP, Pascual M, Mont L, Willems R, Splett V, Stromberg K, Sheldon T, Kristiansen N, Steinwender C. Behavior of AV synchrony pacing mode in a leadless pacemaker during variable AV conduction and arrhythmias. J Cardiovasc Electrophysiol. 2021 Jul;32(7):1947-1957. doi: 10.1111/jce.15061. Epub 2021 May 20.
Results Reference
derived
PubMed Identifier
32717315
Citation
Garweg C, Khelae SK, Steinwender C, Chan JYS, Ritter P, Johansen JB, Sagi V, Epstein LM, Piccini JP, Pascual M, Mont L, Willems R, Sheldon T, Splett V, Stromberg K, Wood N, Chinitz L. Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study. Heart Rhythm. 2020 Dec;17(12):2037-2045. doi: 10.1016/j.hrthm.2020.07.024. Epub 2020 Jul 24.
Results Reference
derived
PubMed Identifier
31709982
Citation
Steinwender C, Khelae SK, Garweg C, Chan JYS, Ritter P, Johansen JB, Sagi V, Epstein LM, Piccini JP, Pascual M, Mont L, Sheldon T, Splett V, Stromberg K, Wood N, Chinitz L. Atrioventricular Synchronous Pacing Using a Leadless Ventricular Pacemaker: Results From the MARVEL 2 Study. JACC Clin Electrophysiol. 2020 Jan;6(1):94-106. doi: 10.1016/j.jacep.2019.10.017. Epub 2019 Nov 11.
Results Reference
derived

Learn more about this trial