Efficacy Comparison of High and Very High Power Short Duration Pulmonary Vein Isolation (HPSD-Remap)
Primary Purpose
Atrial Fibrillation
Status
Recruiting
Phase
Not Applicable
Locations
Hungary
Study Type
Interventional
Intervention
QDot Micro Catheter
Sponsored by
About this trial
This is an interventional treatment trial for Atrial Fibrillation focused on measuring atrial fibrillation, catheter ablation, high-power, short-duration, pulmonary vein isolation, durability
Eligibility Criteria
Inclusion Criteria:
- Symptomatic paroxysmal/persistent AF
- Age >18 years
- Willingness to sign informed consent form.
Exclusion Criteria:
- Contraindication to ablation
- Contraindication of long-term anticoagulation
- Long-standing persistent AF
- History of PVI
- History of cardiac surgery
- Pregnancy
- Active malignancy
- Life expectancy <1 year
- Valvular AF
- Reversible cause of AF (e.g. hyperthyroidism).
Sites / Locations
- Semmelweis UniversityRecruiting
Arms of the Study
Arm 1
Arm 2
Arm Type
Active Comparator
Active Comparator
Arm Label
Pulmonary vein isolation with 50W energy setting
Pulmonary vein isolation with 90W energy setting
Arm Description
In this group of subjects, the initial pulmonary vein isolation procedure will be performed using 50W radiofrequency energy. This power setting will be used for all the ablation points.
In this group of subjects, the initial pulmonary vein isolation procedure will be performed using 90W radiofrequency energy. This power setting will be used for all the ablation points.
Outcomes
Primary Outcome Measures
Pulmonary vein reconnection
Presence or absence of pulmonary vein reconnection at 3 month remapping procedure
Secondary Outcome Measures
Procedure time
Procedure time of the initial pulmonary vein isolation procedure
Fluoroscopy time
Fluoroscopy time of the initial pulmonary vein isolation procedure
Fluoroscopy dose
Fluoroscopy dose of the initial pulmonary vein isolation procedure
Major complication
Major complication of the initial pulmonary vein isolation procedure
First pass isolation
First pass isolation during the initial pulmonary vein isolation procedure
Acute reconnection
Acute reconnection during the initial pulmonary vein isolation procedure
Major complication of the second (remapping) procedure
Major complication of the second (remapping) procedure
Full Information
1. Study Identification
Unique Protocol Identification Number
NCT05459831
Brief Title
Efficacy Comparison of High and Very High Power Short Duration Pulmonary Vein Isolation
Acronym
HPSD-Remap
Official Title
Efficacy Comparison of High and Very High Power Short Duration Pulmonary Vein Isolation
Study Type
Interventional
2. Study Status
Record Verification Date
July 2022
Overall Recruitment Status
Recruiting
Study Start Date
February 15, 2022 (Actual)
Primary Completion Date
December 2022 (Anticipated)
Study Completion Date
February 2023 (Anticipated)
3. Sponsor/Collaborators
Responsible Party, by Official Title
Principal Investigator
Name of the Sponsor
Semmelweis University
4. Oversight
Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
No
Data Monitoring Committee
Yes
5. Study Description
Brief Summary
Objective: To evaluate the long-term durability of PVI performed with 90W and 50W (QDot catheter).
Primary hypothesis: Pulmonary vein isolation is equally highly effective when performed with 50W guided by ablation index (QMODE) and inter-tag distance <5 mm and when performed with 90W (QMODE+) and inter-tag distance <5 mm.
Study Design: 40 patients will be randomized in a 1:1 ratio to 50W and 90W PVI.
Study procedures: Initial PVI will be performed via femoral access and double transseptal puncture guided by fluoroscopy and pressure monitoring. If necessary intracardiac echocardiography (ICE) will be used for the transseptal puncture. A fast anatomical left atrial map will be created with a multipolar mapping catheter; then, point-by-point PVI will be performed with QDot catheter with 50W or 90W according to randomization. Three months after the initial procedure, patients will undergo a repeated high density left atrial mapping with PentaRay or OctaRay catheter to evaluate the durability of the PVI. Exact gap numbers and locations will be registered. If necessary, reablation of the pulmonary veins (rePVI) will be performed.
Inclusion criteria: Symptomatic paroxysmal/persistent AF, >18 years, willingness to sign informed consent form.
Exclusion criteria: Long-standing persistent AF, active malignancy, contraindication for anticoagulant therapy, life expectancy <1 year, valvular AF, hyperthyroidism.
Endpoints: At repeat procedure, a high density left atrial map will be created with PentaRay or if available OctaRay catheter to evaluate the durability of the PVI. Primary endpoint will be pulmonary vein reconnection.
Detailed Description
Background
Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia in adults. Its prevalence increases with age, affecting 3/100 people, and its incidence is also correlated with age, affecting 30/100,000 people per year. Its clinical significance stems from common complications (stroke, heart failure) and concomitant higher mortality.
The most effective treatment method of atrial fibrillation is catheter ablation, which seeks to maintain sinus rhythm. Isolation of the triggering pulmonary veins (PVI) from the left atrium is the cornerstone of AF ablation. PVI has undergone significant technical development over the past decade; however, the chances of arrhythmia recurrence and the occurrence of procedural complications are not negligible.
Experiments have shown that lesions with a larger surface diameter but less depth develop during HPSD ablation; the higher the energy is, the shallower and wider the lesions are. These lesion properties make it easier to create a solid line from the ablation points, while the shallower lesion causes less damage to the surrounding tissues. The latest application of HPSD technology is the so-called "very high power short duration" (vHPSD) ablation when we apply 90W power. No data are available on long-term efficacy or safety comparing ablation with HPSD vs. vHPSD.
Objective: To evaluate the durability of radiofrequency PVI performed with 90W (QMODE plus) and 50W (QMODE) with the QDot catheter.
Primary hypothesis: Pulmonary vein isolation is equally highly effective when performed with 50W guided by ablation index (QMODE) and inter-tag distance <5 mm and when performed with 90W (QMODE+) and inter-tag distance <5 mm.
Study Design: Single-center, prospective, randomized study. Patients will be randomly assigned in a 1:1 ratio to HPSD (50W) and vHPSD (90W) PVI.
Research period Planned research period from the second quarter of 2022 for a maximum of 1 year.
Study subjects
• Planned number of included subjects: 40.
Study procedures:
Initial PVI will be performed via femoral access after double transseptal puncture guided by fluoroscopy and pressure monitoring. If necessary ICE will be used for the transseptal puncture. A fast anatomical left atrial map will be created with a multipolar mapping catheter; then, point-by-point PVI will be performed with QDot catheter with 50W or 90W according to randomization. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed by multipolar catheters. If PVI is not complete at this point, it will be completed to reach complete isolation of all veins. After that, a 20 minutes waiting period will be the next step to evaluate presence or absence of acute reconnection of the PV-s. In case of acute reconnection, touch-up applications will be delivered to complete the PVI.
Three months after the initial procedure, patients will undergo a repeated high density left atrial mapping with PentaRay or, if available OctaRay catheter to evaluate the durability of the PVI. Exact gap numbers and locations will be registered (based on the 16 segment PV model). If necessary, rePVI will be performed.
Statistical analysis Continuous variables are expressed as mean and standard deviation, or median and interquartile ranges, depending on the normal of non-normal distribution. Categorical variables are expressed in numbers and percentages. Continuous variables are compared by parametric or non-parametric tests, depending on the distribution, while categorical variables are compared by Chi-square test or Fisher's exact test. The predictive value of different ablation techniques in outcome is planned to be analyzed by logistic regression. A two-tailed p-value of <0.05 will be considered statistically significant. Statistical analyses will be performed using IBM SPSS 25 (Apache Software Foundation, USA) and GraphPad Prism 7.1 (GraphPad Softwares Inc., USA) software products.
Data management All study patient data is entered into Semmelweis University's electronic database (accessible only from Semmelweis University's network) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data is restricted to institutional leaders and program professional leaders.
Personal and material requirements The personal and material requirements of the research are available at the Heart and Vascular Center of Semmelweis University. The personal requirements are given by the staff of the Electrophysiological Laboratory, and the material conditions are provided by the equipment of the same laboratory.
Ethics approval
An ethics approval was obtained from the Hungarian National Public Health Center (No. 8119-2/2022/EÜIG).
6. Conditions and Keywords
Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Atrial Fibrillation
Keywords
atrial fibrillation, catheter ablation, high-power, short-duration, pulmonary vein isolation, durability
7. Study Design
Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Parallel Assignment
Masking
None (Open Label)
Allocation
Randomized
Enrollment
40 (Anticipated)
8. Arms, Groups, and Interventions
Arm Title
Pulmonary vein isolation with 50W energy setting
Arm Type
Active Comparator
Arm Description
In this group of subjects, the initial pulmonary vein isolation procedure will be performed using 50W radiofrequency energy. This power setting will be used for all the ablation points.
Arm Title
Pulmonary vein isolation with 90W energy setting
Arm Type
Active Comparator
Arm Description
In this group of subjects, the initial pulmonary vein isolation procedure will be performed using 90W radiofrequency energy. This power setting will be used for all the ablation points.
Intervention Type
Device
Intervention Name(s)
QDot Micro Catheter
Intervention Description
Atrial fibrillation radiofrequency ablation with 90W or 50W energy setting.
Primary Outcome Measure Information:
Title
Pulmonary vein reconnection
Description
Presence or absence of pulmonary vein reconnection at 3 month remapping procedure
Time Frame
3 month
Secondary Outcome Measure Information:
Title
Procedure time
Description
Procedure time of the initial pulmonary vein isolation procedure
Time Frame
Immediately after the procedure
Title
Fluoroscopy time
Description
Fluoroscopy time of the initial pulmonary vein isolation procedure
Time Frame
Immediately after the procedure
Title
Fluoroscopy dose
Description
Fluoroscopy dose of the initial pulmonary vein isolation procedure
Time Frame
Immediately after the procedure
Title
Major complication
Description
Major complication of the initial pulmonary vein isolation procedure
Time Frame
3 month
Title
First pass isolation
Description
First pass isolation during the initial pulmonary vein isolation procedure
Time Frame
Immediately after the procedure
Title
Acute reconnection
Description
Acute reconnection during the initial pulmonary vein isolation procedure
Time Frame
Immediately after the procedure
Title
Major complication of the second (remapping) procedure
Description
Major complication of the second (remapping) procedure
Time Frame
1 week
10. Eligibility
Sex
All
Minimum Age & Unit of Time
18 Years
Maximum Age & Unit of Time
100 Years
Accepts Healthy Volunteers
No
Eligibility Criteria
Inclusion Criteria:
Symptomatic paroxysmal/persistent AF
Age >18 years
Willingness to sign informed consent form.
Exclusion Criteria:
Contraindication to ablation
Contraindication of long-term anticoagulation
Long-standing persistent AF
History of PVI
History of cardiac surgery
Pregnancy
Active malignancy
Life expectancy <1 year
Valvular AF
Reversible cause of AF (e.g. hyperthyroidism).
Central Contact Person:
First Name & Middle Initial & Last Name or Official Title & Degree
Nándor Szegedi, Ph.D.
Phone
+36206663442
Email
nandorszegedi@gmail.com
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
László Gellér, Ph.D., DSc
Organizational Affiliation
Semmelweis University
Official's Role
Study Chair
Facility Information:
Facility Name
Semmelweis University
City
Budapest
ZIP/Postal Code
1122
Country
Hungary
Individual Site Status
Recruiting
Facility Contact:
First Name & Middle Initial & Last Name & Degree
Nándor Szegedi, Ph.D.
Phone
206663442
Email
nandorszegedi@gmail.com
12. IPD Sharing Statement
Learn more about this trial
Efficacy Comparison of High and Very High Power Short Duration Pulmonary Vein Isolation
We'll reach out to this number within 24 hrs