Comparison of Blood Pressure Measurement With an Omron HEM-907 Device in a Clinical Setting and a Research Setting (OMRONHEM-907)

Primary Purpose

Hypertension

Status
Completed
Phase
Not Applicable
Locations
Canada
Study Type
Interventional
Intervention
Sequence: Clinical/Research
Sequence: Research/Clinical
Sponsored by
Institut de Recherches Cliniques de Montreal
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional diagnostic trial for Hypertension focused on measuring hypertension, blood pressure, systolic, diastolic, AOBP, Omron HEM-907, research, clinical

Eligibility Criteria

18 Years - undefined (Adult, Older Adult)All SexesDoes not accept healthy volunteers

Inclusion Criteria:

  • At least 18 years old
  • Hypertensive or presumed hypertensive
  • Patients with a scheduled visit at the IRCM hypertension clinic.

Exclusion Criteria:

  • Limitation preventing the measurement of blood pressure
  • Introduction or withdrawal of any blood pressure-modifying medication starting 4 weeks before the first and until the last study visit.
  • Recreative drug use (except cannabis)
  • Alcohol use of more than 4 drinks during the day
  • New symptoms during the research visit, that warrants a medical evaluation
  • Mean systolic blood pressure >180 mm Hg or mean diastolic blood pressure >110 mm Hg (using AOBP)

Sites / Locations

  • Institut de Recherches Cliniques de Montréal

Arms of the Study

Arm 1

Arm 2

Arm Type

Experimental

Active Comparator

Arm Label

Sequence: Clinical/Research

Sequence: Research/Clinical

Arm Description

Participants assigned to this arm will have their blood pressure measured in a clinical setting first, and in a research setting second. The sequence randomization corresponds to the intervention. Visits will be at least a day apart but within a two-week period. During the clinical visit, they will have their blood pressure measured with the Omron HEM-907, an automated office blood pressure (AOBP) device. During the research setting, participants will be guided through a series of research-driven steps such as study questionnaires and completion of consent forms. They will have their blood pressure measured in both arms with a mercury sphygmomanometer, and then 3 measurements with a mercury sphygmomanometer. AOBP measurements (Omron HEM-907) will be performed at the end of the visit.

Participants assigned to this arm will go through the same measurements and procedures exception made of the research-first and clinical-second sequence. The intervention to which they are randomized corresponds to the sequence of the visits.

Outcomes

Primary Outcome Measures

Mean automated office blood pressure systolic blood pressure estimates

Secondary Outcome Measures

Mean automated office blood pressure diastolic blood pressure estimates.
Individual unaveraged automated office systolic and diastolic blood pressure measurements (1, 2 and 3)

Full Information

First Posted
November 17, 2017
Last Updated
April 10, 2018
Sponsor
Institut de Recherches Cliniques de Montreal
search

1. Study Identification

Unique Protocol Identification Number
NCT03360669
Brief Title
Comparison of Blood Pressure Measurement With an Omron HEM-907 Device in a Clinical Setting and a Research Setting
Acronym
OMRONHEM-907
Official Title
Comparison of Blood Pressure Measurement With an Omron HEM-907 Device in a Clinical Setting and a Research Setting
Study Type
Interventional

2. Study Status

Record Verification Date
April 2018
Overall Recruitment Status
Completed
Study Start Date
November 27, 2017 (Actual)
Primary Completion Date
March 29, 2018 (Actual)
Study Completion Date
March 29, 2018 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
Institut de Recherches Cliniques de Montreal

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
Automated office blood pressure (AOBP) devices are recommended for high blood pressure diagnosis. One of those devices is the Omron HEM-907 (Omron Healthcare). It is currently not known if blood pressure measurements performed with this device in a clinical setting are equivalent to those made in a research setting. Therefore, this randomized-controlled trial was designed. It aims to compare the blood pressure measurements performed in a clinical and a research setting. Seventy patients will be included and randomized to a clinical/research or research/clinical sequence.
Detailed Description
Hypertension is a common condition, with an estimated prevalence around 20 percent in Canada. It is a major risk factor for cardiovascular disease. Recently, Hypertension Canada has updated its guidelines and recommend we should consider a 120 mm Hg systolic blood pressure target in high-risk patients aged 50 years or older. This recommendation derives mainly from the SPRINT trial. In this trial, blood pressure was measured with an automated oscillometric blood pressure (AOBP) device, the Omron HEM-907 (Omron Healthcare). However, it is not known if measurements performed with this device in a clinical setting are equivalent to those made in a research setting. Therefore, this randomized-controlled trial was designed. It aims to compare the blood pressure measurements of 70 patients in a clinical and a research setting. Patients will be randomized to a clinical/research or a research/clinical sequence.

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Hypertension
Keywords
hypertension, blood pressure, systolic, diastolic, AOBP, Omron HEM-907, research, clinical

7. Study Design

Primary Purpose
Diagnostic
Study Phase
Not Applicable
Interventional Study Model
Parallel Assignment
Masking
None (Open Label)
Allocation
Randomized
Enrollment
70 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Sequence: Clinical/Research
Arm Type
Experimental
Arm Description
Participants assigned to this arm will have their blood pressure measured in a clinical setting first, and in a research setting second. The sequence randomization corresponds to the intervention. Visits will be at least a day apart but within a two-week period. During the clinical visit, they will have their blood pressure measured with the Omron HEM-907, an automated office blood pressure (AOBP) device. During the research setting, participants will be guided through a series of research-driven steps such as study questionnaires and completion of consent forms. They will have their blood pressure measured in both arms with a mercury sphygmomanometer, and then 3 measurements with a mercury sphygmomanometer. AOBP measurements (Omron HEM-907) will be performed at the end of the visit.
Arm Title
Sequence: Research/Clinical
Arm Type
Active Comparator
Arm Description
Participants assigned to this arm will go through the same measurements and procedures exception made of the research-first and clinical-second sequence. The intervention to which they are randomized corresponds to the sequence of the visits.
Intervention Type
Other
Intervention Name(s)
Sequence: Clinical/Research
Intervention Description
Participants will be assigned to a sequence of: research visit first/clinical visit second.
Intervention Type
Other
Intervention Name(s)
Sequence: Research/Clinical
Intervention Description
Participants will be assigned to a sequence of: clinical visit first/research visit second.
Primary Outcome Measure Information:
Title
Mean automated office blood pressure systolic blood pressure estimates
Time Frame
Up to 2 weeks
Secondary Outcome Measure Information:
Title
Mean automated office blood pressure diastolic blood pressure estimates.
Time Frame
Up to 2 weeks
Title
Individual unaveraged automated office systolic and diastolic blood pressure measurements (1, 2 and 3)
Time Frame
Up to 2 weeks

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria: At least 18 years old Hypertensive or presumed hypertensive Patients with a scheduled visit at the IRCM hypertension clinic. Exclusion Criteria: Limitation preventing the measurement of blood pressure Introduction or withdrawal of any blood pressure-modifying medication starting 4 weeks before the first and until the last study visit. Recreative drug use (except cannabis) Alcohol use of more than 4 drinks during the day New symptoms during the research visit, that warrants a medical evaluation Mean systolic blood pressure >180 mm Hg or mean diastolic blood pressure >110 mm Hg (using AOBP)
Facility Information:
Facility Name
Institut de Recherches Cliniques de Montréal
City
Montreal
State/Province
Quebec
ZIP/Postal Code
H2W1R7
Country
Canada

12. IPD Sharing Statement

Plan to Share IPD
Yes
IPD Sharing Plan Description
according to specific scientifically pertinent demands
Citations:
PubMed Identifier
26711315
Citation
Padwal RS, Bienek A, McAlister FA, Campbell NR; Outcomes Research Task Force of the Canadian Hypertension Education Program. Epidemiology of Hypertension in Canada: An Update. Can J Cardiol. 2016 May;32(5):687-94. doi: 10.1016/j.cjca.2015.07.734. Epub 2015 Aug 15.
Results Reference
background
PubMed Identifier
8622248
Citation
Kannel WB. Blood pressure as a cardiovascular risk factor: prevention and treatment. JAMA. 1996 May 22-29;275(20):1571-6.
Results Reference
background
PubMed Identifier
26724178
Citation
Ettehad D, Emdin CA, Kiran A, Anderson SG, Callender T, Emberson J, Chalmers J, Rodgers A, Rahimi K. Blood pressure lowering for prevention of cardiovascular disease and death: a systematic review and meta-analysis. Lancet. 2016 Mar 5;387(10022):957-967. doi: 10.1016/S0140-6736(15)01225-8. Epub 2015 Dec 24.
Results Reference
background
PubMed Identifier
3336140
Citation
Pickering TG, James GD, Boddie C, Harshfield GA, Blank S, Laragh JH. How common is white coat hypertension? JAMA. 1988 Jan 8;259(2):225-8.
Results Reference
background
PubMed Identifier
12799100
Citation
Myers MG, Valdivieso MA. Use of an automated blood pressure recording device, the BpTRU, to reduce the "white coat effect" in routine practice. Am J Hypertens. 2003 Jun;16(6):494-7. doi: 10.1016/s0895-7061(03)00058-x.
Results Reference
background
PubMed Identifier
23969286
Citation
Edwards C, Hiremath S, Gupta A, McCormick BB, Ruzicka M. BpTRUth: do automated blood pressure monitors outperform mercury? J Am Soc Hypertens. 2013 Nov-Dec;7(6):448-53. doi: 10.1016/j.jash.2013.07.002. Epub 2013 Aug 19.
Results Reference
background
PubMed Identifier
28449828
Citation
Leung AA, Daskalopoulou SS, Dasgupta K, McBrien K, Butalia S, Zarnke KB, Nerenberg K, Harris KC, Nakhla M, Cloutier L, Gelfer M, Lamarre-Cliche M, Milot A, Bolli P, Tremblay G, McLean D, Tran KC, Tobe SW, Ruzicka M, Burns KD, Vallee M, Prasad GVR, Gryn SE, Feldman RD, Selby P, Pipe A, Schiffrin EL, McFarlane PA, Oh P, Hegele RA, Khara M, Wilson TW, Penner SB, Burgess E, Sivapalan P, Herman RJ, Bacon SL, Rabkin SW, Gilbert RE, Campbell TS, Grover S, Honos G, Lindsay P, Hill MD, Coutts SB, Gubitz G, Campbell NRC, Moe GW, Howlett JG, Boulanger JM, Prebtani A, Kline G, Leiter LA, Jones C, Cote AM, Woo V, Kaczorowski J, Trudeau L, Tsuyuki RT, Hiremath S, Drouin D, Lavoie KL, Hamet P, Gregoire JC, Lewanczuk R, Dresser GK, Sharma M, Reid D, Lear SA, Moullec G, Gupta M, Magee LA, Logan AG, Dionne J, Fournier A, Benoit G, Feber J, Poirier L, Padwal RS, Rabi DM; Hypertension Canada. Hypertension Canada's 2017 Guidelines for Diagnosis, Risk Assessment, Prevention, and Treatment of Hypertension in Adults. Can J Cardiol. 2017 May;33(5):557-576. doi: 10.1016/j.cjca.2017.03.005. Epub 2017 Mar 10. Erratum In: Can J Cardiol. 2017 Dec;33(12 ):1733-1734.
Results Reference
background
PubMed Identifier
26551272
Citation
SPRINT Research Group; Wright JT Jr, Williamson JD, Whelton PK, Snyder JK, Sink KM, Rocco MV, Reboussin DM, Rahman M, Oparil S, Lewis CE, Kimmel PL, Johnson KC, Goff DC Jr, Fine LJ, Cutler JA, Cushman WC, Cheung AK, Ambrosius WT. A Randomized Trial of Intensive versus Standard Blood-Pressure Control. N Engl J Med. 2015 Nov 26;373(22):2103-16. doi: 10.1056/NEJMoa1511939. Epub 2015 Nov 9. Erratum In: N Engl J Med. 2017 Dec 21;377(25):2506.
Results Reference
background
PubMed Identifier
21300709
Citation
Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Grant FC, Kaczorowski J. Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial. BMJ. 2011 Feb 7;342:d286. doi: 10.1136/bmj.d286.
Results Reference
background
PubMed Identifier
7662244
Citation
Myers MG, Oh PI, Reeves RA, Joyner CD. Prevalence of white coat effect in treated hypertensive patients in the community. Am J Hypertens. 1995 Jun;8(6):591-7. doi: 10.1016/0895-7061(95)00049-U.
Results Reference
background
PubMed Identifier
19417634
Citation
Myers MG, Valdivieso M, Kiss A. Consistent relationship between automated office blood pressure recorded in different settings. Blood Press Monit. 2009 Jun;14(3):108-11. doi: 10.1097/MBP.0b013e32832c5167.
Results Reference
background
PubMed Identifier
28159816
Citation
Agarwal R. Implications of Blood Pressure Measurement Technique for Implementation of Systolic Blood Pressure Intervention Trial (SPRINT). J Am Heart Assoc. 2017 Feb 3;6(2):e004536. doi: 10.1161/JAHA.116.004536.
Results Reference
background

Learn more about this trial

Comparison of Blood Pressure Measurement With an Omron HEM-907 Device in a Clinical Setting and a Research Setting