Optimizing Closed-Loop Control of Type 1 Diabetes Mellitus in Adolescents

Primary Purpose

Diabetes Mellitus, Type 1

Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
Diabetes Assistant (DiAs)
Sponsored by
University of Virginia
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Diabetes Mellitus, Type 1 focused on measuring Type 1 Diabetes Mellitus (T1DM), Closed Loop Control (CLC), Diabetes Assistant (DiAs), Continuous Glucose Monitor (CGM)

Eligibility Criteria

13 Years - 18 Years (Child, Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  1. ≥13 and ≤18 years old
  2. Clinical diagnosis of type 1 diabetes mellitus for at least 2 years as noted by the following:

    Criteria for documented hyperglycemia (at least 1 criterion must be met):

    • Fasting glucose ≥126 mg/dL - confirmed
    • Two-hour Oral Glucose Tolerance Test (OGTT) ≥200 mg/dL - confirmed
    • Hemoglobin A1c (HbA1c) ≥6.5% documented by history - confirmed
    • Random glucose ≥200 mg/dL with symptoms
    • No data at diagnosis is available but the participant has a convincing history of hyperglycemia consistent with diabetes

    Criteria for requiring insulin at diagnosis (at least 1 criterion must be met):

    • Participant required insulin at diagnosis and continually thereafter
    • Participant did not start insulin at diagnosis but upon investigator review likely needed insulin (significant hyperglycemia that did not respond to oral agents) and did require insulin eventually and used continually
    • Participant did not start insulin at diagnosis but continued to be hyperglycemic, had positive islet cell antibodies - consistent with Latent Autoimmune Diabetes in Adults (LADA) and did require insulin eventually and used continually
  3. Use of an insulin pump to treat his/her diabetes for at least 6 months
  4. Actively using a carbohydrate (CHO) /insulin ratio for insulin bolus adjustments in order to keep blood glucose in a predefined range
  5. Tanner stage II or greater based on physician exam
  6. HbA1c between <10.5% as measured with DCA2000 or equivalent device
  7. Not currently known to be pregnant, breast feeding, or intending to become pregnant (females)
  8. Demonstration of proper mental status and cognition for the study
  9. Willingness to avoid consumption of acetaminophen-containing products during the study visits involving DexCom use
  10. Self-reported behavior of snacking without insulin coverage or under bolus for meals
  11. Willingness to consume a 30 grams snack without insulin coverage and willingness to bolus 75 % of usual treatment for an 80 grams lunch during both admissions
  12. Willingness to remove home CGM for admissions if the subject typically wears a CGM
  13. Medication stability in the preceding two months if taking antihypertensive, thyroid, anti-depressant or lipid lowering medication
  14. Reported history of missed or inaccurate bolus treatments at meal time

Exclusion Criteria:

  1. Diabetic ketoacidosis (DKA) within the 6 months prior to enrollment
  2. In adherence with the One Touch Ultra 2 User Guide, subjects with hematocrit levels below 30% or above 55% will be excluded. Severe hypoglycemia resulting in seizure or loss of consciousness in the 12 months prior to enrollment
  3. Pregnancy; breast feeding, or intention of becoming pregnant
  4. Uncontrolled arterial hypertension (based on resting blood pressure > 95 percentile as listed according to age and height percentile in the Harriet Lane Handbook of Pediatrics)
  5. Conditions which may increase the risk of hypoglycemia such as uncontrolled coronary artery disease during the previous year (e.g. history of myocardial infarction, acute coronary syndrome, therapeutic coronary intervention, coronary bypass or stenting procedure, stable or unstable angina, episode of chest pain of cardiac etiology with documented electrocardiogram (EKG) changes, or positive stress test or catheterization with coronary blockages >50%), congestive heart failure, history of cerebrovascular event, seizure disorder, syncope, adrenal insufficiency, neurologic disease or atrial fibrillation
  6. History of a systemic or deep tissue infection with methicillin-resistant staph aureus or Candida albicans
  7. History of arrythmia
  8. Use of a device that may pose electromagnetic compatibility issues and/or radiofrequency interference with the DexCom CGM (implantable cardioverter-defibrillator, electronic pacemaker, neurostimulator, intrathecal pump, and cochlear implants)
  9. Anticoagulant therapy other than aspirin
  10. Oral steroids
  11. Medical condition requiring use of an acetaminophen-containing medication that cannot be withheld for for 24 hours before CGM insertions until the end of each the study admissions.
  12. Psychiatric disorders that would interfere with study tasks (e.g. inpatient psychiatric treatment within 6 months prior to enrollment)
  13. Mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation
  14. Known current or recent alcohol or drug abuse
  15. Medical conditions that would make operating a CGM, the DiAs cell phone, or insulin pump difficult (e.g. blindness, severe arthritis, immobility)
  16. Any skin condition that prevents sensor or pump placement on the abdomen or arm (e.g. bad sunburn, pre-existing dermatitis, intertrigo, psoriasis, extensive scarring, cellulitis)
  17. Impaired hepatic function measured as alanine aminotransferase or aspartate aminotransferase ≥ three times the upper reference limit
  18. Impaired renal function measured as creatinine >1.2 times above the upper limit of normal
  19. Uncontrolled microvascular (diabetic) complications (other than diabetic non-proliferative retinopathy), such as history of laser coagulation, proliferative diabetic retinopathy, known diabetic nephropathy (other than microalbuminuria with normal creatinine) or neuropathy requiring treatment
  20. Active gastroparesis requiring current medical therapy
  21. Uncontrolled thyroid disease
  22. Known bleeding diathesis or dyscrasia
  23. Known allergy to medical adhesives, components of the insulin pump insertion set or continuous glucose monitor sensor
  24. Active enrollment in another clinical trial
  25. Use of anti-diabetic agents other than continuous subcutaneous insulin infusion (CSII) including long-acting insulin, intermediate-acting insulin, metformin, sulfonylureas, meglitinides, thiazolidinediones, dipeptidyl peptidase 4 (DPP-4) inhibitors, glucagon-like peptide 1 agonists and alpha-glucosidase inhibitors
  26. Subjects with basal rates less than 0.01 units/hour

RESTRICTIONS ON USE OF OTHER DRUGS OR TREATMENTS

  1. Use of anti-diabetic agents other than CSII including long-acting insulin, intermediate-acting insulin, metformin, sulfonylureas, meglitinides, thiazolidinediones, DPP-4, inhibitors, glucagon-like peptide 1 agonists, and alpha-glucosidase inhibitors.
  2. Acetaminophen will be restricted 24 hours before the insertion of not be allowed while the continuous glucose monitor is in use.
  3. Medications that block symptoms of hypoglycemia, including but not limited to beta blockers.
  4. Subjects on amylin analogs will be asked to withhold the medication during the outpatient admissions.

Sites / Locations

  • University of Virginia Center for Diabetes Technology

Arms of the Study

Arm 1

Arm 2

Arm Type

Active Comparator

No Intervention

Arm Label

DiAs-closed-loop system

Sensor Augmented Pump

Arm Description

eight hours using Closed Loop Control (DiAs) under both, miss insulin bolus for 30 grams of carbohydrates snack or under bolus for an 80 grams of carbohydrates lunch

8 hours observational under both, missed insulin bolus for 30 gr carbohydrates snack and under bolus for an 80 gr carbohydrates lunch. Sensor augmented pump used the patients own insulin settings (basal rate, meal boluses and correction boluses) and a CGM provided by the study team to be used during admission. No DiAs use during this admission.

Outcomes

Primary Outcome Measures

Percent of Time Spent Near Normoglycemia
Percentage of time that blood glucose (BG) values (measured with both finger-stick and CGM) were near normoglycemia (70-180 mg/dL).

Secondary Outcome Measures

Full Information

First Posted
June 27, 2013
Last Updated
August 23, 2022
Sponsor
University of Virginia
search

1. Study Identification

Unique Protocol Identification Number
NCT01890954
Brief Title
Optimizing Closed-Loop Control of Type 1 Diabetes Mellitus in Adolescents
Official Title
Optimizing Closed-Loop Control of Type 1 Diabetes Mellitus in Adolescents
Study Type
Interventional

2. Study Status

Record Verification Date
August 2022
Overall Recruitment Status
Completed
Study Start Date
August 2013 (undefined)
Primary Completion Date
February 2014 (Actual)
Study Completion Date
February 2014 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
University of Virginia

4. Oversight

Data Monitoring Committee
No

5. Study Description

Brief Summary
The purpose of this study is to use a closed-loop Control-to-Range (CTR) system in adolescents with type 1 diabetes in an outpatient setting, and to evaluate the CTR system's ability to significantly improve blood glucose levels when an insulin bolus is omitted for a 30 gram carbohydrate snack and when insulin bolus is insufficient for the amount of carbohydrates consumed for a meal. The primary objective of this study is to use a closed-loop Control-to-Range (CTR) system to significantly reduce the post-prandial blood glucose excursion in adolescents with type 1 diabetes who omit and/or under-bolus insulin for either snacks or meals. Up to 20 subjects aged ≥13 and ≤18 years old will be tested.
Detailed Description
The purpose of this study is to use a closed-loop Control-to-Range (CTR) system in adolescents with type 1 diabetes in an outpatient setting, and to evaluate the CTR system's ability to significantly improve blood glucose levels when an insulin bolus is omitted for a 30 gram carbohydrate snack and when insulin bolus is insufficient for the amount of carbohydrates consumed for a meal. The CTR system is comprised of two algorithmic layers: (i) A Safety Supervision Module (SSM) which contains a predictive insulin request dampener (or brakes); (ii) a Range Correction Module (RCM), consisting in (a) a Hyperglycemia Mitigation System, and (b) Insulin on Board controller. Both modules will receive continuous glucose monitoring (CGM) and historical insulin delivery data. The SSM will monitor the safety of the subject's continuous subcutaneous insulin infusion pump (CSII) to prevent hypoglycemia. The RCM will be responsible for optimizing blood glucose (BG) control and mitigating postprandial hyperglycemic excursions through a mix of increased basal rate and, potentially, isolated insulin boluses. To run CTR, we will use our wearable artificial pancreas platform, known as DiAs (Diabetes Assistant) which consists of a smart phone running CTR and connected to standard insulin delivery and CGM devices.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Diabetes Mellitus, Type 1
Keywords
Type 1 Diabetes Mellitus (T1DM), Closed Loop Control (CLC), Diabetes Assistant (DiAs), Continuous Glucose Monitor (CGM)

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Crossover Assignment
Masking
None (Open Label)
Allocation
Randomized
Enrollment
17 (Actual)

8. Arms, Groups, and Interventions

Arm Title
DiAs-closed-loop system
Arm Type
Active Comparator
Arm Description
eight hours using Closed Loop Control (DiAs) under both, miss insulin bolus for 30 grams of carbohydrates snack or under bolus for an 80 grams of carbohydrates lunch
Arm Title
Sensor Augmented Pump
Arm Type
No Intervention
Arm Description
8 hours observational under both, missed insulin bolus for 30 gr carbohydrates snack and under bolus for an 80 gr carbohydrates lunch. Sensor augmented pump used the patients own insulin settings (basal rate, meal boluses and correction boluses) and a CGM provided by the study team to be used during admission. No DiAs use during this admission.
Intervention Type
Device
Intervention Name(s)
Diabetes Assistant (DiAs)
Other Intervention Name(s)
Closed-Loop system
Intervention Description
Diabetes Assistance (DiAs) is a software residing in a Smartphone that contains the algorithms to regulate and control insulin deliveries (insulin bolus for: Basal rate, meal insulin and correction bolus) with inputs glucose values from a CGM and outputs insulin infusion by an insulin pump
Primary Outcome Measure Information:
Title
Percent of Time Spent Near Normoglycemia
Description
Percentage of time that blood glucose (BG) values (measured with both finger-stick and CGM) were near normoglycemia (70-180 mg/dL).
Time Frame
8 hours

10. Eligibility

Sex
All
Minimum Age & Unit of Time
13 Years
Maximum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: ≥13 and ≤18 years old Clinical diagnosis of type 1 diabetes mellitus for at least 2 years as noted by the following: Criteria for documented hyperglycemia (at least 1 criterion must be met): Fasting glucose ≥126 mg/dL - confirmed Two-hour Oral Glucose Tolerance Test (OGTT) ≥200 mg/dL - confirmed Hemoglobin A1c (HbA1c) ≥6.5% documented by history - confirmed Random glucose ≥200 mg/dL with symptoms No data at diagnosis is available but the participant has a convincing history of hyperglycemia consistent with diabetes Criteria for requiring insulin at diagnosis (at least 1 criterion must be met): Participant required insulin at diagnosis and continually thereafter Participant did not start insulin at diagnosis but upon investigator review likely needed insulin (significant hyperglycemia that did not respond to oral agents) and did require insulin eventually and used continually Participant did not start insulin at diagnosis but continued to be hyperglycemic, had positive islet cell antibodies - consistent with Latent Autoimmune Diabetes in Adults (LADA) and did require insulin eventually and used continually Use of an insulin pump to treat his/her diabetes for at least 6 months Actively using a carbohydrate (CHO) /insulin ratio for insulin bolus adjustments in order to keep blood glucose in a predefined range Tanner stage II or greater based on physician exam HbA1c between <10.5% as measured with DCA2000 or equivalent device Not currently known to be pregnant, breast feeding, or intending to become pregnant (females) Demonstration of proper mental status and cognition for the study Willingness to avoid consumption of acetaminophen-containing products during the study visits involving DexCom use Self-reported behavior of snacking without insulin coverage or under bolus for meals Willingness to consume a 30 grams snack without insulin coverage and willingness to bolus 75 % of usual treatment for an 80 grams lunch during both admissions Willingness to remove home CGM for admissions if the subject typically wears a CGM Medication stability in the preceding two months if taking antihypertensive, thyroid, anti-depressant or lipid lowering medication Reported history of missed or inaccurate bolus treatments at meal time Exclusion Criteria: Diabetic ketoacidosis (DKA) within the 6 months prior to enrollment In adherence with the One Touch Ultra 2 User Guide, subjects with hematocrit levels below 30% or above 55% will be excluded. Severe hypoglycemia resulting in seizure or loss of consciousness in the 12 months prior to enrollment Pregnancy; breast feeding, or intention of becoming pregnant Uncontrolled arterial hypertension (based on resting blood pressure > 95 percentile as listed according to age and height percentile in the Harriet Lane Handbook of Pediatrics) Conditions which may increase the risk of hypoglycemia such as uncontrolled coronary artery disease during the previous year (e.g. history of myocardial infarction, acute coronary syndrome, therapeutic coronary intervention, coronary bypass or stenting procedure, stable or unstable angina, episode of chest pain of cardiac etiology with documented electrocardiogram (EKG) changes, or positive stress test or catheterization with coronary blockages >50%), congestive heart failure, history of cerebrovascular event, seizure disorder, syncope, adrenal insufficiency, neurologic disease or atrial fibrillation History of a systemic or deep tissue infection with methicillin-resistant staph aureus or Candida albicans History of arrythmia Use of a device that may pose electromagnetic compatibility issues and/or radiofrequency interference with the DexCom CGM (implantable cardioverter-defibrillator, electronic pacemaker, neurostimulator, intrathecal pump, and cochlear implants) Anticoagulant therapy other than aspirin Oral steroids Medical condition requiring use of an acetaminophen-containing medication that cannot be withheld for for 24 hours before CGM insertions until the end of each the study admissions. Psychiatric disorders that would interfere with study tasks (e.g. inpatient psychiatric treatment within 6 months prior to enrollment) Mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation Known current or recent alcohol or drug abuse Medical conditions that would make operating a CGM, the DiAs cell phone, or insulin pump difficult (e.g. blindness, severe arthritis, immobility) Any skin condition that prevents sensor or pump placement on the abdomen or arm (e.g. bad sunburn, pre-existing dermatitis, intertrigo, psoriasis, extensive scarring, cellulitis) Impaired hepatic function measured as alanine aminotransferase or aspartate aminotransferase ≥ three times the upper reference limit Impaired renal function measured as creatinine >1.2 times above the upper limit of normal Uncontrolled microvascular (diabetic) complications (other than diabetic non-proliferative retinopathy), such as history of laser coagulation, proliferative diabetic retinopathy, known diabetic nephropathy (other than microalbuminuria with normal creatinine) or neuropathy requiring treatment Active gastroparesis requiring current medical therapy Uncontrolled thyroid disease Known bleeding diathesis or dyscrasia Known allergy to medical adhesives, components of the insulin pump insertion set or continuous glucose monitor sensor Active enrollment in another clinical trial Use of anti-diabetic agents other than continuous subcutaneous insulin infusion (CSII) including long-acting insulin, intermediate-acting insulin, metformin, sulfonylureas, meglitinides, thiazolidinediones, dipeptidyl peptidase 4 (DPP-4) inhibitors, glucagon-like peptide 1 agonists and alpha-glucosidase inhibitors Subjects with basal rates less than 0.01 units/hour RESTRICTIONS ON USE OF OTHER DRUGS OR TREATMENTS Use of anti-diabetic agents other than CSII including long-acting insulin, intermediate-acting insulin, metformin, sulfonylureas, meglitinides, thiazolidinediones, DPP-4, inhibitors, glucagon-like peptide 1 agonists, and alpha-glucosidase inhibitors. Acetaminophen will be restricted 24 hours before the insertion of not be allowed while the continuous glucose monitor is in use. Medications that block symptoms of hypoglycemia, including but not limited to beta blockers. Subjects on amylin analogs will be asked to withhold the medication during the outpatient admissions.
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Daniel R. Chernavvsky, MD, CRC
Organizational Affiliation
University of Virginia
Official's Role
Principal Investigator
First Name & Middle Initial & Last Name & Degree
Mark DeBoer, MD
Organizational Affiliation
University of Virginia
Official's Role
Principal Investigator
Facility Information:
Facility Name
University of Virginia Center for Diabetes Technology
City
Charlottesville
State/Province
Virginia
ZIP/Postal Code
22904
Country
United States

12. IPD Sharing Statement

Plan to Share IPD
Yes
IPD Sharing Plan Description
We will make de-identified individual participant data (IPD) collected in this study available to other researchers at the end of the study. Paper published in Pediatric Diabetes
Citations:
PubMed Identifier
25348683
Citation
Chernavvsky DR, DeBoer MD, Keith-Hynes P, Mize B, McElwee M, Demartini S, Dunsmore SF, Wakeman C, Kovatchev BP, Breton MD. Use of an artificial pancreas among adolescents for a missed snack bolus and an underestimated meal bolus. Pediatr Diabetes. 2016 Feb;17(1):28-35. doi: 10.1111/pedi.12230. Epub 2014 Oct 27.
Results Reference