Observation of ImageReady™ MR Conditional Defibrillation System in China
Primary Purpose
Arrhythmia
Status
Completed
Phase
Not Applicable
Locations
China
Study Type
Interventional
Intervention
ImageReady™ MR Conditional Defibrillation System
Sponsored by
About this trial
This is an interventional treatment trial for Arrhythmia
Eligibility Criteria
Inclusion Criteria:
- Subject is indicated to Class I/II indications per guidelines/consensus released by Chinese Society of Cardiac Pacing and Electrophysiology
- Subject must have the ImageReady System as their initial (de novo) defibrillation system implant
- Subject will receive an ICD or CRT-D pulse generator in the left or right pectoral region ICD
- Subject is able and willing to undergo an MR scan
- Subject is willing and capable of providing informed consent and participating in all testing/ visits associated with this clinical study at an approved clinical study center and at the intervals defined by this protocol
- Subject is age 18 or above
Exclusion Criteria:
- Subject has other active or abandoned implanted cardiac rhythm devices, components or accessories present such as pulse generators, leads, lead adaptors or extenders
- Presence of metallic objects that represent a contraindication to MR imaging at the discretion of the Radiologist and impacting the ability to conduct the study protocol
- Subject needs or will need a medically necessary MR scan, before completing the 1-month post-MR follow-up visit
Subject with:
- A history of syncope related to brady-arrhythmia
- A history of syncope of unknown etiology
- Sinus pauses (Pause > 2 s)
- Permanent or intermittent complete AV block
- Documentation of progressive AV nodal block over time
Trifascicular block (alternating bundle branch block or PR > 200 ms with LBBB or other bifascicular block)
- Note: It is required to run a 12 lead ECG and a 10s rhythm strip to document this exclusion criterion.
- Subject is not clinically capable of tolerating the absence of pacing or Resynchronization therapy support in a supine position for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
- Subject is not clinically capable of tolerating the absence of Tachycardia therapy support for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
- Subjects currently requiring dialysis
- Subject has a mechanical heart valve
- Subject has a known or suspected sensitivity to dexamethasone acetate (DXA)
- Subject is currently on the active heart transplant list
- Subject has documented life expectancy of less than 12 months
- Subject is enrolled in any other concurrent study that might interfere with this study
- Women of childbearing potential who are or might be pregnant at the time of this study
Sites / Locations
- Zhongshan Hospital Fudan University
Arms of the Study
Arm 1
Arm Type
Experimental
Arm Label
ImageReady™ MR Conditional Defibrillation System
Arm Description
Subject is indicated to Class I/II indications per guidelines/consensus released by Chinese Society of Cardiac Pacing and Electrophysiology Subject must have the ImageReady System as their initial (de novo) defibrillation system implant
Outcomes
Primary Outcome Measures
The Percent of Participants Who Were MR Scan-Related ImageReady System Complication -Free
The primary safety endpoint of the MR ICD study will be assessed for all subjects who undergo any portion of the study-required MR scan sequences. Safety will be confirmed by evaluating the MR scan related ImageReady System complication-free rate (CFR) between the MR Scan and the MRI Visit + 1 Month.MR scan related ImageReady System complication is relation with ImageReady System and MR Scan.
The Percentage of Participant Whose the Average RV Shocking Impedance is >200 Ohm 1 Month Post Scan, While it is ≤ 200 Ohm Before Scan
The normal RV shocking impedance measured by the system should be≤200 Ohm, with >200 Ohm considered abnormal. The primary effectiveness endpoint 1 is defined as that the average RV shocking impedance is >200 Ohm at 1 month post scan, while it is ≤ 200 Ohm before scan.There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 1 occur.
The Percentage of Participant That Have an Increase in Average RV Pacing Thresholds > 0.5V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up
Subjects that have an increase in average RV pacing thresholds ≤ 0.5V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 2 event.The performance goal of this endpoint for the ENABLE MRI study was 87%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 2 occur.
The Percentage of Participant That Have the Average RV Sensed Amplitude at the MRI + 1 Month Visit Remains < 5.0 mV or Less Than 50% of the Pre-MR Scan Value
Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 3 event.The performance goal of this endpoint for the ENABLE MRI study was 85%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 3 occur.
The Percentage of Participant Have an Increase in Average LV Pacing Thresholds >1.0V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up
Subjects that have an increase in average LV pacing thresholds ≤1.0V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 4 event.The performance goal of this endpoint for the ENABLE MRI study was 87%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 4 occur.
the Percentage of Participant That Have the Average LV Sensed Amplitude at the MRI + 1 Month Visit Remains <5.0 mV or Above 50% of the Pre-MR Scan Value
Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 5 event.The performance goal of this endpoint for the ENABLE MRI study was 85%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 5 occur.
Secondary Outcome Measures
Full Information
NCT ID
NCT03451721
First Posted
September 4, 2017
Last Updated
October 20, 2021
Sponsor
Boston Scientific Corporation
1. Study Identification
Unique Protocol Identification Number
NCT03451721
Brief Title
Observation of ImageReady™ MR Conditional Defibrillation System in China
Official Title
Observation of ImageReady™ MR Conditional Defibrillation System in China (MR ICD)
Study Type
Interventional
2. Study Status
Record Verification Date
September 2019
Overall Recruitment Status
Completed
Study Start Date
April 3, 2018 (Actual)
Primary Completion Date
March 12, 2019 (Actual)
Study Completion Date
December 10, 2019 (Actual)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Sponsor
Name of the Sponsor
Boston Scientific Corporation
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Product Manufactured in and Exported from the U.S.
No
Data Monitoring Committee
No
5. Study Description
Brief Summary
To observe the safety and effectiveness of ImageReady™ MR conditional defibrillation system in a Chinese population.
Detailed Description
This is a multi-center, prospective, single-arm study, aimed to observe the safety and effectiveness of ImageReady™ MR conditional defibrillation system in a Chinese population.It will enroll 20 subjects in 5 sites in China.The primary safety endpoint is MR scan-related Complication-free rate between the MR Scan and the MRI + 1 Month Visit. The primary effectiveness endpoints include:1.Abnormal increase in RV shocking impedance from the pre-MR scan to the 1 Month post-MR scan. 2.Increase in RV pacing threshold from the pre-MR scan to the 1 Month post-MR scan. 3.Decrease in RV sensed amplitude from the pre-MR scan to the 1 Month post-MR scan. 4.Increase in LV pacing threshold from the pre-MR scan to the 1 Month post-MR scan. 5.Decrease in LV sensed amplitude from the pre-MR scan to the 1 Month post- MR scan.
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Arrhythmia
7. Study Design
Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
20 (Actual)
8. Arms, Groups, and Interventions
Arm Title
ImageReady™ MR Conditional Defibrillation System
Arm Type
Experimental
Arm Description
Subject is indicated to Class I/II indications per guidelines/consensus released by Chinese Society of Cardiac Pacing and Electrophysiology
Subject must have the ImageReady System as their initial (de novo) defibrillation system implant
Intervention Type
Device
Intervention Name(s)
ImageReady™ MR Conditional Defibrillation System
Intervention Description
Subject will receive an ICD or CRT-D pulse generator in the left or right pectoral region ICD
Primary Outcome Measure Information:
Title
The Percent of Participants Who Were MR Scan-Related ImageReady System Complication -Free
Description
The primary safety endpoint of the MR ICD study will be assessed for all subjects who undergo any portion of the study-required MR scan sequences. Safety will be confirmed by evaluating the MR scan related ImageReady System complication-free rate (CFR) between the MR Scan and the MRI Visit + 1 Month.MR scan related ImageReady System complication is relation with ImageReady System and MR Scan.
Time Frame
MRI + 1 Month Visit(10-13 weeks from 0 day)
Title
The Percentage of Participant Whose the Average RV Shocking Impedance is >200 Ohm 1 Month Post Scan, While it is ≤ 200 Ohm Before Scan
Description
The normal RV shocking impedance measured by the system should be≤200 Ohm, with >200 Ohm considered abnormal. The primary effectiveness endpoint 1 is defined as that the average RV shocking impedance is >200 Ohm at 1 month post scan, while it is ≤ 200 Ohm before scan.There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 1 occur.
Time Frame
MRI+1 Month visit ( 10~13 weeks from 0 day)
Title
The Percentage of Participant That Have an Increase in Average RV Pacing Thresholds > 0.5V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up
Description
Subjects that have an increase in average RV pacing thresholds ≤ 0.5V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 2 event.The performance goal of this endpoint for the ENABLE MRI study was 87%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 2 occur.
Time Frame
MRI+1Month visit( 10~14 weeks from 0 day)
Title
The Percentage of Participant That Have the Average RV Sensed Amplitude at the MRI + 1 Month Visit Remains < 5.0 mV or Less Than 50% of the Pre-MR Scan Value
Description
Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 3 event.The performance goal of this endpoint for the ENABLE MRI study was 85%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 3 occur.
Time Frame
MR+1 Month visit( 10~14 weeks from 0 day)
Title
The Percentage of Participant Have an Increase in Average LV Pacing Thresholds >1.0V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up
Description
Subjects that have an increase in average LV pacing thresholds ≤1.0V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 4 event.The performance goal of this endpoint for the ENABLE MRI study was 87%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 4 occur.
Time Frame
MR +1 Month visit( 10~14 weeks from 0 day)
Title
the Percentage of Participant That Have the Average LV Sensed Amplitude at the MRI + 1 Month Visit Remains <5.0 mV or Above 50% of the Pre-MR Scan Value
Description
Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 5 event.The performance goal of this endpoint for the ENABLE MRI study was 85%. There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 5 occur.
Time Frame
MR +1 Month visit( 10~14 weeks from 0 day)
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
Subject is indicated to Class I/II indications per guidelines/consensus released by Chinese Society of Cardiac Pacing and Electrophysiology
Subject must have the ImageReady System as their initial (de novo) defibrillation system implant
Subject will receive an ICD or CRT-D pulse generator in the left or right pectoral region ICD
Subject is able and willing to undergo an MR scan
Subject is willing and capable of providing informed consent and participating in all testing/ visits associated with this clinical study at an approved clinical study center and at the intervals defined by this protocol
Subject is age 18 or above
Exclusion Criteria:
Subject has other active or abandoned implanted cardiac rhythm devices, components or accessories present such as pulse generators, leads, lead adaptors or extenders
Presence of metallic objects that represent a contraindication to MR imaging at the discretion of the Radiologist and impacting the ability to conduct the study protocol
Subject needs or will need a medically necessary MR scan, before completing the 1-month post-MR follow-up visit
Subject with:
A history of syncope related to brady-arrhythmia
A history of syncope of unknown etiology
Sinus pauses (Pause > 2 s)
Permanent or intermittent complete AV block
Documentation of progressive AV nodal block over time
Trifascicular block (alternating bundle branch block or PR > 200 ms with LBBB or other bifascicular block)
Note: It is required to run a 12 lead ECG and a 10s rhythm strip to document this exclusion criterion.
Subject is not clinically capable of tolerating the absence of pacing or Resynchronization therapy support in a supine position for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
Subject is not clinically capable of tolerating the absence of Tachycardia therapy support for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
Subjects currently requiring dialysis
Subject has a mechanical heart valve
Subject has a known or suspected sensitivity to dexamethasone acetate (DXA)
Subject is currently on the active heart transplant list
Subject has documented life expectancy of less than 12 months
Subject is enrolled in any other concurrent study that might interfere with this study
Women of childbearing potential who are or might be pregnant at the time of this study
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Yangang Su, Doctor
Organizational Affiliation
Fudan University
Official's Role
Principal Investigator
Facility Information:
Facility Name
Zhongshan Hospital Fudan University
City
Shanghai
State/Province
Shanghai
ZIP/Postal Code
200032
Country
China
12. IPD Sharing Statement
Plan to Share IPD
No
Citations:
PubMed Identifier
21969340
Citation
Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011 Oct 4;155(7):415-24. doi: 10.7326/0003-4819-155-7-201110040-00004.
Results Reference
background
PubMed Identifier
28225684
Citation
Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, Boyle NG, Frabizzio JV, Birgersdotter-Green U, Higgins SL, Lampert R, Machado CE, Martin ET, Rivard AL, Rubenstein JC, Schaerf RH, Schwartz JD, Shah DJ, Tomassoni GF, Tominaga GT, Tonkin AE, Uretsky S, Wolff SD. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. N Engl J Med. 2017 Feb 23;376(8):755-764. doi: 10.1056/NEJMoa1603265.
Results Reference
background
Citation
ASTM Standard F2503-08, "Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment" ASTM International, West Conshohocken, PA, 2008, DOI: 10.1520/F2503-08, www.astm.org..
Results Reference
background
Citation
Chinese Medical Association ECG Physiology and Pacing Branch, Chinese Medical Association Cardiology Branch, Chinese Medical Association Cardiology Professional Committee implanted cardioverter defibrillator treatment. Implanted cardioverter defibrillator treatment of China Expert consensus. Chinese Journal of Cardiac Arrhythmias 2014;18:242-53.
Results Reference
background
Citation
Shu Zhang, Dejia Huang, Wei Hua, et al. Advice on cardiac resynchronization therapy for chronic heart failure (revised in 2013). Chinese Journal of Cardiac Arrhythmias 2013;17:247-61.
Results Reference
background
PubMed Identifier
18547191
Citation
Wolf SM, Lawrenz FP, Nelson CA, Kahn JP, Cho MK, Clayton EW, Fletcher JG, Georgieff MK, Hammerschmidt D, Hudson K, Illes J, Kapur V, Keane MA, Koenig BA, Leroy BS, McFarland EG, Paradise J, Parker LS, Terry SF, Van Ness B, Wilfond BS. Managing incidental findings in human subjects research: analysis and recommendations. J Law Med Ethics. 2008 Summer;36(2):219-48, 211. doi: 10.1111/j.1748-720X.2008.00266.x.
Results Reference
background
PubMed Identifier
17515363
Citation
Kanal E, Barkovich AJ, Bell C, Borgstede JP, Bradley WG Jr, Froelich JW, Gilk T, Gimbel JR, Gosbee J, Kuhni-Kaminski E, Lester JW Jr, Nyenhuis J, Parag Y, Schaefer DJ, Sebek-Scoumis EA, Weinreb J, Zaremba LA, Wilcox P, Lucey L, Sass N; ACR Blue Ribbon Panel on MR Safety. ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol. 2007 Jun;188(6):1447-74. doi: 10.2214/AJR.06.1616. No abstract available.
Results Reference
background
Learn more about this trial
Observation of ImageReady™ MR Conditional Defibrillation System in China
We'll reach out to this number within 24 hrs