Microfidelity (MIFI) Ablation Technology Versus Standard Ablation Catheter for Atrioventricular Nodal Ablation

Primary Purpose

Atrial Fibrillation

Status
Completed
Phase
Not Applicable
Locations
United States
Study Type
Interventional
Intervention
MIFI Catheter
Standard Ablation Catheter
Sponsored by
University of Florida
About
Eligibility
Locations
Arms
Outcomes
Full info

About this trial

This is an interventional treatment trial for Atrial Fibrillation focused on measuring ablation, atrioventricular node, microfidelity

Eligibility Criteria

18 Years - 95 Years (Adult, Older Adult)All SexesAccepts Healthy Volunteers

Inclusion Criteria:

  1. Patients with a diagnosis of persistent or permanent atrial fibrillation documented on electrocardiography
  2. Patients must meet American College of Cardiology and Heart Rhythm Society "ACC/HRS" guidelines for atrioventricular nodal ablation procedure
  3. Patients must be available for at least 1 month post procedure
  4. Patients must be greater than or equal to 18 years old.

Exclusion Criteria:

1. Patients who do not meet ACC/HRS indications for av nodal ablation

Sites / Locations

  • University of Florida

Arms of the Study

Arm 1

Arm 2

Arm Type

Active Comparator

Experimental

Arm Label

Bipolar Ablation Catheter

MIFI Ablation Catheter

Arm Description

Subjects will be randomized by random computer programming to receive standard ablation catheter for their procedure.

Subjects will be randomized by random computer programming to receive an ablation with MIFI technology. MIFI Catheter contains tightly spaced multielectrode pattern

Outcomes

Primary Outcome Measures

The Time Differences Between the 2 Arms for Acute Success of Ablation Identified by a Junctional Rhythm or Complete Heart Block
Identified by a junctional rhythm or complete heart block. Time zero is when the ablation procedure begins and the successful outcomes are either complete heart block or junctional rhythm.

Secondary Outcome Measures

Time From Application of Radio-frequency Energy to Acute Success
The target will be where the HIS electrogram amplitude is greatest and where the QRS complex from pacing is narrowest.

Full Information

First Posted
March 21, 2017
Last Updated
August 31, 2021
Sponsor
University of Florida
search

1. Study Identification

Unique Protocol Identification Number
NCT03114124
Brief Title
Microfidelity (MIFI) Ablation Technology Versus Standard Ablation Catheter for Atrioventricular Nodal Ablation
Official Title
Microfidelity (MIFI) Ablation Technology Versus Standard Ablation Catheter for Atrioventricular Nodal Ablation A Comparative Pilot Study of Time to Success
Study Type
Interventional

2. Study Status

Record Verification Date
August 2021
Overall Recruitment Status
Completed
Study Start Date
May 23, 2017 (Actual)
Primary Completion Date
September 26, 2019 (Actual)
Study Completion Date
November 9, 2019 (Actual)

3. Sponsor/Collaborators

Responsible Party, by Official Title
Sponsor
Name of the Sponsor
University of Florida

4. Oversight

Studies a U.S. FDA-regulated Drug Product
No
Studies a U.S. FDA-regulated Device Product
Yes
Product Manufactured in and Exported from the U.S.
No
Data Monitoring Committee
No

5. Study Description

Brief Summary
The aim of our study is to investigate the comparative efficacy of high fidelity multi electrode ablation catheters vs the standard bipolar configuration in success of AV nodal ablation This study will provide insights on the use of new technology where application may increase efficacy, promote patient and physician safety and decrease costs.
Detailed Description
Introduction In some elderly patients with atrial fibrillation (AF), especially in combination with heart failure, a rate control strategy may be preferred. When pharmacological therapy is ineffective or not tolerated, it is reasonable to perform atrioventricular (AV) node ablation with ventricular pacing as a class IIA indication per current guidelines. Usually, the procedure is simple and straightforward and complete heart block can be achieved without any difficulty. However, this "simple" procedure can sometimes prove to be a most difficult case. The most common reason for failure to achieve complete heart block is the inability to localize the compact AV node using the His signal with standard intracardiac electrograms. As these patients come to the laboratory in AF, the His signal may be obscured by AF waves. In some patients with a deeper intramyocardial location of the His bundle and compact AV node it becomes necessary to produce deeper ablation lesions using an irrigated catheter to achieve block. In patients with AF, the target of ablation for the "ablate and pace" approach is the compact AV node, located at the apex of the triangle of Koch. Ideally, ablation is performed at the most proximal penetrating part of the His bundle in order to maintain a proximal automatic junctional rhythm and avoid pacemaker dependence. Para Hisian pacing is most commonly used to reveal the presence of a septal accessory pathway. The His bundle is a deep insulated structure and it is difficult to capture it at usual energy outputs. Using a high- output pacing (usually 20 mA at 2 msec) it is possible to directly capture the deeply situated His bundle, which is confirmed by a narrower QRS complex on the paced electrograms. Thus, high-output pacing can be utilized to map the His bundle area in difficult situations. By applying this electrophysiologic principle of differential tissue capture to help identify the location of the compact AV node, which is in close proximity to the His bundle. Ventricular pacing was performed initially at high output to capture both the basal right ventricular myocardium and the His bundle and the output was gradually lowered to lose His bundle capture. The QRS duration is relatively narrow with high output pacing and increases when the pacing output is lowered, representing ventricular myocardial capture alone. Finally, loss of ventricular capture is seen with further reduction of pacing output. This maneuver has been shown to aid in determining the proximity of the ablation catheter to compact AV node as was validated by successful ablation at this site. Parahisian pacing in conjunction with av nodal ablation has recently been described in the literature. A novel catheter with three mini electrodes within the ablation tip (IntellaTip MiFi, Boston Scientific, Boston, MA) may enhance the available data for such a signal dependent technique. In this catheter, bipolar signals can be recorded between the three 0.8-mm-wide electrodes that are arranged radially 1.3 mm from the end of the catheter alongside the standard distal and proximal bipolar recordings. Animal studies have already demonstrated that the mini electrodes in this novel catheter are more accurate in identifying conducting gaps in linear ablations than conventional electrode recordings. Study Aims: The aim of our study is to investigate the comparative efficacy of high fidelity multi electrode ablation catheters vs the standard bipolar configuration in success of AV nodal ablation This study will provide insights on the use of new technology where application may increase efficacy, promote patient and physician safety and decrease costs. Primary and secondary objectives Primary endpoints Acute success of ablation identified by a junctional rhythm or complete heart block Time from application of radiofrequency energy to acute success Secondary endpoints include Procedure time Radiation time Frequency of ablation application Duration of ablation application

6. Conditions and Keywords

Primary Disease or Condition Being Studied in the Trial, or the Focus of the Study
Atrial Fibrillation
Keywords
ablation, atrioventricular node, microfidelity

7. Study Design

Primary Purpose
Treatment
Study Phase
Not Applicable
Interventional Study Model
Parallel Assignment
Model Description
Standard ablation catheter vs MIFI catheter
Masking
Investigator
Masking Description
Variable block sizes will be used to improve the blinding of the clinical team. This information will then be used to design a fully-powered randomized study to compare these two arms. Patients will be randomly assigned to study arms. The project biostatistician and team will generate a set of envelopes which contain the random assignments for each patient. Once a patient has consented to participate, the next envelope will be opened for that patient and the catheter assignment will be revealed. This will maintain the blinding of the clinical team.
Allocation
Randomized
Enrollment
30 (Actual)

8. Arms, Groups, and Interventions

Arm Title
Bipolar Ablation Catheter
Arm Type
Active Comparator
Arm Description
Subjects will be randomized by random computer programming to receive standard ablation catheter for their procedure.
Arm Title
MIFI Ablation Catheter
Arm Type
Experimental
Arm Description
Subjects will be randomized by random computer programming to receive an ablation with MIFI technology. MIFI Catheter contains tightly spaced multielectrode pattern
Intervention Type
Device
Intervention Name(s)
MIFI Catheter
Intervention Description
Multielectrode Catheter: standard ablation catheter that has the ability to record low and high frequency signals with high fidelity
Intervention Type
Device
Intervention Name(s)
Standard Ablation Catheter
Intervention Description
Standard of care for patients with certain cardiac arrhythmias.
Primary Outcome Measure Information:
Title
The Time Differences Between the 2 Arms for Acute Success of Ablation Identified by a Junctional Rhythm or Complete Heart Block
Description
Identified by a junctional rhythm or complete heart block. Time zero is when the ablation procedure begins and the successful outcomes are either complete heart block or junctional rhythm.
Time Frame
60 seconds
Secondary Outcome Measure Information:
Title
Time From Application of Radio-frequency Energy to Acute Success
Description
The target will be where the HIS electrogram amplitude is greatest and where the QRS complex from pacing is narrowest.
Time Frame
up to 3000 seconds

10. Eligibility

Sex
All
Minimum Age & Unit of Time
18 Years
Maximum Age & Unit of Time
95 Years
Accepts Healthy Volunteers
Accepts Healthy Volunteers
Eligibility Criteria
Inclusion Criteria: Patients with a diagnosis of persistent or permanent atrial fibrillation documented on electrocardiography Patients must meet American College of Cardiology and Heart Rhythm Society "ACC/HRS" guidelines for atrioventricular nodal ablation procedure Patients must be available for at least 1 month post procedure Patients must be greater than or equal to 18 years old. Exclusion Criteria: 1. Patients who do not meet ACC/HRS indications for av nodal ablation
Overall Study Officials:
First Name & Middle Initial & Last Name & Degree
John N Catanzaro, MD
Organizational Affiliation
University of Florida
Official's Role
Principal Investigator
Facility Information:
Facility Name
University of Florida