EM1.1: vestibulo-ocular reflex (VOR) gain during video head impulse testing (vHIT) averaged for the three implanted semicircular canals (VHITG) at 6 months post-implantation, relative to pre-operative baseline, aggregated for all implanted participants.
During video head impulse testing (vHIT), rotational eye and head movement are measured by a goggle-mounted camera and motion sensor during brief, fast ~10 degree head rotations about the axis of a semicircular canal being tested. Performance is quantified by vestibulo-ocular reflex (VOR) gain, which a ratio of eye movement to head movement (in this case the integral of eye velocity relative to the head and the additive inverse of the integral of head velocity, respectively). VHITG, the average vHIT VOR gain for each of the three semicircular canals of the implanted ear, typically ranges from 0 to 1, with higher numbers indicating better performance.
EM1.2: Bruininks-Oseretsky Test of Motor Proficiency 2nd Edition Balance Subtest 5 (BOT) score 6 months post-implantation, relative to pre-operative baseline, aggregated for all implanted participants.
BOT comprises 9 postural tasks scored by an examiner: standing with both feet on a line, standing on one leg, tandem walking, walking on a line with eyes open and eyes closed, and standing on a balance beam. Each task is performed twice and scored (0-4 points, 4 is best). The BOT score is the sum of the best scores for each task. It ranges from 0 to 36 points, with higher scores meaning better outcome and minimally important difference 4.5 points.
EM1.3: Dynamic Gait Index (DGI) score at 6 months post-implantation, relative to pre-operative baseline, aggregated for all implanted participants.
DGI comprises 8 gait tasks scored by an examiner: walking on a level surface, changing gait speed, walking during horizontal head turns, walking during vertical head turns, pivoting, stepping over an obstacle, stepping around obstacles, and waling up and down stairs. Each task is scored (0-3 points, 3 is best). Raw DGI score is the sum of the eight subtest scores and ranges from 0 to 24, with higher scores meaning better outcome and minimally important difference 3.2 points.
EM1.4: Dizziness Handicap Inventory (DHI) score at 6 months post-implantation, relative to pre-operative baseline, aggregated for all implanted participants.
Dizziness Handicap Inventory (DHI) is a 25-question survey designed to quantify self-rated disability due to dizziness and unsteadiness, with scores ranging from 0 (least disabled, best outcome) to 100 (worst outcome). The minimally important difference is 18 points.
EM1.5: SF6D utility (SF6DU) score at 6 months post-implantation, relative to pre-operative baseline, aggregated for all implanted participants.
SF6DU is a unitless health utility score computed from patient-reported replies to the SF6 health-related quality of life survey and ranging from 0 to 1, with higher scores meaning better outcome and minimally important difference 0.03 points.
EM2.1: vestibulo-ocular reflex (VOR) gain during video head impulse testing (vHIT) averaged for the three implanted semicircular canals (VHITG), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
During video head impulse testing (vHIT), rotational eye and head movement are measured by a goggle-mounted camera and motion sensor during brief, fast ~10 degree head rotations about the axis of a semicircular canal being tested. Performance is quantified by vestibulo-ocular reflex (VOR) gain, which a ratio of eye movement to head movement (in this case the integral of eye velocity relative to the head and the additive inverse of the integral of head velocity, respectively). VHITG, the average vHIT VOR gain for each of the three semicircular canals of the implanted ear, typically ranges from 0 to 1, with higher numbers indicating better performance.
EM2.2: Bruininks-Oseretsky Test of Motor Proficiency 2nd Edition Balance Subtest 5 (BOT) score 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
BOT includes 9 postural tasks scored by an examiner: standing with both feet on a line, standing on one leg, tandem walking, walking on a line with eyes open and eyes closed, and standing on a balance beam. Each task is performed twice and scored (0-4 points). The BOT score is the sum of the best scores for each task. It ranges from 0 to 36 points, with higher scores meaning better outcome and minimally important difference 4.5 points.
EM2.3: Dynamic Gait Index (DGI) score at 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
DGI comprises 8 gait tasks scored by an examiner: walking on a level surface, changing gait speed, walking during horizontal head turns, walking during vertical head turns, pivoting, stepping over an obstacle, stepping around obstacles, and waling up and down stairs. Each task is scored (0-3 points, 3 is best). Raw DGI score is the sum of the eight subtest scores and ranges from 0 to 24, with higher scores meaning better outcome and minimally important difference 3.2 points.
EM2.4: Dizziness Handicap Inventory (DHI) score at 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
Dizziness Handicap Inventory (DHI) is a 25-question survey designed to quantify self-rated disability due to dizziness and unsteadiness, with scores ranging from 0 (least disabled, best outcome) to 100 (worst outcome). The minimally important difference is 18 points.
EM2.5: SF6D utility (SF6DU) score at 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
SF6DU is a unitless health utility score computed from patient-reported replies to the SF6 health-related quality of life survey and ranging from 0 to 1, with higher scores meaning better outcome and minimally important difference 0.03 points.
EM3.1: vestibulo-ocular reflex (VOR) gain during video head impulse testing (vHIT) averaged for the three implanted semicircular canals (VHITG), for the subgroup of participants who have ototoxic adult-onset BVH
During video head impulse testing (vHIT), rotational eye and head movement are measured by a goggle-mounted camera and motion sensor during brief, fast ~10 degree head rotations about the axis of a semicircular canal being tested. Performance is quantified by vestibulo-ocular reflex (VOR) gain, which a ratio of eye movement to head movement (in this case the integral of eye velocity relative to the head and the additive inverse of the integral of head velocity, respectively). VHITG, the average vHIT VOR gain for each of the three semicircular canals of the implanted ear, typically ranges from 0 to 1, with higher numbers indicating better performance.
EM3.2: Bruininks-Oseretsky Test of Motor Proficiency 2nd Edition Balance Subtest 5 (BOT) score 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have ototoxic adult-onset BVH
BOT includes 9 postural tasks scored by an examiner: standing with both feet on a line, standing on one leg, tandem walking, walking on a line with eyes open and eyes closed, and standing on a balance beam. Each task is performed twice and scored (0-4 points). The BOT score is the sum of the best scores for each task. It ranges from 0 to 36 points, with higher scores meaning better outcome and minimally important difference 4.5 points.
EM3.3: Dynamic Gait Index (DGI) score at 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have ototoxic adult-onset BVH
DGI comprises 8 gait tasks scored by an examiner: walking on a level surface, changing gait speed, walking during horizontal head turns, walking during vertical head turns, pivoting, stepping over an obstacle, stepping around obstacles, and waling up and down stairs. Each task is scored (0-3 points, 3 is best). Raw DGI score is the sum of the eight subtest scores and ranges from 0 to 24, with higher scores meaning better outcome and minimally important difference 3.2 points.
EM3.4: Dizziness Handicap Inventory (DHI) score at 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have ototoxic adult-onset BVH
Dizziness Handicap Inventory (DHI) is a 25-question survey designed to quantify self-rated disability due to dizziness and unsteadiness, with scores ranging from 0 (least disabled, best outcome) to 100 (worst outcome). The minimally important difference is 18 points.
EM3.5: SF6D utility (SF6DU) score at 6 months post-implantation, relative to pre-operative baseline, for the subgroup of participants who have ototoxic adult-onset BVH
SF6DU is a unitless health utility score computed from patient-reported replies to the SF6 health-related quality of life survey and ranging from 0 to 1, with higher scores meaning better outcome and minimally important difference 0.03 points.
SM0.1: Type of unanticipated adverse device effects as of 6 months post-implantation
Assess safety of the intervention, aggregating data across all implanted participants (diversified across ototoxic and non-ototoxic/non-central adult-onset BVH), as determined by (type of) changes from pre-intervention to 6 months post-implantation
SM0.2: Incidence of unanticipated adverse device effects as of 6 months post-implantation
Assess safety of the intervention, aggregating data across all implanted participants (diversified across ototoxic and non-ototoxic/non-central adult-onset BVH), as determined by incidence of adverse device-related events from pre-intervention to 6 months post-implantation
SM0.3: Severity of unanticipated adverse device effects as of 6 months post-implantation
Assess safety of the intervention, aggregating data across all implanted participants (diversified across ototoxic and non-ototoxic/non-central adult-onset BVH), as determined by severity of adverse device-related events from pre-intervention to 6 months post-implantation
SM1.1: Changes in implant-side bone-conduction 4-frequency (0.5, 1, 2, 4 kHz) pure tone average detection threshold (BonePTAi), aggregated across all implant users
BonePTAi is the 4-frequency average of pure tone detection thresholds, in dBHL, for implant-side bone-conduction presentation of pure tones at 0.5, 1, 2 and 4 kHz. Range is -10 to 80 dBHL, with change toward smaller absolute values representing better (closer to normal) outcome.
SM1.2: Changes in implant-side air-conduction 4-frequency pure tone average detection threshold (AirPTAi), aggregated across all implant users
AirPTAi is the 4-frequency average of pure tone detection thresholds, in dBHL, for implant-side air-conduction presentation of pure tones at 0.5, 1, 2 and 4 kHz. Range is -10 to 120 dBHL, with change toward smaller absolute values representing better (closer to normal) outcome.
SM1.3: Changes in implant-side consonant-nucleus-consonant word discrimination score (CNCWi), aggregated across all implant users
CNCWi is the percentage of monosyllabic consonant-nucleus-consonant words correctly repeated when presented via an earphone speaker on the implanted side in a sound-isolation booth while a masking noise is played to the contralateral ear at 40 dB above its bone-conduction 4-frequency pure tone detection threshold if that threshold is at least 10 dB better than that of the ear being tested. CNCWi ranges from 0-100% correct, with higher score meaning better outcome.
SM1.4: Changes in sound-field Arizona Biology (AzBios) sentences-in-noise discrimination score, aggregated across all implant users
AzBios is the percentage of words correctly repeated when a set of 20 Arizona Biology sentences (randomly chosen without repeats from 15 sets) is presented at a moderate level (60 dB SPL) along with masking noise (simultaneous presentation of ten sentences from another set at aggregate sound level 55 dBSPL) to both ears via a sound field speaker in a sound-isolation booth. Scores range from 0-100% correct, with higher score meaning better outcome.
SM1.5: Change in proportion of implanted ears with AAO-HNS 1995 Class A/B hearing or changes of ≤30dB and ≤30% from preoperative baseline AirPTAi ≥ 20dBHL and CNCWi ≤ 80%, respectively, aggregating data across all implanted participants (ClassABCNCWa)
For ClassABCNCW, AAO-HNS (American Academy of Otolaryngology - Head & Neck Surgery) 1995 Class A or B hearing is defined as AirPTAi not worse than 50 dBHL and CNCWi not worse than 50%. ClassABCNCW proportion ranges from 0-100%, with higher values meaning better outcomes.
SM1.6: Change in proportion of implanted ears with AAO-HNS 1995 Class A/B hearing or changes of ≤30dB and ≤30% from preoperative baseline AirPTAi ≥ 20dBHL and AzBios ≤ 80%, respectively, aggregating across all implanted participants (ClassABAzBioa)
For ClassABAzBio, AAO-HNS (American Academy of Otolaryngology - Head & Neck Surgery) 1995 Class A or B hearing is defined as AirPTAi not worse than 50 dBHL and sound-field-presentation Arizona Biology sentences-in-noise discrimination score (AzBios) not worse than 50%. ClassABAzBio proportion ranges from 0-100%, with higher values meaning better outcomes.
SM1.7: Change in implant-side ear distortion-product otoacoustic emission (DPOAE) signal-to-noise ratio for responses to tones centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz (an 8-valued vector), aggregated across all implant users
Distortion-product otoacoustic emissions (DPOAE) are sounds generated by a normal cochlea and detectable using a microphone in the ear canal when pairs of pure tones (centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz) are presented to the ear. DPOAE signals (and 8-valued vector for responses to stimuli centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz) are quantified in dB relative to the frequency-specific noise floor recorded by the same microphone before onset of stimulation. Response range is 0-15 dB, with higher values meaning better outcome.
SM1.8: Changes in implant-side tympanometry responses (a 3-valued vector comprising peak compliance in mL, pressure in decaPascals at peak admittance, and ear canal volume in mL), aggregated across all implant users
Tympanometry involves using an ear canal microphone to measure ear canal acoustic admittance (conventionally reported in units of mL) and volume (in mL) by sensing reflection of a 226 Hz tone presented via a speaker in the ear canal while ear canal pressure is slowly varied from -400 to 200 decaPascals. Negative bias pressures pull the ear drum outward, positive bias pressures push the ear drum inward, and acoustic admittance is greatest when the ear drum is in a neutral anatomic position, neither retracted inward nor ballooned outward. Results are conventionally represented graphically then summarized by a three-valued vector comprising peak compliance (range 0-2mL, values closer to normal range of 0.3-1.4 mL meaning more normal outcome), pressure at which compliance peaks (range -400 to 200 decaPascal, values closer to 0 meaning more normal outcome), and ear canal volume (range 0-3 mL, values closer to normal adult range of 0.6-1.4 meaning more normal outcome).
SM1.9: Changes in Tinnitus Handicap Inventory (THI), aggregated across all implant users
The Tinnitus Handicap Inventory (THI) survey is set of questions designed to quantify self-perceived handicap due to tinnitus. THI score ranges from 0-100. Higher scores mean worse outcome.
SM1.10: Changes in Autophony Index (AI), aggregated across all implant users
The Autophony Index (AI) survey is set of questions designed to quantify self-perceived autophony (hearing one's voice or other internally-generated body sounds). AI score ranges from 0-104. Higher scores mean worse outcomes.
SM1.11: Changes in cervical vestibular-evoked myogenic potential (cVEMP) response amplitude, aggregated across all implant users
Cervical vestibular-evoked myogenic potentials (cVEMPs) are a measure of saccule-mediated inhibition of sternocleidomastoid muscle electromyographic (EMG) activity during presentation of loud tones to the ipsilateral ear. After normalizing to average rectified EMG activity (in microvolts) prior to sound stimulation, cVEMP amplitudes (in microvolts) are unitless. Values range from 0 to ~10, with larger values suggesting increased saccule activity and generally indicating better outcomes, except that values above 3.0 suggest abnormally high saccule sensitivity to sound.
SM1.12: Changes in ocular vestibular-evoked myogenic potential (oVEMP) response amplitude, aggregated across all implant users
Ocular vestibular-evoked myogenic potentials (oVEMPs) are a measure of utricle-mediated electromyographic (EMG) activity of inferior oblique and inferior rectus extraocular muscles during presentation of loud tones to the contralateral ear. Amplitude ranges from 0 to ~50 microvolts, with larger values generally suggesting increased utricle activity and better outcomes but values above 17 microvolts suggesting abnormally high utricle sensitivity to sound.
SM1.13: Changes in subjective visual vertical (SVV) as assessed using the Bucket Test, aggregated across all implant users
A participant's vision is occluded by a bucket oriented so that its axis aligns with the participant's nasooccipital axis. The examiner rotates the bucket about the axis to a new orientation, the participant reorients it until a line drawn on the floor of the bucket is judges by the participant to be Earth-vertical and deviation of that line from true vertical is recorded. Values range over -15 to 15 degrees, with smaller absolute values meaning more normal outcomes.
SM1.14: Changes in scalp thickness at the location of implanted magnets (Tscalp), aggregated across all implant users
Scalp thickness at the location of implanted magnets (Tscalp) is measured in mm using a magnetic sensor. Range is 0 to 10 mm, with values closer to the range 3-7mm indicating better outcomes.
SM1.15: Changes in all adverse events, including but not limited to events related to the surgical procedure, device and device systems, aggregated across all implant users
Assess safety of the intervention, aggregating data across all implanted participants (diversified across ototoxic and non-ototoxic/non-central adult-onset BVH), as determined by changes in hearing and otolith endorgan function indicated by changes in all adverse events, including but not limited to events related to the surgical procedure, device and device systems
SM2.1: Changes in implant-side bone-conduction 4-frequency (0.5, 1, 2, 4 kHz) pure tone average detection threshold (BonePTAi), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
BonePTAi is the 4-frequency average of pure tone detection thresholds, in dBHL, for implant-side bone-conduction presentation of pure tones at 0.5, 1, 2 and 4 kHz. Range is -10 to 80 dBHL, with change toward smaller absolute values representing better (closer to normal) outcome.
SM2.2: Changes in implant-side air-conduction 4-frequency pure tone average detection threshold (AirPTAi), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
AirPTAi is the 4-frequency average of pure tone detection thresholds, in dBHL, for implant-side air-conduction presentation of pure tones at 0.5, 1, 2 and 4 kHz. Range is -10 to 120 dBHL, with change toward smaller absolute values representing better (closer to normal) outcome.
SM2.3: Changes in implant-side consonant-nucleus-consonant word discrimination score (CNCWi), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
CNCWi is the percentage of monosyllabic consonant-nucleus-consonant words correctly repeated when presented via an earphone speaker on the implanted side in a sound-isolation booth while a masking noise is played to the contralateral ear at 40 dB above its bone-conduction 4-frequency pure tone detection threshold if that threshold is at least 10 dB better than that of the ear being tested. CNCWi ranges from 0-100% correct, with higher score meaning better outcome.
SM2.4: Changes in sound-field Arizona Biology (AzBios) sentences-in-noise discrimination score, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
AzBios is the percentage of words correctly repeated when a set of 20 Arizona Biology sentences (randomly chosen without repeats from 15 sets) is presented at a moderate level (60 dB SPL) along with masking noise (simultaneous presentation of ten sentences from another set at aggregate sound level 55 dBSPL) to both ears via a sound field speaker in a sound-isolation booth. Scores range from 0-100% correct, with higher score meaning better outcome.
SM2.5: Change in proportion of implanted ears with AAO-HNS 1995 Class A/B hearing or changes of ≤30dB and ≤30% from preop baseline AirPTAi ≥ 20dBHL and CNCWi ≤ 80%, respectively, for non-ototoxic/non-central adult-onset BVH participants (ClassABCNCWn)
For ClassABCNCW, AAO-HNS (American Academy of Otolaryngology - Head & Neck Surgery) 1995 Class A or B hearing is defined as AirPTAi not worse than 50 dBHL and CNCWi not worse than 50%. ClassABCNCW proportion ranges from 0-100%, with higher values meaning better outcomes.
SM2.6: Change in proportion of implanted ears with AAO-HNS 1995 Class A/B hearing or changes of ≤30dB and ≤30% from preop baseline AirPTAi ≥ 20dBHL and AzBios ≤ 80%, respectively, for non-ototoxic/non-central BVH participants (ClassABAzBion)
For ClassABAzBio, AAO-HNS (American Academy of Otolaryngology - Head & Neck Surgery) 1995 Class A or B hearing is defined as AirPTAi not worse than 50 dBHL and sound-field-presentation Arizona Biology sentences-in-noise discrimination score (AzBios) not worse than 50%. ClassABAzBio proportion ranges from 0-100%, with higher values meaning better outcomes.
SM2.7: Change in implant-side distortion-product otoacoustic emission (DPOAE) signal/noise ratio for response to tones centered on 0.75/1/1.5/2/3/4/6/8 kHz, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
Distortion-product otoacoustic emissions (DPOAE) are sounds generated by a normal cochlea and detectable using a microphone in the ear canal when pairs of pure tones (centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz) are presented to the ear. DPOAE signals (and 8-valued vector for responses to stimuli centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz) are quantified in dB relative to the frequency-specific noise floor recorded by the same microphone before onset of stimulation. Response range is 0-15 dB, with higher values meaning better outcome.
SM2.8: Changes in implant-side tympanometry response (3-valued vector comprising peak admittance in mL, pressure in decaPascals at peak admittance, and ear canal volume in mL), for the participants who have non-ototoxic/non-central adult-onset BVH
Tympanometry involves using an ear canal microphone to measure ear canal acoustic admittance (conventionally reported in units of mL) and volume (in mL) by sensing reflection of a 226 Hz tone presented via a speaker in the ear canal while ear canal pressure is slowly varied from -400 to 200 decaPascals. Negative bias pressures pull the ear drum outward, positive bias pressures push the ear drum inward, and acoustic admittance is greatest when the ear drum is in a neutral anatomic position, neither retracted inward nor ballooned outward. Results are conventionally represented graphically then summarized by a three-valued vector comprising peak compliance (range 0-2mL, values closer to normal range of 0.3-1.4 mL meaning more normal outcome), pressure at which compliance peaks (range -400 to 200 decaPascal, values closer to 0 meaning more normal outcome), and ear canal volume (range 0-3 mL, values closer to normal adult range of 0.6-1.4 meaning more normal outcome).
SM2.9: Changes in Tinnitus Handicap Inventory (THI), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
The Tinnitus Handicap Inventory (THI) survey is set of questions designed to quantify self-perceived handicap due to tinnitus. THI score ranges from 0-100. Higher scores mean worse outcomes.
SM2.10: Changes in Autophony Index (AI), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
The Autophony Index (AI) survey is set of questions designed to quantify self-perceived autophony (hearing one's voice or other internally-generated body sounds). AI score ranges from 0-104. Higher scores mean worse outcomes.
SM2.11: Changes in cervical vestibular-evoked myogenic potential (cVEMP) response amplitude, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
Cervical vestibular-evoked myogenic potentials (cVEMPs) are a measure of saccule-mediated inhibition of sternocleidomastoid muscle electromyographic (EMG) activity during presentation of loud tones to the ipsilateral ear. After normalizing to average rectified EMG activity (in microvolts) prior to sound stimulation, cVEMP amplitudes (in microvolts) are unitless. Values range from 0 to ~10, with larger values suggesting increased saccule activity and generally indicating better outcomes, except that values above 3.0 suggest abnormally high saccule sensitivity to sound.
SM2.12: Changes in ocular vestibular-evoked myogenic potential (oVEMP) response amplitude, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
Ocular vestibular-evoked myogenic potentials (oVEMPs) are a measure of utricle-mediated electromyographic (EMG) activity of inferior oblique and inferior rectus extraocular muscles during presentation of loud tones to the contralateral ear. Amplitude ranges from 0 to ~50 microvolts, with larger values generally suggesting increased utricle activity and better outcomes but values above 17 microvolts suggesting abnormally high utricle sensitivity to sound.
SM2.13: Changes in subjective visual vertical (SVV) as assessed using the Bucket Test, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
A participant's vision is occluded by a bucket oriented so that its axis aligns with the participant's nasooccipital axis. The examiner rotates the bucket about the axis to a new orientation, the participant reorients it until a line drawn on the floor of the bucket is judges by the participant to be Earth-vertical and deviation of that line from true vertical is recorded. Values range over -15 to 15 degrees, with smaller absolute values meaning more normal outcomes.
SM2.14: Changes in scalp thickness at the location of implanted magnets (Tscalp), for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
Scalp thickness at the location of implanted magnets (Tscalp) is measured in mm using a magnetic sensor. Range is 0 to 10 mm, with values closer to the range 3-7mm indicating better outcomes.
SM2.15: Changes in all adverse events, including but not limited to events related to the surgical procedure, device and device systems, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH
Assess safety of the intervention, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH, as determined by changes in hearing and otolith endorgan function indicated by changes in all adverse events, including but not limited to events related to the surgical procedure, device and device systems
SM3.1: Changes in implant-side bone-conduction 4-frequency (0.5, 1, 2, 4 kHz) pure tone average detection threshold (BonePTAi), for the subgroup of participants who have ototoxic adult-onset BVH
BonePTAi is the 4-frequency average of pure tone detection thresholds, in dBHL, for implant-side bone-conduction presentation of pure tones at 0.5, 1, 2 and 4 kHz. Range is -10 to 80 dBHL, with change toward smaller absolute values representing better (closer to normal) outcome.
SM3.2: Changes in implant-side air-conduction 4-frequency pure tone average detection threshold (AirPTAi), for the subgroup of participants who have ototoxic adult-onset BVH
AirPTAi is the 4-frequency average of pure tone detection thresholds, in dBHL, for implant-side air-conduction presentation of pure tones at 0.5, 1, 2 and 4 kHz. Range is -10 to 120 dBHL, with change toward smaller absolute values representing better (closer to normal) outcome.
SM3.3: Changes in implant-side consonant-nucleus-consonant word discrimination score (CNCWi), for the subgroup of participants who have ototoxic adult-onset BVH
CNCWi is the percentage of monosyllabic consonant-nucleus-consonant words correctly repeated when presented via an earphone speaker on the implanted side in a sound-isolation booth while a masking noise is played to the contralateral ear at 40 dB above its bone-conduction 4-frequency pure tone detection threshold if that threshold is at least 10 dB better than that of the ear being tested. CNCWi ranges from 0-100% correct, with higher score meaning better outcome.
SM3.4: Changes in sound-field Arizona Biology (AzBios) sentences-in-noise discrimination score, for the subgroup of participants who have ototoxic adult-onset BVH
AzBios is the percentage of words correctly repeated when a set of 20 Arizona Biology sentences (randomly chosen without repeats from 15 sets) is presented at a moderate level (60 dB SPL) along with masking noise (simultaneous presentation of ten sentences from another set at aggregate sound level 55 dBSPL) to both ears via a sound field speaker in a sound-isolation booth. Scores range from 0-100% correct, with higher score meaning better outcome.
SM3.5: Change in proportion of implanted ears with AAO-HNS 1995 Class A/B hearing or changes of ≤30dB and ≤30% from preoperative baseline AirPTAi ≥ 20dBHL and CNCWi ≤ 80%, respectively, aggregating across participants with ototoxic BVH (ClassABCNCWo)
For ClassABCNCW, AAO-HNS (American Academy of Otolaryngology - Head & Neck Surgery) 1995 Class A or B hearing is defined as AirPTAi not worse than 50 dBHL and CNCWi not worse than 50%. ClassABCNCW proportion ranges from 0-100%, with higher values meaning better outcomes.
SM3.6: Change in proportion of implanted ears with AAO-HNS 1995 Class A/B hearing or changes of ≤30dB and ≤30% from preoperative baseline AirPTAi ≥ 20dBHL and AzBios ≤ 80%, respectively, aggregating across participants with ototoxic BVH (ClassABAzBioo)
For ClassABAzBio, AAO-HNS (American Academy of Otolaryngology - Head & Neck Surgery) 1995 Class A or B hearing is defined as AirPTAi not worse than 50 dBHL and sound-field-presentation Arizona Biology sentences-in-noise discrimination score (AzBios) not worse than 50%. ClassABAzBio proportion ranges from 0-100%, with higher values meaning better outcomes.
SM3.7: Change in implant-side distortion-product otoacoustic emission (DPOAE) signal/noise ratio for response to tones centered on 0.75/1/1.5/2/3/4/6/8 kHz, for participants who have ototoxic adult-onset BVH
Distortion-product otoacoustic emissions (DPOAE) are sounds generated by a normal cochlea and detectable using a microphone in the ear canal when pairs of pure tones (centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz) are presented to the ear. DPOAE signals (and 8-valued vector for responses to stimuli centered on 0.75, 1, 1.5, 2, 3, 4, 6 and 8 kHz) are quantified in dB relative to the frequency-specific noise floor recorded by the same microphone before onset of stimulation. Response range is 0-15 dB, with higher values meaning better outcome.
SM3.8: Changes in implant-side tympanometry responses (a 3-valued vector comprising peak compliance in mL, pressure in decaPascals at peak admittance, and ear canal volume in mL), for the subgroup of participants who have ototoxic adult-onset BVH
Tympanometry involves using an ear canal microphone to measure ear canal acoustic admittance (conventionally reported in units of mL) and volume (in mL) by sensing reflection of a 226 Hz tone presented via a speaker in the ear canal while ear canal pressure is slowly varied from -400 to 200 decaPascals. Negative bias pressures pull the ear drum outward, positive bias pressures push the ear drum inward, and acoustic admittance is greatest when the ear drum is in a neutral anatomic position, neither retracted inward nor ballooned outward. Results are conventionally represented graphically then summarized by a three-valued vector comprising peak compliance (range 0-2mL, values closer to normal range of 0.3-1.4 mL meaning more normal outcome), pressure at which compliance peaks (range -400 to 200 decaPascal, values closer to 0 meaning more normal outcome), and ear canal volume (range 0-3 mL, values closer to normal adult range of 0.6-1.4 meaning more normal outcome).
SM3.9: Changes in Tinnitus Handicap Inventory (THI), for the subgroup of participants who have ototoxic adult-onset BVH
The Tinnitus Handicap Inventory (THI) survey is set of questions designed to quantify self-perceived handicap due to tinnitus. THI score ranges from 0-100. Higher scores mean worse outcomes.
SM3.10: Changes in Autophony Index (AI), for the subgroup of participants who have ototoxic adult-onset BVH
The Autophony Index (AI) survey is set of questions designed to quantify self-perceived autophony (hearing one's voice or other internally-generated body sounds). AI score ranges from 0-104. Higher scores mean worse outcomes.
SM3.11: Changes in cervical vestibular-evoked myogenic potential (cVEMP) response amplitude, for the subgroup of participants who have ototoxic adult-onset BVH
Cervical vestibular-evoked myogenic potentials (cVEMPs) are a measure of saccule-mediated inhibition of sternocleidomastoid muscle electromyographic (EMG) activity during presentation of loud tones to the ipsilateral ear. After normalizing to average rectified EMG activity (in microvolts) prior to sound stimulation, cVEMP amplitudes (in microvolts) are unitless. Values range from 0 to ~10, with larger values suggesting increased saccule activity and generally indicating better outcomes, except that values above 3.0 suggest abnormally high saccule sensitivity to sound.
SM3.12: Changes in ocular vestibular-evoked myogenic potential (oVEMP) response amplitude, for the subgroup of participants who have ototoxic adult-onset BVH
Ocular vestibular-evoked myogenic potentials (oVEMPs) are a measure of utricle-mediated electromyographic (EMG) activity of inferior oblique and inferior rectus extraocular muscles during presentation of loud tones to the contralateral ear. Amplitude ranges from 0 to ~50 microvolts, with larger values generally suggesting increased utricle activity and better outcomes but values above 17 microvolts suggesting abnormally high utricle sensitivity to sound.
SM3.13: Changes in subjective visual vertical (SVV) as assessed using the Bucket Test, for the subgroup of participants who have ototoxic adult-onset BVH
A participant's vision is occluded by a bucket oriented so that its axis aligns with the participant's nasooccipital axis. The examiner rotates the bucket about the axis to a new orientation, the participant reorients it until a line drawn on the floor of the bucket is judges by the participant to be Earth-vertical and deviation of that line from true vertical is recorded. Values range over -15 to 15 degrees, with smaller absolute values meaning more normal outcomes.
SM3.14: Changes in scalp thickness at the location of implanted magnets (Tscalp), for the subgroup of participants who have ototoxic adult-onset BVH
Scalp thickness at the location of implanted magnets (Tscalp) is measured in mm using a magnetic sensor. Range is 0 to 10 mm, with values closer to the range 3-7mm indicating better outcomes.
SM3.15: Changes in all adverse events, including but not limited to events related to the surgical procedure, device and device systems, for the subgroup of participants who have ototoxic adult-onset BVH
Assess safety of the intervention, for the subgroup of participants who have ototoxic adult-onset BVH, as determined by changes in hearing and otolith endorgan function indicated by changes in all adverse events, including but not limited to events related to the surgical procedure, device and device systems
TM1: Proportion of all implanted participants who continue to use the MVI system daily at 6 months post-implantation
Assess tolerance of the intervention, aggregating data across all implanted participants, as determined by the proportion of all implant recipients who continue to use the MVI system daily at 6 months post-implantation
TM2: Proportion of implanted participants with non-ototoxic/non-central adult-onset BVH who continue to use the MVI system daily at 6 months post-implantation
Assess tolerance of the intervention, for the subgroup of participants who have non-ototoxic/non-central adult-onset BVH, as determined by the proportion of implanted participants with non-ototoxic/non-central adult-onset BVH who continue to use the MVI system daily at 6 months post-implantation
TM3: Proportion of implanted participants with ototoxic adult-onset BVH who continue to use the MVI system daily at 6 months post-implantation
Assess tolerance of the intervention, for the subgroup of participants who have ototoxic adult-onset BVH, as determined by the proportion of implanted participants with ototoxic adult-onset BVH who continue to use the MVI system daily at 6 months post-implantation