Quartet Lead With Defibrillator Multisite Algorithmic Cardiac Resynchronisation Therapy Optimisation (QUARTRMASTER)

Primary Purpose

Heart Failure, Systolic

Status
Withdrawn
Phase
Not Applicable
Locations
United Kingdom
Study Type
Interventional
Intervention
Algorithmic Cardiac Resynchronisation Therapy Optimisation
Sponsored by
Cardiff and Vale University Health Board
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Heart Failure, Systolic focused on measuring Cardiac resynchronization pacing therapy

Eligibility Criteria

18 Years - undefined (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • Successful implantation of CRT-D device with quadripolar left ventricular lead and dual coil right ventricular lead.

Exclusion Criteria:

  • failed device implantation
  • inability to complete the follow-up process
  • inability to provide full written consent. Patients unable to perform cardiopulmonary exercise testing will be excluded from this assessment only and will be able to complete the remaining assessments.

Sites / Locations

  • University Hospital of Wales

Arms of the Study

Arm 1

Arm Type

Experimental

Arm Label

Main patient cohort

Arm Description

Algorithmic Cardiac Resynchronisation Therapy Optimisation; Patients requiring CRT-D enrolled to receive algorithmic CRT optimisation at their device implantation. They will then be programmed to standard CRT settings for 12 weeks. Assessments will be performed and they will again undergo CRT optimisation using the specified algorithm. This time, they will be programmed to the settings giving maximal cardiac output on echocardiography, as assessed by LVOT VTI. After a further 12 weeks, they will again undergo assessment and the study will terminate.

Outcomes

Primary Outcome Measures

Change in left ventricular end-systolic volume in patients undergoing algorithmic optimisation of CRT-D compared with usual device settings
End-systolic volume will be measured on echocardiography using Simpson's method. Where image quality precludes a full biplane assessment, single plane will be used with the apical 4-chamber view, or alternatively left ventricular end-systolic diameter will be used to calculate volume using the Teicholz method.

Secondary Outcome Measures

Feasibility of algorithmic, echocardiography-guided CRT-D optimisation with quadripolar left ventricular leads in clinical practice
To see whether our optimisation algorithm is acceptable to patients and easy to perform in clinical practice.
Correlation between invasive haemodynamic studies with non-invasive ultrasound-based methods in optimising CRT-D
Attempt to establish a gold standard for optimisation of CRT devices by comparing invasive haemodynamic monitoring, echocardiographic parameters and cardiac output by ultrasound cardiac output monitoring (USCOM)
Change in left ventricular end diastolic pressure with algorithmic optimisation of CRT-D compared with usual device settings
A measurement of left ventricular end-diastolic pressure will be made via invasive left ventricular monitoring after programming with usual settings and again after programming using algorithmic optimisation of the CRT-D device.
Feasibility of ultrasound cardiac output monitoring (USCOM) in optimisation of CRT devices in clinical practice
Use of USCOM has been documented in optimisation of CRT, but not in such rapidly-changing algorithmic optimisation. We plan to assess whether it is feasible to use this in clinical practice as a replacement for a full echocardiogram.
Change in peak oxygen consumption with algorithmic CRT-D optimisation compared with usual device settings
Assessment of peak oxygen consumption as measured by cardiopulmonary exercise testing after usual settings compared with settings after algorithmic optimisation.
Change in left ventricular ejection fraction with algorithmic CRT-D optimisation compared with usual device settings.
Ejection fraction will be measured on echocardiography by Simpson's biplane method of discs. Where image quality precludes a full biplane assessment, single plane will be used with the apical 4-chamber view. Participants whose views are not sufficiently good quality for either of these measurements will be excluded from this outcome measure.
Change in left ventricular end-diastolic volume with algorithmic CRT-D optimisation compared with usual device settings.
End-diastolic volume will be measured on echocardiography using Simpson's method. Where image quality precludes a full biplane assessment, single plane will be used with the apical 4-chamber view, or the Teicholz method will be used to estimate end-diastolic volume by measuring left ventricular end-diastolic diameter.
Change in left ventricular diastolic filling time with algorithmic CRT-D optimisation compared with usual device settings.
Diastolic filling time will be measured on echocardiography using pulsed wave Doppler over the mitral inflow of the left ventricle.
Change in left ventricular outflow tract velocity time integral (LVOT VTI) with algorithmic CRT-D optimisation compared with usual device settings.
LVOT VTI will be measured on echocardiography using pulsed wave Doppler within the left ventricular outflow tract, within 1cm of the aortic valve where a clear envelope is visible.
Change in longitudinal and global longitudinal strain (GLS) with algorithmic CRT-D optimisation compared with usual device settings.
Longitudinal strain of the left ventricle will be measured on echocardiography using speckle tracking software
Change in radial strain with algorithmic CRT-D optimisation compared with usual device settings.
Radial strain of the left ventricle will be measured on echocardiography using speckle tracking software
Change in distance walked on a 6-minute hall walk test with algorithmic CRT-D optimisation compared with usual device settings
Patients will undergo a hall walk test at baseline, 12 weeks (with usual settings) and 24 weeks (following algorithmic optimisation) and the difference in distance walked will be analysed.
Change in New York Heart Association functional class with algorithmic CRT-D optimisation compared with usual device settings.
Functional status will be assessed at baseline, after 12 weeks (with usual settings) and after 24 weeks (with optimised settings) and the change in class will be recorded.
Change in score in the Minnesota Living with Heart Failure Questionnaire (MLHFQ) between algorithmically optimised CRT-D and usual device settings.
Patients will complete a questionnaire at baseline, 12 weeks (with usual settings) and 24 weeks (following algorithmic optimisation) and the difference in scores will be analysed. The Minnesota Living with Heart Failure Questionnaire is a validated and commonly used tool for assessing symptoms and activities of daily living in heart failure patients.
Change in cardiac output measured by the novel technique of Ultra-Sound Cardiac Output Monitoring (USCOM) with algorithmic CRT-D optimisation compared with usual device settings
USCOM is a validated technique that correlates well with echocardiographic and Doppler measurements of cardiac output and is performed using a simple probe placed at several points over the chest wall.
Number of patients suffering adverse events during the study period
Hospital patient record systems and telephone calls will be used to find out about hospital admissions and mortality during the study period. An adverse event will be defined as an admission to hospital or death from any cause.

Full Information

First Posted
December 7, 2016
Last Updated
January 21, 2020
Sponsor
Cardiff and Vale University Health Board
Collaborators
Abbott Medical Devices
search

1. Study Identification

Unique Protocol Identification Number
NCT02997670
Brief Title
Quartet Lead With Defibrillator Multisite Algorithmic Cardiac Resynchronisation Therapy Optimisation
Acronym
QUARTRMASTER
Official Title
Quartet Lead With Defibrillator Multisite Algorithmic Cardiac Resynchronisation Therapy Optimisation
Study Type
Interventional

2. Study Status

Record Verification Date
January 2020
Overall Recruitment Status
Withdrawn
Why Stopped
Work focused in other areas, other groups working on this.
Study Start Date
July 1, 2019 (Actual)
Primary Completion Date
July 1, 2019 (Actual)
Study Completion Date
July 1, 2019 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
Cardiff and Vale University Health Board
Collaborators
Abbott Medical Devices

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
Cardiac resynchronisation therapy (CRT) improves outcomes and symptoms in selected patients with heart failure. However, around one third of suitable patients do not demonstrate benefit following device implantation when assessed by echocardiography (heart scanning). This group has poorer outcomes. Response rate can be enhanced by altering timing delays between the pacing leads, but some patients still fail to improve. Quadripolar left ventricular leads are now widely used in CRT. The lead's four poles increase the number of conformations available to the programmer, allowing multiple vectors to be programmed simultaneously or sequentially. This allows programming to avoid, for example, a patch of scar and find an area that will respond better to pacing. This technique is known as multi-site pacing. CRT is often implanted along with a defibrillator lead in the right ventricle, known as CRT-D. The defibrillator lead offers further combinations for pacing. Goal of Research To evaluate an algorithm for assessing different multi-site pacing combinations in optimisation of CRT Outline The investigators will recruit 24 consecutive patients undergoing CRT-D implantation for conventional indications at our hospital. At baseline, patients will undergo echocardiography, exercise testing and assessments of functional ability and quality of life. The device will be implanted as standard. Optimisation will be performed with an algorithm using different vector combinations and assessing the heart's efficiency through echocardiography and invasive pressure monitoring. The pacemaker will be programmed with standard settings. After twelve weeks, the baseline investigations and optimisation algorithm will be repeated and the device programmed according to the maximum efficiency. After a further 12 weeks, the same parameters will be measured to look for improved response to CRT. Potential Benefit To increase the response rate to cardiac resynchronisation therapy and improve reliability of the technique
Detailed Description
Study Design: This is an open-label, single centre, prospective, cohort study to assess the effect of algorithmic, echocardiography-guided optimisation of CRT-D following implantation of a left ventricular quadripolar lead. Introduction and background: There is now considerable evidence that cardiac resynchronisation therapy (CRT) improves outcomes and symptoms in patients with heart failure. However, around a third of patients do not demonstrate any haemodynamic or functional benefit following device implantation. Earlier studies have used a cut-off of 15% reduction in left ventricular end-systolic volume to define response to CRT. Failure to respond to CRT is felt to be multifactorial. Issues include: Anatomical limitations in terms of lead placement (the lead must be placed within a branch of the coronary sinus vein and therefore targeting to the site of maximum contraction delay can be difficult) Presence of areas of scar tissue, which are resistant to being paced Phrenic nerve stimulation with some pacing sites. The phrenic nerve runs close to the heart and may be stimulated by the pacemaker, with the effect that the patient's diaphragm is stimulated and they experience persistent hiccups. High pacing thresholds, which means that increased power from the pacemaker must be used to create a successful pacing stimulus (capture) and that in some instances, capture may not be achieved. This can be due to scar or to poor contact with the heart tissue. Previous work has demonstrated that response can be improved in some patients by optimisation of inter- and intra-ventricular dyssynchrony and atrio-ventricular delay. However, this optimisation is limited by the fixed location of the leads after implantation, with pacing possible from the lead tip only. Multi-site pacing (MSP) with a quadripolar left ventricular lead has been introduced to increase the number of conformations available to the programmer and is especially useful in reducing phrenic nerve capture. Additionally, multiple vectors can be programmed sequentially or simultaneously, allowing for incorporation of a greater number of myocardial segments. Quadripolar left ventricular leads are now being routinely used in many hospitals. They offer new opportunities for optimising CRT-D (resynchronisation-defibrillator) devices by altering the pacing vectors between the four different poles on the lead and the two defibrillator coils to give hundreds of possible pacing combinations. Several recent studies have been published looking at the effect of multi-site pacing on effectiveness of CRT therapy. Generally these employ limited vector combinations, but have already demonstrated beneficial effects on haemodynamics and echocardiographic measures of heart function. The evaluation of an algorithm that examines many more vector combinations, including in combination with the right ventricular defibrillator coils, has not been performed. The researchers propose to investigate this method of optimisation. A gold standard for optimisation of CRT has yet to be defined. Methods employed predominantly include invasive haemodynamic measurement and echocardiography. The ideal method would have low inter-observer variability, high ease of use and rapid sampling rate to allow adjustments to be made and evaluated quickly. USCOM, a continuous-wave Doppler method of continually assessing cardiac output, has been successfully used in optimisation of CRT and has gained a patent, however has so far been used to optimise atrio-ventricular delay only. This method may prove a more accurate, rapid and convenient way to rapidly assess response to changes in pacemaker parameters. The investigators plan to evaluate these three methods of assessment and correlate with each other and with cardiovascular outcomes. Hypothesis: Use of an algorithm in optimisation of cardiac resynchronisation therapy systems containing quadripolar left ventricular leads and dual coil right ventricular leads will increase response rate to this therapy. Study structure: Patients will be recruited consecutively amongst those undergoing implantation of a CRT-D device. At baseline, they will undergo assessment of functional capacity and echocardiographic parameters as well as invasive left ventricular pressure monitoring. The device will be implanted under normal laboratory conditions, aiming for a postero-lateral or lateral cardiac vein left ventricular lead position. Algorithmic echocardiography-guided optimisation of the devices will be conducted at the time of device implantation. Further assessment will be conducted simultaneously with haemodynamic left ventricular pressure monitoring and Ultrasound Cardiac Output Monitoring (USCOM) during device optimisation. The devices will then be programmed with standard CRT parameters. Subjects will be seen at 12 weeks for echocardiographic and functional assessment, following which algorithmic optimisation will be repeated with haemodynamic and echocardiographic monitoring concurrently. Further assessment will be performed at 24 weeks as per baseline, following which the study will end. Study population: 24 men and women, 18 years and older, who are able to attend follow-up assessment 12 and 24 weeks after implantation.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Heart Failure, Systolic
Keywords
Cardiac resynchronization pacing therapy

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Single Group Assignment
Masking
None (Open Label)
Allocation
N/A
Enrollment
0 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Main patient cohort
Arm Type
Experimental
Arm Description
Algorithmic Cardiac Resynchronisation Therapy Optimisation; Patients requiring CRT-D enrolled to receive algorithmic CRT optimisation at their device implantation. They will then be programmed to standard CRT settings for 12 weeks. Assessments will be performed and they will again undergo CRT optimisation using the specified algorithm. This time, they will be programmed to the settings giving maximal cardiac output on echocardiography, as assessed by LVOT VTI. After a further 12 weeks, they will again undergo assessment and the study will terminate.
Intervention Type
Other
Intervention Name(s)
Algorithmic Cardiac Resynchronisation Therapy Optimisation
Intervention Description
Use of a computer algorithm to run through multiple combinations of pacing options in patients who have a Cardiac resynchronisation pacemaker with defibrillator where a quadripolar lead has been implanted. These quadripolar leads offer extra options for pacing locations across the heart, as well as for different combinations and sequences of pacing stimuli.
Primary Outcome Measure Information:
Title
Change in left ventricular end-systolic volume in patients undergoing algorithmic optimisation of CRT-D compared with usual device settings
Description
End-systolic volume will be measured on echocardiography using Simpson's method. Where image quality precludes a full biplane assessment, single plane will be used with the apical 4-chamber view, or alternatively left ventricular end-systolic diameter will be used to calculate volume using the Teicholz method.
Time Frame
24 weeks
Secondary Outcome Measure Information:
Title
Feasibility of algorithmic, echocardiography-guided CRT-D optimisation with quadripolar left ventricular leads in clinical practice
Description
To see whether our optimisation algorithm is acceptable to patients and easy to perform in clinical practice.
Time Frame
24 weeks
Title
Correlation between invasive haemodynamic studies with non-invasive ultrasound-based methods in optimising CRT-D
Description
Attempt to establish a gold standard for optimisation of CRT devices by comparing invasive haemodynamic monitoring, echocardiographic parameters and cardiac output by ultrasound cardiac output monitoring (USCOM)
Time Frame
24 weeks
Title
Change in left ventricular end diastolic pressure with algorithmic optimisation of CRT-D compared with usual device settings
Description
A measurement of left ventricular end-diastolic pressure will be made via invasive left ventricular monitoring after programming with usual settings and again after programming using algorithmic optimisation of the CRT-D device.
Time Frame
24 weeks
Title
Feasibility of ultrasound cardiac output monitoring (USCOM) in optimisation of CRT devices in clinical practice
Description
Use of USCOM has been documented in optimisation of CRT, but not in such rapidly-changing algorithmic optimisation. We plan to assess whether it is feasible to use this in clinical practice as a replacement for a full echocardiogram.
Time Frame
At baseline and at 12 weeks
Title
Change in peak oxygen consumption with algorithmic CRT-D optimisation compared with usual device settings
Description
Assessment of peak oxygen consumption as measured by cardiopulmonary exercise testing after usual settings compared with settings after algorithmic optimisation.
Time Frame
24 weeks
Title
Change in left ventricular ejection fraction with algorithmic CRT-D optimisation compared with usual device settings.
Description
Ejection fraction will be measured on echocardiography by Simpson's biplane method of discs. Where image quality precludes a full biplane assessment, single plane will be used with the apical 4-chamber view. Participants whose views are not sufficiently good quality for either of these measurements will be excluded from this outcome measure.
Time Frame
24 weeks
Title
Change in left ventricular end-diastolic volume with algorithmic CRT-D optimisation compared with usual device settings.
Description
End-diastolic volume will be measured on echocardiography using Simpson's method. Where image quality precludes a full biplane assessment, single plane will be used with the apical 4-chamber view, or the Teicholz method will be used to estimate end-diastolic volume by measuring left ventricular end-diastolic diameter.
Time Frame
24 weeks
Title
Change in left ventricular diastolic filling time with algorithmic CRT-D optimisation compared with usual device settings.
Description
Diastolic filling time will be measured on echocardiography using pulsed wave Doppler over the mitral inflow of the left ventricle.
Time Frame
24 weeks
Title
Change in left ventricular outflow tract velocity time integral (LVOT VTI) with algorithmic CRT-D optimisation compared with usual device settings.
Description
LVOT VTI will be measured on echocardiography using pulsed wave Doppler within the left ventricular outflow tract, within 1cm of the aortic valve where a clear envelope is visible.
Time Frame
24 weeks
Title
Change in longitudinal and global longitudinal strain (GLS) with algorithmic CRT-D optimisation compared with usual device settings.
Description
Longitudinal strain of the left ventricle will be measured on echocardiography using speckle tracking software
Time Frame
24 weeks
Title
Change in radial strain with algorithmic CRT-D optimisation compared with usual device settings.
Description
Radial strain of the left ventricle will be measured on echocardiography using speckle tracking software
Time Frame
24 weeks
Title
Change in distance walked on a 6-minute hall walk test with algorithmic CRT-D optimisation compared with usual device settings
Description
Patients will undergo a hall walk test at baseline, 12 weeks (with usual settings) and 24 weeks (following algorithmic optimisation) and the difference in distance walked will be analysed.
Time Frame
24 weeks
Title
Change in New York Heart Association functional class with algorithmic CRT-D optimisation compared with usual device settings.
Description
Functional status will be assessed at baseline, after 12 weeks (with usual settings) and after 24 weeks (with optimised settings) and the change in class will be recorded.
Time Frame
24 weeks
Title
Change in score in the Minnesota Living with Heart Failure Questionnaire (MLHFQ) between algorithmically optimised CRT-D and usual device settings.
Description
Patients will complete a questionnaire at baseline, 12 weeks (with usual settings) and 24 weeks (following algorithmic optimisation) and the difference in scores will be analysed. The Minnesota Living with Heart Failure Questionnaire is a validated and commonly used tool for assessing symptoms and activities of daily living in heart failure patients.
Time Frame
24 weeks
Title
Change in cardiac output measured by the novel technique of Ultra-Sound Cardiac Output Monitoring (USCOM) with algorithmic CRT-D optimisation compared with usual device settings
Description
USCOM is a validated technique that correlates well with echocardiographic and Doppler measurements of cardiac output and is performed using a simple probe placed at several points over the chest wall.
Time Frame
24 weeks
Title
Number of patients suffering adverse events during the study period
Description
Hospital patient record systems and telephone calls will be used to find out about hospital admissions and mortality during the study period. An adverse event will be defined as an admission to hospital or death from any cause.
Time Frame
From patient enrollment through study completion, an average of 1 year.

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: Successful implantation of CRT-D device with quadripolar left ventricular lead and dual coil right ventricular lead. Exclusion Criteria: failed device implantation inability to complete the follow-up process inability to provide full written consent. Patients unable to perform cardiopulmonary exercise testing will be excluded from this assessment only and will be able to complete the remaining assessments.
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
Freya M Lodge, MBBS
Organizational Affiliation
Cardiff and Vale University Health Board
Official's Role
Principal Investigator
First Name & Middle Initial & Last Name & Degree
Zaheer R Yousef, MBBS
Organizational Affiliation
Cardiff and Vale University Health Board
Official's Role
Study Director
Facility Information:
Facility Name
University Hospital of Wales
City
Cardiff
State/Province
Wales
ZIP/Postal Code
CF14 4XW
Country
United Kingdom

12. IPD Sharing Statement

Plan to Share IPD
No
IPD Sharing Plan Description
Individual participant data available on individual written request at the discretion of the investigators.
Citations:
PubMed Identifier
12084604
Citation
Auricchio A, Stellbrink C, Sack S, Block M, Vogt J, Bakker P, Huth C, Schondube F, Wolfhard U, Bocker D, Krahnefeld O, Kirkels H; Pacing Therapies in Congestive Heart Failure (PATH-CHF) Study Group. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol. 2002 Jun 19;39(12):2026-33. doi: 10.1016/s0735-1097(02)01895-8.
Results Reference
background
PubMed Identifier
12063368
Citation
Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E, Kocovic DZ, Packer M, Clavell AL, Hayes DL, Ellestad M, Trupp RJ, Underwood J, Pickering F, Truex C, McAtee P, Messenger J; MIRACLE Study Group. Multicenter InSync Randomized Clinical Evaluation. Cardiac resynchronization in chronic heart failure. N Engl J Med. 2002 Jun 13;346(24):1845-53. doi: 10.1056/NEJMoa013168.
Results Reference
background
PubMed Identifier
12771115
Citation
Young JB, Abraham WT, Smith AL, Leon AR, Lieberman R, Wilkoff B, Canby RC, Schroeder JS, Liem LB, Hall S, Wheelan K; Multicenter InSync ICD Randomized Clinical Evaluation (MIRACLE ICD) Trial Investigators. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial. JAMA. 2003 May 28;289(20):2685-94. doi: 10.1001/jama.289.20.2685.
Results Reference
background
PubMed Identifier
19800193
Citation
Daubert C, Gold MR, Abraham WT, Ghio S, Hassager C, Goode G, Szili-Torok T, Linde C; REVERSE Study Group. Prevention of disease progression by cardiac resynchronization therapy in patients with asymptomatic or mildly symptomatic left ventricular dysfunction: insights from the European cohort of the REVERSE (Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction) trial. J Am Coll Cardiol. 2009 Nov 10;54(20):1837-46. doi: 10.1016/j.jacc.2009.08.011. Epub 2009 Oct 1.
Results Reference
background
PubMed Identifier
19038680
Citation
Linde C, Abraham WT, Gold MR, St John Sutton M, Ghio S, Daubert C; REVERSE (REsynchronization reVErses Remodeling in Systolic left vEntricular dysfunction) Study Group. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms. J Am Coll Cardiol. 2008 Dec 2;52(23):1834-1843. doi: 10.1016/j.jacc.2008.08.027. Epub 2008 Nov 7.
Results Reference
background
PubMed Identifier
20150328
Citation
Chung ES, Katra RP, Ghio S, Bax J, Gerritse B, Hilpisch K, Peterson BJ, Feldman DS, Abraham WT. Cardiac resynchronization therapy may benefit patients with left ventricular ejection fraction >35%: a PROSPECT trial substudy. Eur J Heart Fail. 2010 Jun;12(6):581-7. doi: 10.1093/eurjhf/hfq009. Epub 2010 Feb 11.
Results Reference
background
PubMed Identifier
26333378
Citation
Osca J, Alonso P, Cano O, Andres A, Miro V, Tello MJ, Olague J, Martinez L, Salvador A. The use of multisite left ventricular pacing via quadripolar lead improves acute haemodynamics and mechanical dyssynchrony assessed by radial strain speckle tracking: initial results. Europace. 2016 Apr;18(4):560-7. doi: 10.1093/europace/euv211. Epub 2015 Sep 1.
Results Reference
background
PubMed Identifier
25214507
Citation
Rinaldi CA, Burri H, Thibault B, Curnis A, Rao A, Gras D, Sperzel J, Singh JP, Biffi M, Bordachar P, Leclercq C. A review of multisite pacing to achieve cardiac resynchronization therapy. Europace. 2015 Jan;17(1):7-17. doi: 10.1093/europace/euu197. Epub 2014 Sep 11.
Results Reference
background
PubMed Identifier
24291411
Citation
Pappone C, Calovic Z, Vicedomini G, Cuko A, McSpadden LC, Ryu K, Romano E, Saviano M, Baldi M, Pappone A, Ciaccio C, Giannelli L, Ionescu B, Petretta A, Vitale R, Fundaliotis A, Tavazzi L, Santinelli V. Multipoint left ventricular pacing improves acute hemodynamic response assessed with pressure-volume loops in cardiac resynchronization therapy patients. Heart Rhythm. 2014 Mar;11(3):394-401. doi: 10.1016/j.hrthm.2013.11.023. Epub 2013 Nov 28.
Results Reference
background
PubMed Identifier
17241315
Citation
Siu CW, Tse HF, Lee K, Chan HW, Chen WH, Yung C, Lee S, Lau CP. Cardiac resynchronization therapy optimization by ultrasonic cardiac output monitoring (USCOM) device. Pacing Clin Electrophysiol. 2007 Jan;30(1):50-5. doi: 10.1111/j.1540-8159.2007.00579.x.
Results Reference
background
PubMed Identifier
25625721
Citation
Zanon F, Baracca E, Pastore G, Marcantoni L, Fraccaro C, Lanza D, Picariello C, Aggio S, Roncon L, Dell'Avvocata F, Rigatelli G, Pacetta D, Noventa F, Prinzen FW. Multipoint pacing by a left ventricular quadripolar lead improves the acute hemodynamic response to CRT compared with conventional biventricular pacing at any site. Heart Rhythm. 2015 May;12(5):975-81. doi: 10.1016/j.hrthm.2015.01.034. Epub 2015 Jan 24.
Results Reference
background

Learn more about this trial

Quartet Lead With Defibrillator Multisite Algorithmic Cardiac Resynchronisation Therapy Optimisation